基于图像的全自动尿液有形成分分析仪US-2026诊断性能评价 被引量:3

Automated urine particle analyzer US 2026 in detecting different kinds of cells
下载PDF
导出
摘要 目的评价全自动US-2026尿液有形成分分析仪诊断性能。方法收集本院2013年3月1日-4月1日我院门诊和住院患者423份尿液样本,其中350份用于细胞诊断评估,73份用于管型诊断评估。结果研究表明US-2026尿液有形成分分析仪对红细胞、白细胞的检出灵敏度和特异性较高,均>80%;与人工镜检相比符合率分别为红细胞84.5%,白细胞82.3%,管型60.6%,上皮细胞78.8%。结论全自动US-2026尿液有形成分分析仪与人工显微镜检测结果有较好的一致性,可减少尿液人工镜检率,缩短标本检测周期。 Objective To assess the diagnostic performance of automated urine particle analyzer US-2026. Methods Of the 423 urine samples collected from patients who visited the outpatient department or admitted to our hospital from March 1,2013 to April 1, 2013, 350 were used in cellular diagnosis using US-2026 and 73 were used in cast cell diagnosis using US-2026. Results The sensitivity and specificity of US-2026 were higher than 80% for detecting red blood cells and white blood cells. The rate of red blood cells, white blood cells, cast and epithelial cells that detected using US-2026 was 84.5%, 82.3%, 60.6%, and 78.8%, respectively, which was consistent with that detected by manual microscopy. Conclusion The consistency of US 2026 and manual microscopy is rather good in detecting different types of cells. US-2026 can shorten the detection cycle of urine samples.
出处 《解放军医学院学报》 CAS 2014年第9期893-895,共3页 Academic Journal of Chinese PLA Medical School
关键词 尿液有形成分分析仪 显微镜 临床实验室技术 US-2026 microscope clinical laboratory technique
  • 相关文献

参考文献17

  • 1European Confederation of Laboratory Medicine.European urinalysis guidelines:summary[J].Scand J Clin Lab Invest,2000,60:1-96.
  • 2Mundt LA,Shanahan K.Textbook of urinalysis and body fluids[M].2rid ed.Philadelphia:Philadelphia:Lippincott Williams & Wilkins,2011:55-192.
  • 3Kurup R,Leich M.Comparison of urine analysis using manual and sedimentation methods[J].West Indian Med J,2012,61 (3):240-244.
  • 4Yü ksel H,Kili(c) E,Ekinci A,et al.Comparison of fully automated urine sediment analyzers H800-FUS100 and labumat-urised with manual microscopy[J].J Clin Lab Anal,2013,27 (4):312-316.
  • 5丛玉隆.尿液沉渣检查标准化的建议[J].中华检验医学杂志,2002,25(4):249-250. 被引量:246
  • 6丛玉隆,马骏龙,张时民,岳秀玲,张清,丁建文.尿液细胞成分定量分析方法学研究[J].中华检验医学杂志,2006,29(3):211-214. 被引量:130
  • 7Delanghe J.New screening diagnostic techniques in urinalysis[J].Acta Clin Belg,2007,62 (3):155-161.
  • 8Xiang D,Cong Y,Wang C,et al.Development of microscopic review criteria by comparison urine flow cytometer,strip and manual microscopic examination[J].Clin Lab,2012,58 (9-10):979-985.
  • 9Zhao F,Jin Y,Chen X,et al.Clinical application of UF-1000i in combination with AX-4280 for the screening test ability of urinary formed elements[J].J Clin Pathol,2013,66 (3):229-231.
  • 10Shayanfar N,Tobler U,yon Eckardstein A,et al.Automated urinalysis:first experiences and a comparison between the Iris iQ200 urine microscopy system,the Sysmex UF-100 flow cytometer and manual microscopic particle counting[J].Clin Chem Lab Med,2007,45 (9):1251-1256.