加速器机载kV-CBCT成像参数与图像质量的定量分析 被引量:2

Quantitative analysis of the impact of preset imaging parameters on image quality of Linac-integrated kV-CBCT unit
下载PDF
导出
摘要 采用Catphan 503模体,选用不同的扫描预设条件成像,对加速器机载千伏级锥形束CT(kV-CBCT)图像的均匀性、噪声、低对比度分辨率和空间分辨率进行评估。结果表明提高X线球管管电压管电流、增大机架旋转扫描范围、减小准直器尺寸、提高重建精度等参数有利于获得高质量锥形束CT(CBCT)图像,但同时患者将多受额外的X射线辐射,也加大X线球管和探测器等元器件的损耗。因此,临床使用中应根据患者解剖条件合理选择成像参数,这样既能保证图像质量,又尽可能降低患者的辐射剂量。 Abstract This study scanned the Catphan 503 phantoms to evaluate the uniformity, noise, low contrast resolutionand spatial resolution of Linac-integrated kV-cone beam CT image with choosing different scanning preset parame- ters, and analyzed the image variations changed with the tube voltage, tube current, collimator type, gantry scan- ning range and reconstruction precision. The results showed that improving X-ray tube voltage and current, expen- ding the scanning gantry rotation range and reducing collimator size, increasing reconstruction precision and other imaging parameters were conducive to high quality cone beam CT images. As a result, patients suffered more addi- tional X-ray radiation and X-ray tube and detector components loss increased, simultaneously Imaging preset parameters should be reasonably chosen according to the patient' s anatomy, which can guarantee quality and minimize patient radiation dose. scanning the image
出处 《安徽医科大学学报》 CAS 北大核心 2014年第9期1336-1339,共4页 Acta Universitatis Medicinalis Anhui
基金 高校省级自然科学研究项目(编号:KJ2010B380)
关键词 千伏级锥形束CT X线容积成像系统 均匀性 噪声 低对比度分辨率 空间分辨率 kV cone beam computed tomography X-ray volume imaging uniformity noise low contrast visibility spatial resolution
  • 相关文献

参考文献11

二级参考文献44

  • 1戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 2邵军明,徐晓东,孔军.CT成像质量影响因素综述[J].CT理论与应用研究(中英文),2006,15(3):61-67. 被引量:18
  • 3尹勇,袁双虎,刘同海,卢洁,于金明.加速器附加锥形束CT图像质量评价[J].中华放射肿瘤学杂志,2007,16(3):225-228. 被引量:25
  • 4Jaffray DA,Siewerdsen JH.Cone beam computed tomography with a flat-panel imager:initial performance characterization.Med Phys,2000,27(6):1311-1323.
  • 5McBain CA,Henry AM,Sykes J,et al.X-ray volumetric imaging in image-guided radiotherapy:the new standard in on-treatment imaging.Int J Radiat Oncol Biol Phys,2006,64(2):625-634.
  • 6Ding GX,Duggan DM.Coffey CW.Characteristics of kilovoltage X-ray beams used for cone-beam computed tomography in radiation therapy.Phys Med Biol,2007,52(6):1595-1615.
  • 7Palomo JM,Subramanyan K,Hans M.Influence of mA settings and a copper filter in CBCT image resolution.Int J Comput Assist Radiol Surg,2006,1(9):391-393.
  • 8Saw CB, Yang Y, Li F, et al. Performance characteristic and quality assurance aspects of kilovoltage cone-beam CT on medical linear accelerator[J]. Med Dosim, 2007, 32(2): 80-85.
  • 9Oelfke U, Tucking T, Nill S, et al. Linac-integrated kv-cone beam CT: Technical features and first applications[J]. Med Dosim, 2006, 31(1):52-70.
  • 10Moores BM. Patient dose measurements in radiological practices[J]. Phys Med Biol, 2006, 51 : L31-L34.

共引文献36

同被引文献8

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部