国产回弹式眼压计在儿童眼压测量中的应用

Application of the domestic rebounder tonometer in children
下载PDF
导出
摘要 目的:观察国产回弹式眼压计在儿童眼压监测中的应用。方法:对本院眼视光门诊就诊的108例儿童进行两眼眼压测量,年龄大于13岁的儿童回弹式眼压计(rebound tonometer,RBT)测量完成后左眼进行Goldmann压平式眼压计(Goldmann applanation tonometry,GAT)眼压测量。结果:两眼均顺利完成RBT眼压测量的儿童100例,成功率92.6%(100/108),其中13岁以上24例。24例左眼RBT和GAT眼压平均值分别为16.0±3.0mmHg(1kPa=7.5mmHg)和15.7±2.8mmHg,两者差异无统计学意义(P>0.05),两者具有良好的相关性(r=0.849,P<0.05)。100例儿童的左右眼眼压均呈正态分布,平均值分别为16.0±2.8mmHg和16.2±2.6mmHg,两者差异无统计学意义(P>0.05),两者具有良好的相关性(r=0.863,P<0.05)。结论:国产回弹式眼压计在儿童眼压测量中简单易操作,无需麻醉,准确性较高,能够较好满足临床儿童眼压测量的需求。 AIM :To evaluate the clinica rebound tonometer (RBT) application of the domestic in measuring intraocular pressure (IOP) in healthy children. METHODS: IOP measurement was measured in bilateral eyes of 108 cases with domestic RBT. Then the IOOP for the left eyes of the children older than 13 years was obtained with Goldmann applanation tonometry (GAT). RESULTS: Of the 108 children, 100 (92.6%) had IOP measurement both eyes successfully. Of the 100 children, 24 were older than 13 years. The mean IOP values for the left eyes of the 24 children of RBT and GAT were (16. 0± 3.0) mmHg (1kPa=7.5mmHg) and (15.7±2.8) mmHg, respectively. There was no significant difference between the IOP of the two tonometers (P〉0.05). The IOP of RBT had a good liner relationship with that of GAT (r= 0. 849, P〈0.05). The IOP values were normally distributed for the both eyes. Mean IOP values were (16.0±2.8) mmHg and (16. 2± 2. 6) mmHg for the left eyes and right eyes, respectively. There was no statistically significant difference in mean IOP readings between them, which had a good liner relationship (r=0.863, P〈0.05). CONCLUSION: Domestic RBT is a good instrument in detecting IOP in children and it seems to be very comfortable and easy when performing IOP measurement in children without an anesthetic.
作者 潘旭冲
出处 《国际眼科杂志》 CAS 2014年第7期1322-1324,共3页 International Eye Science
关键词 眼压 回弹式眼压计 儿童 intraocular pressure rebound tonometer children
  • 相关文献

参考文献10

二级参考文献53

  • 1张宇燕,孙兴怀.常见眼压计的临床评价[J].眼视光学杂志,2004,6(2):132-135. 被引量:31
  • 2Coleman AL, Caprioli J. The logic behind target intraoeular pressure. Am J Ophthalmol,2009,147:379-380.
  • 3Tonnu PA, Ho T, Newson T, et al. The influence of central corneal thickness and age on intraocular pressure measured by pneumatonometry, noncontact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol, 2005, 89: 851-854.
  • 4Fan DS, Chiu TY, Congdon N, et al. Measurement of intraocular pressure with pressure phosphene tonometry in children. J Pediatr Ophthalmol Strabismus, 2011,48 : 167-173.
  • 5Kageyama M, Hirooka K, Baba T, et al. Comparison of I care rebound tonometer with noncontact tonometer in healthy children. J Glaucoma, 2011, 20:63-66.
  • 6Akinci A, Cetinkaya E, Aycan Z, et al. Relationship between intraocular pressure and obesity in children. J Glaucoma, 2007, 16:627-630.
  • 7Kass MA, Gordon MO, Gao F, et al. Delaying treatment of ocular hypertension: the ocular hypertension treatment study. Arch Ophthalmol, 2010, 128:276-287.
  • 8Kass MA, Heuer DK, Higginbotham EJ, et al. The ocular hypertension treatment study: a randomized trial determines that topical ocular hypotensive medication delays or prevents the onset of primary open-angle glaucoma. Arch Ophthalmol, 2002, 120 : 701-713.
  • 9Cavallotti C, Pescosolido N, Artico M, et al. Occurrence of catecholaminergic nerve fibers in the human uveoscleral tissue in conditions of normal and raised intraocular pressure. Int Ophthalmol, 2001,24 : 133-139.
  • 10Tanaka H, Borres M, Thulesius O, et al. Blood pressure and cardiovascular autonomic function in healthy children and adolescents. J Pediatr, 2000, 137: 63-67.

共引文献29

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部