基于噪声功率谱的CT图像噪声评价 被引量:3

Noise power spectrum-based evaluation of CT image noise
下载PDF
导出
摘要 目的设计并探讨基于噪声功率谱(NPS)的CT图像噪声评价方法。方法改变CT图像扫描参数以及重建参数获取图像,采用NPS与传统标准差(SD)两种方法对该图像进行噪声检测,并对比两种检测结果。结果 NPS曲线幅值与层厚和剂量成反比,与重建算法的增强程度成正比。重建算法、层厚和剂量对噪声指数的影响能够在NPS曲线上得到明显的体现。结论 NPS能够从幅值与频率方面反应噪声变化,与SD方法相结合可形成更为完善的噪声评价体系。 Objective To design and discuss the method for evaluating CT image noise based on noise power spectrum(NPS).Methods CT images were acquired with different parameters.The NPS and standard deviation(SD)were computed for evaluating the image noise,and then two results were compared.Results NPS peak value was negatively related with slice and dose,and positively related with reconstruction algorithm.The factors variation,such as reconstruction algorithm,slice and dose,which can influence the CT image noise level,was reflected in the NPS curves obviously.Conclusion The NPS can reflect the magnitude and frequency content of the noise variations.It will build a relatively perfect evaluation method for noise evaluating system by combining NPS with SD.
出处 《中国医学影像技术》 CSCD 北大核心 2014年第8期1243-1246,共4页 Chinese Journal of Medical Imaging Technology
基金 广州市行业工程中心建设项目(穗科信条[2010]13-1)
关键词 噪声功率谱 标准差 噪声评价 图像分析 Noise power spectrum Standard deviation Noise evaluation Image analysis
  • 相关文献

参考文献14

  • 1余晓锷,龚剑.CT原理与技术.北京:科学出版社,2014:149-153.
  • 2Ikeda M, Makino R, Imai K, et al. A method for estimating noise variance of CT image. Comput Med Imaging Graph, 2010, 34(8) : 642-650.
  • 3白玫,刘彬,郑钧正.多排螺旋CT纵轴分辨力和图像噪声影响因素研究[J].中国医学影像技术,2008,24(7):1110-1113. 被引量:13
  • 4Siewerdsen JH, Cunningham IA, Jaffray DA. A framework for noise-power spectrum analysis of multidimensional images. Med Phys, 2002,29(11) :2655-2671.
  • 5Cupp Scott L. A practical application for noise power spectrum analysis. Philadelphia: Drexel University, 2008.
  • 6姜树勋.CT噪声的影响因素分析[J].中国医学装备,2007,4(8):36-37. 被引量:7
  • 7亓恒涛,秦维昌,王巍,刘传亚.64层螺旋CT噪声测试及其影响因素分析[J].中华放射医学与防护杂志,2007,27(2):195-198. 被引量:17
  • 8Mieville FA, Bolard G, Bulling S, et al. Effects of computing pa- rameters and measurement locations on the estimation of 3D NPS in non-stationary MDCT images. Phys Med, 2013, 29 (6) : 684- 694.
  • 9Wilson JM, Christianson OI, Richard S, et al. A methodology for image quality evaluation of advanced CT systems. Med Phys, 2013,40(3) :031908.
  • 10张彦彩,朱小忠,马国林,王武.螺旋CT低剂量扫描研究进展[J].中国医学影像技术,2010,26(7):1376-1378. 被引量:71

二级参考文献55

共引文献102

同被引文献26

引证文献3

二级引证文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部