期刊文献+

基于STM32的便携式压力测痛仪的设计 被引量:1

Design of Portable Pressure-pained-detector Based on STM32
下载PDF
导出
摘要 针对现有疼痛检测设备体积笨重,不便于携带的局限性,设计了一种基于STM32的便携式高精度压力测痛仪。用高精度的压力传感器为主体的测压触头拾取压力痛信号。通过对放大电路和滤波电路等信号预处理,有效地滤除系统噪声。TFT-LCD实时显示被测的压力值,有效建立被测对象的压痛阈值。用USB接口将所得数据传输到PC机上存储,便于后续医生对疾病的诊断与治疗。 In accordance with the bulky volume and not easy to carry of pain detection equipment limitations, designs a kind of portable high-precision dolonmeter based on STM32. Using the high precision pressure sensor pressure contact as the main body to make the pressure pain signals. Through the signal preprocess of the amplifying circuit and the filter circuit, to eliminate the system noise effectively. Use the TFT-LCD to display the real-time measured pressure value, and establish a pressure pain threshold of the object. The USB interface transmits the data to PC to store, and it's beneficial to the diagnosis and treatment of subsequent.
出处 《中国医疗器械杂志》 CAS 2014年第5期333-336,共4页 Chinese Journal of Medical Instrumentation
基金 国家自然科学基金项目(61271115)
关键词 测痛仪 压痛阈值 便携式 STM32 etector, PPT, portable, STM32
  • 相关文献

参考文献9

二级参考文献74

  • 1赵均海,孙家驹.小信号医用传感器的设计及应用[J].生物医学工程学杂志,2004,21(6):1003-1005. 被引量:2
  • 2代少升,张跃,张和君,刘文煌.便携式远程心电实时监护仪的研制[J].医疗卫生装备,2006,27(9):1-2. 被引量:31
  • 3张奋耿,梁向坚,张继平,安启升,张有剑.弹簧式压力测痛仪在肱骨外上髁炎诊断中的应用[J].中国中西医结合外科杂志,2006,12(6):557-559. 被引量:2
  • 4Winter B B,Webster J G.Driven-right-leg Circuit De-sign.IEEE T rans.Biomed.Eng.1983,BME-30:62-66.
  • 5Cristi,R.著,徐盛等译.现代数字信号处理[M].北京:机械工业出版社,2005.129-148.
  • 6Sjolund BH,Persson AL.Pressure pain threshold changes after repeated mechano-nociceptive stimulation of the trapezius muscle:possible influence of previous pain experience[J].J Pain,2007,8(4):355-362.
  • 7Jones DH,Kilgour RD,Comtois AS.Test-retest reliability of pressure pain threshold measurements of the upper limb and torso in young healthy women[J].J Pain,2007,8(8):650-656.
  • 8Ogimoto T,Ogawa T,Sumiyoshi K,et al.Pressure-pain threshold determination in the oral mucosa:validity and reliability[J].J Oral Rehabil,2002,29(7):620-626.
  • 9Bernhardt O,Schiffman EL,Look JO.Reliability and validity of a new fingertip-shaped pressure algometer for assessing pressure pain thresholds in the temporomandibular joint and masticatory muscles[J].J Orofac Pain,2007,21(1):29-38.
  • 10Dagtekin O,KOnig E0 Gerbershagen H J,et al.Measuring pressure pain thresholds.Comparison of an electromeehanically controlled algometer with established methods[J].Scbmerz,2007,21(5):439-444.