期刊文献+

速差催化荧光法同时测定痕量钴和铜 被引量:5

SIMULTANEOUS DETERMINATION OF COBALT AND COPPER BY DIFFERENTIAL RATE CATALYTIC FLUORIMETRY
下载PDF
导出
摘要 基于在硼酸 硼砂介质中 ,钴和铜都能催化过氧化氢氧化还原型罗丹明 6G的反应 ,并且产生荧光强度具有加和性的特性 ,建立了速差催化荧光法同时测定痕量钴和铜的新方法。钴和铜的检出限分别为 0 .90ng·ml- 1和 0 .0 5 3μg·ml- 1;测定范围分别为 1.0 0~ 2 0 .0 0ng·ml- 1和 0 .10~1.2 0 μg·ml- 1。将方法用于生物样品中痕量钴和铜的同时测定 。 Differential rate catalytic fluorimetry was applied to the simultaneous determination of cobalt and copper, based on the differential rate of the catalytic effect of cobalt and copper on the oxidation of reduced rhodamine 6G with hydrogen peroxide in the presence of triethanolamine in Na 2B 4O 7-H 3BO 3 buffer solution at pH=8.40. The detection limit for cobalt and copper are 0.90ng·ml -1 and 0.053μg·ml -1 respectively. The linear ranges for the determination are 1.00~20.00ng·ml -1 for cobalt and 0.10~1.20μg·ml -1 for copper. The method has been used to determine simultaneously cobalt and coppere in hair, human serum and fingernail samples with satisfactory results.
出处 《理化检验(化学分册)》 CAS CSCD 北大核心 2002年第7期350-351,354,共3页 Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
关键词 速差催化荧光法 同时测定 痕量分析 人体组织 Differential rate catalytic fluorimetry Cobalt Copper Simultaneous determination
  • 相关文献

参考文献3

二级参考文献40

  • 1方国桢,唐晓萍.速差对数外推法同时分光光度测定铂和金及其与双波长法的比较[J].分析试验室,1993,12(4):7-11. 被引量:8
  • 2郭祥群,1991年
  • 3傅献彩,物理化学(第3版),1985年
  • 4方国桢,1992年
  • 5郭样群,高等学校化学学报,1992年,13卷,4期,450页
  • 6刘国慧,1992年
  • 7罗庆尧,分光光度分析,1992年
  • 8方国桢,1991年
  • 9Xing R,Anal Chim Acta,1991年,251卷,313页
  • 10Yang Y,Chem Res Chin Univ,1991年,7卷,79页