期刊文献+

利用射野照相验证片对胸部肿瘤放疗摆位误差的分析 被引量:1

下载PDF
导出
摘要 目的利用射野照相和Photoshop软件对胸部肿瘤适形放射治疗进行位置验证,分析摆位误差,为临床靶区(CTV)外扩计划靶区(PTV)提供参考依据。方法 30例胸部肿瘤患者,每位患者在首次及每周一治疗摆位前拍摄正、侧位射野照相各一张,与数字重建图像进行骨性匹配,测量射野中心点在X、Y、Z方向的位移值。结果通过对150张次射野片对比、分析,摆位误差X为(2.22±2.21)mm,Y为(3.69±2.47)mm,Z为(3.38±2.24)mm;获取CTV外扩的边界:X为5.99 mm,Y为9.11mm,Z为8.33 mm。结论通过对摆位误差数据的分析,为胸部肿瘤的适形放疗提供CTV外扩值,提高放射治疗的准确性。
机构地区 甘肃肿瘤医院
出处 《卫生职业教育》 2014年第15期133-135,共3页 HEALTH VOCATIONAL EDUCATION
  • 相关文献

参考文献9

  • 1Hunt M. The effect of positionail uncertainties on the treatment of primary nasopharyns cancer[J].Med Phys, 1989,16(3) :456-464.
  • 2Jef M M,Mary V G,Waller R B,et al.Prospective clinical ecalution ofelectronic portal imaging device [J].Int J Radiat Oncol Biol Phys,1996 (34) :943.
  • 3Stroom J C, De Boer H C, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability[J]. Int J Radiat Oncol Biol Phys, 1999,43(4):905-919.
  • 4Stroom J C,Heimen B J. Geometrical uncertainties, radiotherapy planning margins and the ICRU-62 report [J].Radiat Oncol, 2002,64( 1 ): 75-83.
  • 5于龙珍,翟振宇,沈君姝.体部肿瘤精确放疗摆位误差分析[J].中国医学物理学杂志,2007,24(3):160-162. 被引量:36
  • 6Humkmans C W,Remeijer P,Lebesque J V, et al. Set-up verification using portal imaging; reiew of current clinical practice [J].Radiother Oncol, 2001,58(2) : 105-120.
  • 7胡逸民 张红志 戴建荣.肿瘤放射物理学[M].北京:原子能出版社,1996.614.
  • 8Alter J M, Ten-Haken R K, Lawrence T S,et al. Uncertainties in CT based radiadion therapy treatment planning associate with patient breathing[J]. Int J Radiat Oncol Biol phys, 1996(36) : 167-174.
  • 9谢志原,林育毅,王永川,吴君心,陈辉林,邵林东,陈素真,潘建基.锥形束CT测量食管癌放射治疗的摆位误差[J].中国癌症防治杂志,2011,3(2):125-129. 被引量:9

二级参考文献14

  • 1戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 2包德强,夏玲,林春丽,董泽华,林凤飞.56例食管癌三维适形放疗的疗效观察[J].中华放射肿瘤学杂志,2006,15(5):382-383. 被引量:23
  • 3韩春,王澜,祝淑钗,迟子锋,曹彦坤.食管癌三维适形放疗中肺和食管损伤的相关因素分析[J].中华放射肿瘤学杂志,2007,16(2):103-107. 被引量:21
  • 4Uematsu M,Akira S,Kazuhiko T,et al..Focal,high dose,and fractionated modified stereotactic radiationtherapy for lung carcinoma patients[J].Cancer,1998,82:1062-1070.
  • 5MeertensH,BijholdJ,StrackeeJ.A method of the measurement of field placement errors in digital portal images[J].Phys Med Biol,1990,35:299-323.
  • 6Hurkmans C W,Remeijer P,Lebesque J V,et al.Set-up verification using portal imaging; review of current clinical practice[J].Radiather Oncol,2001,58(2):105-120.
  • 7Hans CJ,John R.Analysis and reduction of 3D systematic and random setup deviations during the simulation and treatment of lung cancer patients with CT based external beam radiotherapy dose planning[J].Int Radiat Oncol Biol Phys,2001,49:857-868.
  • 8Lax I,Blomgren H,Naslund I,et al.S tereotactic radiotherapy of malignacies in the abdomen[J].Acta Oncologia,1994,33:677-703.
  • 9Onishi H,Kuriyama K,Komiyama T,et al.CT evaluation of patient deep inspiration self-breath-holding:how precisely can patients reproduce the tumor position in the absence of respiratory monitoring devices?Med Phys[J].2003,30(6):1183-1187.
  • 10Balter JM,Ten-Haken RK,Lawrence TS,et al.Uncertainties in CT-based radiation therapy treatment planning associated with patient breathing[J].Int Radiat Oncol Biol Phys,1996,36:167-174.