摘要
目的评估在独立通风笼盒(IVC)所处环境不同湿度条件下,IVC不同换气次数时,大、小鼠IVC微环境湿度和氨浓度的变化。方法屏障环境内,以7 d为换笼周期,分别设定室内环境低湿(40%)、中湿(50%)、高湿(60%),IVC换气次数为40次/h、60次/h时,对3个品牌大、小鼠IVC微环境进行测定,观察笼盒内湿度和氨浓度的变化。结果当室内环境为低湿条件,小鼠和大鼠IVC换气次数设置为40次/h时;当室内环境为中湿条件,小鼠换气次数设置为40次/h、大鼠换气次数设置为60次/h时;当室内环境为高湿条件,小鼠换气次数设置为60次/h时,笼盒内微环境均基本能够满足GB14925-2010对湿度和氨浓度的要求。而在高湿情况下,即使大鼠IVC换气次数置为60次/h也不能满足要求。结论室内环境湿度和IVC换气次数是影响IVC微环境的关键指标,只有依据室内环境条件合理设置IVC换气次数才能较好地维护IVC微环境和达到有效管理的目的。
Objective To assess the changes of humidity and ammonia concentration in rat and mouse individually ventilated cages (IVC) based on macroenvironmental humidity and air ventilation changes .Methods Three kinds of rat and mouse IVC in barrier facilities were set as research objective .The changes of micronvironmental humidity and ammonia concentration at 40 times/h and 60 times /h air changes were detected continuously for a 7-days-cycle relative to low (40%), moderate (50%), and high (60%) macroenvironmental humidity.Results Mouse and rat IVC with 40 times /h air changes under low macroenvironmental humidity condition , mouse IVC with 40 times/h and rat IVC with 60 times/h air changes under moderate macroenvironmental humidity condition , mouse IVC with 60 times /h air changes under high macroenvironmental humidity condition , basically meet the GB14925-2010 requirements.While under macroenvironmental high humidity condition, the microenvironments of rat and mouse IVC with 60 times/h air changes could not satisfy the requirements.Conclusions The environmental humidity and ventilation frequency are the key index of IVC microenvironment.Only on the basis of external environment conditions to set up reasonable IVC ventilation frequency in order to better maintain the IVC microenvironment so that to achieve the goal of effective management .
出处
《中国比较医学杂志》
CAS
2014年第8期36-40,共5页
Chinese Journal of Comparative Medicine
基金
广东省科技计划项目,项目编号[2011B060300034]
关键词
IVC
外环境
微环境
换气次数
湿度
氨浓度
Individually ventilated cage(IVC)
Macroenvironment
Microenvironment
Air Changes
Humidity
Ammonia concentration
Rat
Mouse