双源CT双能量扫描在肝脏成像中的应用研究 被引量:9

Evaluation of Liver Imaging by Dual-source Dual-energy CT
下载PDF
导出
摘要 目的:比较双源CT双能量扫描与常规扫描在肝脏成像中的图像质量及辐射剂量。方法:随机选择腹部常规平扫、增强扫描患者34例,通过伦理委员会审查征得病人知情同意后,行双源CT平扫及增强扫描,增强扫描门脉期选择双能量扫描方案,记录扫描中患者所接受的辐射剂量。扫描结束后,应用后处理软件重建出肝脏虚拟平扫图像,将肝脏常规平扫与虚拟平扫的图像质量、辐射剂量、图像噪声(SD)、信噪比(Mean/SD)进行对比分析。图像质量等级评价结果采用秩和检验;SD与Mean/SD值、辐射剂量评价若满足正态、方差齐采用t检验,若不满足则采用秩和检验。结果:肝脏常规平扫与虚拟平扫辐射剂量无统计学差异(P>0.05)。常规平扫与虚拟平扫图像质量等级评价结果、噪声(SD)、信噪比(Mean/SD)有统计学差异(P﹤0.05)。常规平扫图像质量优于虚拟平扫,虚拟平扫的噪声低于常规平扫、信噪比高于常规平扫。结论:肝脏双能量虚拟平扫与常规平扫相比,图像噪声低、信噪比高,在肝脏检查中可替代常规平扫,减低患者的扫描时间及辐射剂量。 Purpose:To compare the image quality and radiation dosage of virtual non-contrast for liver imaging with second-generation,dual-source CT in comparison with routine non-contrast images.Methods:Institutional review board approval was obtained.Thirty-four patients underwent liver contrast-enhanced dual-energy CT scan that included routine non-contrast images acquired in single-energy mode,corticomedullary phase images acquired in dual-energy mode,and nephrographic phase images acquired in singleenergy mode.The radiation dosages in patients received by scanning were recorded.The liver virtual noncontrast images were reconstructed after scanning.The acceptance level and image quality of virtual and routine non-contrast images were evaluated.And the image quality,radiation dose,and image noise (SD),signal to noise ratio (Mean/SD) of them were analyzed.The image quality evaluation results were analyzed with rank sum test.The differences of radiation dosage,SD and Mean/SD value between virtual and routine non-contrast images were analyzed with t-test or rank sum test.Results:The differences of image quality level,SD and Mean/SD between liver routine non-contrast and virtual non-contrast images were with statistical significant (P<0.05).The radiation dosage between routine non-contrast and virtual non-contrast images was with no significant difference (P>0.05).Conclusion:Dual-source dual-energy CT is a reliable imaging technique in the evaluation of liver.Routine non-contrast images can be replaced by virtual noncontrast images with considerable radiation dose reduction.
出处 《中国医学计算机成像杂志》 CSCD 北大核心 2014年第3期250-252,共3页 Chinese Computed Medical Imaging
关键词 双源CT双能量虚拟平扫 常规平扫 图像质量 辐射剂量 Virtual noncontrast with dual-source CT Routine noncontrast images Image quality Radiation doses
  • 相关文献

参考文献12

  • 1Prakash P,Kalra MK,Kambadakone AK,et al.Reducing abdominal CT radiation dose with adaptive statistical iterative reconstruction technique.Invest Radiol,2010,45:202-210.
  • 2石磊,张欢,潘自来,刘欢欢,李向亭,杜联军,丁蓓,宋琦,凌华威,陈克敏,严福华.双能量CT在胃癌虚拟平扫应用的初步探讨[J].中国医学计算机成像杂志,2013,19(1):33-37. 被引量:11
  • 3Menzel H,Schibilla H,Teunen D,et a.l European guidelines on quality criteria for computed tomography.Luxembourg:European Commission,2000:1626.
  • 4Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radiation dose optimization.Radiology,2004,230:619-628.
  • 5SlovisTL.The ALARA conception in pediatric CT:myth or reality.Radiology,2002,223:5.
  • 6Graser A.Dual energy CT in the assessment of renal masses:Can dual energy virtually unenhanced images replace noncontrast scanning.RSNA,2007 Chicago,IL,USA.
  • 7赵艳娥,卢光明,张龙江等.双源双能量CT在肝脏富血供肿瘤中的初步临床研究.中华放射学杂志,2013,能量CT成像临床应用论文集:52-55.
  • 8Silva AC,Lawder HJ,Hara A,et al.Innovations in CT dose reduction strategy:application of the adaptive statistical iterative reconstruction algorithm.A JR,2010,25:194-199.
  • 9王晓梅,边杰,杨俊武,白建新,张策,张健飞,隋鸿锦.加权窗宽及加权窗位的选择在双源CT双能门静脉成像中的应用价值[J].中国临床医学影像杂志,2013,24(8):560-563. 被引量:1
  • 10Graser A.Johnson T R,Chandarana H,et al.Dual energy CT:preliminary observations and potential clinical applications in the abdomen.Eur Radiol,2009,19:13-23.

二级参考文献21

  • 1Behrendt FF, Schmidt B, Plumhans C, et al. hnage fusion in dual energy computed tomography :effect on contrast enhancement, signal-tonoise ratio and image quality in computed tomography angiography. Invest Radiol,2009,44 : 1.
  • 2Nakayama Y. Low tube voltage reduces contrast material and radiation doses on 16-MDCT aortography. A JR,2006,187:490.
  • 3Menzel H, Schibilla H, Teunen D, et al. European guidelines on quality criteria for computed tomography. Luxembourg: European Commission ,2000 : 16262.
  • 4Deetjen A, Mollmann S,Conradi G, et al. Use of automatic exposure control in muhislice computed tomography of the coronaries : comparison of 16-slice and 64-slice scanner data with conventional coronary angiography. Heart ,2007,93 : 1040.
  • 5Scheffel H, Stolzmann P, Frauenfelde TR, et al. Dual-energy contrastenhanced computed tomography for the detection of urinary stone disease. Invest Radio1,2007 ,42 :823.
  • 6Graser A,Johnson TR, Chandarana H, et al. Dual energy CT: preliminary observations and potential clinical applications in the abdomen. Eur Radio1,2009,19 : 13.
  • 7Petersilka M, Bruder H, Krauss B, et al. Technical principles of dual source CT. Eur J Radiol,2008 ,68 :362.
  • 8Flohr TG, McCollough CH, Bruder H, et al. First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol,2006 ,16 :256.
  • 9Johnson TR, Nikolaou K, Wintersperger BJ, et al. Dual source CT cardiac imaging : initial experience. Eur Radiol,2006,16 : 1409.
  • 10Slovis TL. The ALARA conception in pediatric CT:myth or reality? Radiology ,2002,223:5.

共引文献51

同被引文献48

引证文献9

二级引证文献50

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部