摘要 综述了美国NIST元素汞发生器的基准溯源传递方法,通过在输出的汞标准气体中混合加入已知高纯度的201Hg0气体(201Hg2+标准溶液还原雾化),并采用同位素稀释电感耦合等离子体/质谱法(ID ICP/MS)测量加标后201Hg/202Hg同位素丰度比例,计算出待溯源元素汞发生器输出的Hg0质量浓度。该过程实现了从元素汞一级标准物质SRM 3133到201Hg2+标准溶液,再到元素汞发生器输出标准气体的溯源传递,完成了从NIST元素汞基准到仪器供应商生产基准的一级溯源传递,为建立符合我国需求的元素汞溯源传递方法和形成元素汞发生器溯源传递能力提供借鉴。 Overviewing the method of prime traceability and delivery on element mercry generator from the NIST in USA. USA NIST uses ID ICP/MS method to calculate the candidate mercry generator output concentration, it adds known high pure 201Hg0 gas to candidate element mercury generator output calibration gas, then abundance of the spiked isotope ^201Hg will be changed,the abundance ratio of ^201Hg/^202Hg will be measured by ICP/ MS. This process completes mercury traceability which is from the prime SRM 3133 to ^201Hg^202 solution concen- tration,and from ^201Hg^2+ solution concentration to element mercmty generator. This method is tot the vender prime to trace to NIST prime,it is good tot us to establish the element mercmty generator traceability method and capacity which is suitable tot Chinese requirement.
出处 《环境监测管理与技术》 2014年第2期7-11,共5页 The Administration and Technique of Environmental Monitoring
基金 环保公益性行业科研专项基金资助项目(201209050) 国家重大科学仪器设备开发专项基金资助项目(2011YQ060100)
关键词 NIST 元素汞发生器 溯源 同位素稀释电感耦合等离子体 质谱法 NIST Element mercury generator Traceability Isotope dilution inductively coupled plasma- mass spectrometry(ID ICP/MS)
  • 相关文献

参考文献13

二级参考文献47

  • 1岑可法,邱坤赞,朱燕群.中国能源与环境可持续发展问题的探讨(一)[J].发电设备,2004,18(5):245-250. 被引量:12
  • 2王军,赵墨田.同位素稀释质谱法测定酸雨标准物质中微量钾和镁[J].质谱学报,1995,16(2):10-14. 被引量:5
  • 3任建莉,周劲松,骆仲泱,徐璋,钟英杰.燃煤电站汞排放分布及控制研究的进展[J].电站系统工程,2006,22(1):44-46. 被引量:38
  • 4郑用熙.分析化学中的数理统计方法[M].北京:科学出版社,1991..
  • 5Jarvis K.E Gray A.L..电感耦合等离子体质谱手册[M].北京,1997.10.
  • 6ABRAMS S A, GRIFFIN I J, HAWTHORNE K M, et al. Vitamin D receptor fokl polymorphisms affect calcium absorption, kinetics, and bone mineralization rates during puberty [J]. Journal of Bone and Mineral Research, 2005, 20 (6) .. 945-953.
  • 7LI Bing, SUN Yali, YIN Ming. Determination of cerium,neodymium and samarium in biological materials at low levels by isotope dilution inductively coupled plasma mass spectrometry[J]. J Anal At Spectrum, 1999, 14:1 843-1 848.
  • 8BEARY E S, PAULSEN P J, JASSIE L B, et al. Determination of environmental lead using continuous-flow microwave digestion isotope dilution inductively coupled plasma mass spectrometry[J]. Anal Chem, 1997, 69(4): 758-766.
  • 9EBDON L, HILL S J, RIVAS C. Lead speciation in rainwater by isotope dilution-high performance liquid chromatography-inductively coupled plasma-mass spectrometry[J]. Spectrochimica Acta Part B, 1998, 53(2): 289-297.
  • 10USEPA. Standards of performance for new and existing stationary[ R/OL]. [ 2005 - 03 - 15 ]. http ://www. epa. gov/air/mercuryrule/pdfs/camr_final_preamble. pdf.

共引文献75

同被引文献89

引证文献2

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部