调强放疗小野小跳数TPS计算值与实际测量值间差异的研究 被引量:3

Study of difference between calculation of TPS and measure of small segment and small monitor unit in IMRT
下载PDF
导出
摘要 目的对调强放疗中的小野和小跳数在TPS计算与加速器实际测量值之间的差异进行研究,以指导调强放疗技术在临床中的应用。方法在TPS中设计小野和小跳数计划并导出,与在加速器上进行的实际测量结果进行比较。结果在小跳数测量中,2MU以上的通过率均>83%;在小野测量中,面积大于2 cm×2 cm的射野通过率均>80%。结论对TPS计算的小野和小跳数需进行手动适当调整;MLC的到位精度对小野的影响比较大,MLC的质量控制不容忽视。 Objective To study the difference between the calculation of TPS and the measurement of small segment and small moni-tor unit in IMRT to instruct the clinical application.Methods The small segment and small monitor unit plan were created and exported in TPS,and compared with the results of measurement in LA.Results In small MU testing the passing rate of over 2MU was more than 83%;in small segment testing the passing rate of the area over 2cm ×2 cm was more than 80%.Conclusion Manual operation should be per-formed in the mall segment and small monitor unit calculated in TPS.MLC leaf positional accuracy is important for the small field,so QA/QC for MLC can not be ignored.
出处 《安徽医学》 2014年第5期556-558,共3页 Anhui Medical Journal
关键词 调强放疗 小野 小跳数 IMRT Small segment Small monitor unit
  • 相关文献

参考文献4

二级参考文献23

  • 1马金利,蒋国梁,傅小龙,廖源,吴开良,周莉钧.窄带野胶片检查在多叶准直器质量保证中作用探讨[J].中华放射肿瘤学杂志,2004,13(2):123-127. 被引量:10
  • 2Ezzell GA, Galvin JM, Low DA, et al. Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee. Med Phys, 2003, 30: 2089-2115.
  • 3Kung JH, Chen GT. Intensity modulated radiotherapy dose delivery error from radiation field offset inaccuracy. Med Phys, 2000, 27: 1617-1622.
  • 4Sharpe MB, Miller MB, Yan D, et al. Monitor unit settings for intensity modulated beams delivered using a step-and-shoot approach. Med Phys, 2000, 27: 2719-2725.
  • 5Chui CS, Spirou S, and LoSasso T. Testing of dynamic multileaf collimation. Med Phys, 1996, 23 : 635-641.
  • 6Low DA, Sohn JW, Klein EE, et al. Characterization of a commercial multileaf collimator used for intensity modulated radiation therapy. Med Phys, 2001, 28: 752-756.
  • 7Childress NL,and Rosen Ⅱ. The design and testing of novel clinical parameters for dose comparison. Int J Radiat Oncol Biol Phys, 2003, 56 : 1464-1479.
  • 8Low DA, Harms WB, Mutic S, et al. A technique for the quantitative evaluation of dose distributions. Med Phys, 1998, 25: 656-661.
  • 9Ezzell GA,Galvin JM,Low D,et al.Guidance document on delivery,treatment planning, and clinical implementation of IMRT: report of the IMRT subcommittee of the AAPM radiation therapy committee[].Medical Physics.2003
  • 10Dong L,Antolak J,Salehpour M,et al.Patient-specific point dose measurement for IMRT monitor unit verification[].International Journal of Radiation Oncology Biology Physics.2003

共引文献118

同被引文献23

引证文献3

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部