摘要
目的:探讨低辐射剂量低对比剂浓度冠脉成像的可行性。方法:对临床拟诊为冠心病的200例患者随机分为A、B、C 3组,A组(120 kV,270 mgI/mL)65例,B组(120 kV,350 mgI/mL)65例,C组(100 kV,270 mgI/mL)70例。采用320排CT行前瞻性心电门控扫描,评价试验所得到的3组图像质量以及患者所接受的有效辐射剂量。结果:3组图像冠脉评分无显著差异(P>0.5),图像噪声和增强效果有明显差异(P<0.05),但都可满足诊断。患者所接受的辐射剂量有显著差异(P<0.01),C组是其中的最低值组。结论:低辐射剂量低对比剂浓度可用于CT冠脉成像,并且具有明显的优越性。
Objective:To evaluate the effect of low radiation dose and low concentration of contrast agent for coronary computed tomography (CCTA) with 320-detector CT.Methods:200 patients with clinically suspected coronary disease were randomly divided into A,B,C three groups.Group A(120 kV,270 mgI/mL) 65 cases,group B(120 kV,350 mgI/mL) 65 cases,group C (100 kV,270 mgI/mL) 70 cases,and performed CCTA with prospective ECG gating by using 320-detector CT.Image and the optimal radiation dose were evaluated.Results:Coronary images of three groups had no significant difference (P〉0.5).Image noise and enhancement was significantly different(P〈0.05),but it can be used for diagnosis.The radiation dose received by the patient had a significant difference (P〈0.01),and the dose received by group C is the lowest.Conclusion:The 320 multi-slice CT with low radiation dose and low concentration of contrast agent can be used for CCTA,and has obvious advantages.