Lenstar 900与A型超声和角膜曲率计在白内障患者眼生物学测量的比较 被引量:13

Comparison on eye biometry of Lenstar 900,A- scan ultrasound and keratometer in patients with cataract
下载PDF
导出
摘要 目的:比较Lenstar 900与A型超声和角膜曲率计在白内障患者眼生物学测量中测量眼轴长度(AL)、前房深度(ACD)及角膜曲率(K1,K2,Km)上的差异,并分析其一致性,为Lenstar 900临床应用提供参考。方法:选取接受白内障手术的患者36例50眼,晶状体核均在Ⅳ级以下,术前分别采用Lenstar 900、A超和角膜曲率计完成AL,ACD和K1,K2,Km的测量,比较测量结果的差异采用配对t检验,测量结果的相关性采用Pearson相关分析,测量结果的一致性采用Bland—Ahamn统计分析。结果:Lenstar 900与A型超声测量AL和ACD值之间的差异无统计学意义(P>0.05)。Lenstar 900与角膜曲率计测量Kl,K2和Km值的差异均无统计学意义(P>0.05)。Lenstar 900与A型超声和角膜曲率计测量的所有参数均具有密切的线性相关(r>0.9,P<0.05),Bland—Ahamn统计分析一致性良好。结论:Lenstar 900与A型超声和角膜曲率计对白内障患者术前生物学测量结果准备可靠,可以相互替代,但是Lenstar 900不仅可以一次完成对AL、ACD和角膜曲率的测量,而且同时可以获得角膜厚度、晶状体厚度、角膜缘白到白距离、瞳孔大小、光轴的偏心率及视网膜厚度等参数,具有测量过程非接触、方便快捷、节约时间等优点,具有广泛的应用前景。 AIM: To investigate the differences among Lenstar 900,A- scan ultrasound and keratometer in measurement of axial length( AL),anterior chamber depth( ACD) and corneal curvature( K1, K2, Km), and evaluate the consistency of the instruments, with the purpose providing references for the clinical application of Lenstar 900.METHODS: In this study we picked up 36 patients(50eyes) underwent cataract surgery, and lens nucleus hardness were under level IV. Before the operation,AL,ACD and K1,K2,Km were measured by Lenstar 900,A-scan ultrasound and keratometer respectively. The differences between the results were compared by the paired t- test. The correlation of the results was analyzed by Pearson correlation analysis,and the consistency was measured by Bland- Ahamn method.RESULTS: The mean AL and ACD values measured by Lenstar 900 and A- scan ultrasound had no significantly statistic differences(P〈0. 05). The K1,K2,Km measured by Lenstar 900 and keratometer were not significantly statistical different( P〈0. 05). The results measured by these three instruments had close linearity correlation(r 0. 9,P〈0. 01). The consistency of the results was well in Bland- Ahamn analysis. CONCLUSION: The preoperatively biometric result of Lenstar 900, A- scan ultrasound and keratometer inpatients with cataract are all reliable,and they can be substituted by each other. However,Lenstar 900 can not only measure AL,ACD and corneal curvature at the same time,but also cornal thickness,lens thickness,white to white, pupil size, optical axis eccentricity, retinal thickness and so on. It has a number of advantages such as non- touching,convenient and efficient,and can be recommended to use widely.
出处 《国际眼科杂志》 CAS 2014年第8期1440-1443,共4页 International Eye Science
关键词 LENSTAR 900 眼生物学测量 白内障 Lenstar 900 eye biometry cataract
  • 相关文献

参考文献13

  • 1Rabsilber TM,Jepsen C,Auffarth GU,et al.Intraocular lens power calculation:clinical comparison of 2 optical biometry devices.J Cataract Refract Surg 2010;36(2):230-234.
  • 2Eibschitz-Tsimhoni M,Tsimhoni O,Archer SM,et al.Effect of axial length and keratometry measurement error on intraocular lens implant power prediction formulas in pediatric patients.J AAPOS 2008;12(2):173-176.
  • 3Olsen T.Sources of error in intraocular lens power calculation.J Cataract Refract Surg 1992;18(2):125-129.
  • 4Bielos Roncevic M,Busic M,Cima I,et al.Intraobserver and interobserver repeatability of ocular components measurement in cataract eyes using a new optical low coherence reflectometer.Graefes Arch Clin Exp Ophthalmol 2011;249(1):83-87.
  • 5Mylonas G,Sacu S,Buehl W,et al.Performance of three biometry devices in patients with different grades of age-related cataract.Acta Ophthalmol 2011;89(3):237-241.
  • 6Rohrer K,Frueh BE,Walti R,et al.Comparison and evaluation of ocular biometry using a new noncontact optical low-coherence reflectometer.Ophthalmology 2009;116(11):2087-2092.
  • 7Jasvinder S,Khang TF,Sarinder KK,et al.Agreement analysis of Lenstar with other techniques of biometry.Eye 2011;25(6):717-724.
  • 8Holzer MP,Mamusa M,Auffarth GU.Accuracy of a new partial coherence interferometry analyser for biometric measurements.Br J Ophthalmol 2009;93(6):807-810.
  • 9Shammas HJ,Hoffer KJ.Repeatability and reproducibility of biometry and keratometry measurements using a noncontact optical low-coherence reflectometer and keratometer.Am J Ophthalmol 2012;153(1):55-61.
  • 10Tehrani M,Krummenauer F,Kumar R,et al.Comparison of biometric measurements using partial coherence interferometry and applanation ultrasound.J Cataract Refract Surg 2003;29(4):747-751.

同被引文献97

  • 1项奕,彭斌,邢怡桥.IOL Master联合A超对3,4级核晶状体进行人工晶状体度数测量[J].武汉大学学报(医学版),2004,25(5):574-576. 被引量:6
  • 2刘伟中.A型超声生物测量、人眼屈光成分的研究与近视眼形成发展规律的探讨(二)[J].中国眼镜科技杂志,2006(12):105-108. 被引量:1
  • 3祁媛媛,张丰菊.光学相干生物测量仪的临床应用[J].中国实用眼科杂志,2007,25(8):809-812. 被引量:9
  • 4Holzer MP, Mamusa M, Auffarth GU.Accuracy of a new partial co- herence interferometry analyser for biometric measurements[J]. Br J Ophthalmol,2009, 93(6):807-810.doi: 10.1136/bjo.2008.152736.
  • 5Zhao J, Chen Z, Zhou Z, et al. Evaluation of the repeatability of the Lenstar and comparison with two other non-contact biometric devic- es in myopes[J]. Clin Exp Optom, 2013, 96(1): 92-99.doi: 10.1111/ j. 1444-0938.2012.00793 .x.
  • 6Thibos LN, Wheeler W, Homer D. Power vectors:an application of Fourier analysis to the description and statistical analysis of re- fractive error[J]. Optom Vis Sci, 1997,74(5):367-375.
  • 7Ferraz FH, Corrente JE, Opromolla P, et al. Refractive errors in a Brazilian population: age and sex distribution[J]. Ophthahnic Physi- ol Opt,2015, 35(1):19-27.doi: 10.1111/opo. 12164.
  • 8Wolfram C, Htihn R, Kottler U, et al. Prevalence of refractive errors in the European adult population: the Gutenberg Health Study (GHS)[J]. Br J Ophthalmol,2014, 98(7):857-861.
  • 9Ziaei H, Katibeh M, Solaimanizad R, et al. Prevalence of refractive errors; the yazd eye study[J]. J Ophthalmic Vis Res,2013, 8(3):227-236.
  • 10Hashemi H, Asgari S, Miraftab M, et al. Agreement study of kerato- metric values measured by Biograph/LENSTAR, auto-kerato-re- fractometer and Pentacam: Decision for IOL calculation[J].Clin Exp Optom, 2014, 97(5):450-455.doi: 10.1111/cxo.12186.

引证文献13

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部