摘要
切光马达使用寿命到期之后就会停转,从而对仪表的测量造成误差,失真的仪表示值又会对工艺人员的操作造成直接的影响,进而影响工艺产品的产品质量。本文通过对这种红外分析仪最常见的且容易发生的故障进行分析,并写出如何更换马达,以及更换后使用效果很好地恢复这一示例,给出了解决马达失效停转问题的正确方法。
In the production process of chemical,petrochemical,real- time on the sample composition analysis was a necessary means of process production,only with scientific and accurate data can achieve economic production,bring considerable economic benefits for the enterprise. The instrument analysis online was a necessary choice,which included infrared analyzer. The most common fault of infrared analyzer and prone to,namely the light cutting motor damage,the method of fault phenomena and treatment were described,so as to offer some reference to the daily maintenance of infrared analyzer.
出处
《广州化工》
CAS
2014年第17期157-159,共3页
GuangZhou Chemical Industry
关键词
红外线
切光马达
光源
切光片
检测器
电磁波
infrared
cut light motor
light source
cut-ray
detector
electromagnetic