医科达Synergy加速器ⅩⅥ系统CBCT扫描头颈部模体的误差研究 被引量:2

The study of error in the head and neck phantom using the X-ray volumetric image system of Elekta Synergy accelerator with CBCT
原文传递
导出
摘要 目的 探讨医科达Synergy加速器ⅩⅥ系统图像自动匹配结果的准确性.方法 应用仿真模体ART头部进行仿真模拟计划设计后进行模体CBCT扫描,测量在不同匹配区域、不同匹配方法以及在不同方向上定量移床后扫描的匹配误差,获取数据进行分析.结果 在定量移床后模体扫描匹配误差与相应移床量数据基本吻合,x、y、z轴向总误差分别为(0.11±0.41)、(-0.04±2.6)、(0.28±0.74) mm.在不同匹配区域、不同匹配方法下进行自动匹配,各组数据差异较小.但对不同匹配区域、不同匹配方法上存在一些配对差异.结论 ⅩⅥ系统的自动匹配功能在头颈部应用上匹配精确,对偏移误差辨识准确可信度高.但对不同匹配区域、不同匹配方法上存在一定优劣对比,在实际运用中可根据实际情况择优应用. Objective To investigate the accuracy and reliability of the image automatic matching using X-ray volumetric image (ⅩⅥ) system with the accelerator Elekta Synergy and to compare the matches with different methods and area.Methods A simulation plan using the head and neck phantom was designed and sent to ⅩⅥ system for CBCT scanning.During the scanning,the couch for certain distance was moved and the data of ⅩⅥ image automatic Matching in different matching methods and area was collected.Results The result of ⅩⅥ image automatic matching is consistent and correct to direction and distance which the couch had been moved (x:(0.11 ± 0.41) mm,y:(-0.04± 2.6) mm,z:(0.28 ± 0.74 mm)).There are not significant differences between the results that matching in different matching methods and different matching area.Conclusions The automatic image Matching of ⅩⅥ is accuracy and high reliability in recognition of offset error.But there are some significant differences on the automatic image Matching in different matching methods and different matching area.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2014年第5期433-436,共4页 Chinese Journal of Radiation Oncology
关键词 体层摄影术 X线计算机 锥形束 自动匹配 头颈部肿瘤 Tomography,X-ray computed,cone beam Automatically match Head and neck neoplasms
  • 相关文献

参考文献13

  • 1Uematsu M,Akira S,Kazuhiko T,et al.Focal,high dose,and fractionate d modified stereotactic radiotherapy for lung carcinoma patients[J].Cancer,1998,82:1062-1070.
  • 2Teh BS,Bastasch MD,Wheeler TM,et al.IMRT for prostate cancer:defining target volume based on correlated pathologic volume of disease[J].Int J Radiat Oncol Biol Phys,2003,56:184-191.
  • 3张爱华,胡健.OBI系统CBCT引导下的腹部调强患者位置验证的应用分析[J].医疗卫生装备,2010,31(11):82-83. 被引量:4
  • 4Jaffray DA,Siewerdsen JH,Wong JW,et al.Flat-panel conebeam computed tomography for image radiation therapy[J].Int J Radiat Oncol Biol Phys,2002,53:1337-1349.
  • 5蒋琳,陈晓品,金丹,宋延波,刘丽.应用OBI系统分析胸部肿瘤调强放疗的摆位误差[J].激光杂志,2009,30(5):92-93. 被引量:10
  • 6Houweling AC,van der Meer S,van der Wal E,et al.Improved immobilization using an individual head support in head and neck cancer patients[J].Radiother Oncol,2010,96:100-103.
  • 7马广栋,李承军,付敬国,吴冰.不同解剖区域配准方法对鼻咽癌体位误差的影响[J].医疗卫生装备,2011,32(7):63-64. 被引量:9
  • 8Lee N,Puri DR,Blanco AI,et al.Intensity-modulated radiation therapy in head and neck cancers:an update[J].Head Neck,2007,9:387-400.
  • 9Eisbruch A.Intensity modulated radiotherapy of head and neck cancer:encouraging early results[J].Int J Radiat Oncol Biol Phys,2002,53:1-3.
  • 10Benner S,Roentgen B.Television monitoring of a 30 MV X-ray beam[J].Phys Med Bid,1962,7:29-33.

二级参考文献47

  • 1沈捷,张福泉,邱杰,胡克,张明杰.应用EPID对放射治疗摆位误差的研究[J].中华放射医学与防护杂志,2006,26(3):308-308. 被引量:39
  • 2邬筱莹,章真,王文超,任珺,郭晓梅,陆惠忠.电子射野影像装置对盆腔肿瘤放疗摆位误差的测定[J].中华放射肿瘤学杂志,2007,16(1):52-54. 被引量:26
  • 3Uematsu M, Akira S, Kazuhiko T, et al. Focal, high dose, and fractionated modified stereotactic radiationtherapy for lung carcinoma patients[J]. Cancer, 1998,82:1062-1070.
  • 4Bijhold J, van Herk M, Vijlbriet R, et al. Fast evaluation ol patient setup during radiotherapy by aligning features in portal and simulator images[J] .Phys Med Biol, 1991, 36(12): 1665- 1679.
  • 5van Herk M. Errors and margins in radiotherapy [J]. Semin Radiat Oneol, 2004,14:52 - 64 .
  • 6van Herk M, Remeijer P, Lebesque JV. Inclusion ot geometric us - certainties in treatment plan evaluation [J]. Int J Radiat Oncol Biol Phys, 2002, 52:1407-1422.
  • 7Jin JY, Ajlouni M, Chen Q, et al. A technique of using gated - CT images to determine internal target volume (ITv)for fractionated stereotactic lung radiotherapy [ J ]. Radiother Oncoi, 2006, 78 : 177 - 184.
  • 8Yah D, Lockman D, Martinez A, et al. Computed tomography guided management of interfractional patient vafiation[J]. Semin Radiat Oncol, 2005,15:168 - 179.
  • 9Hurkmans CW, Remeijer P, Lebesque JV, et al. Set - up verification using portal images, review of current clinical practice [ J]. Radiotherapy and Oncology,2001, 58(2) : 105 - 120.
  • 10Mackie TR, Kapatoes J, Ruchala K, et al. Image guidance for precise confonnal radiotherapy[ J]. Int J Radiat Oncol Biol Phys, 2003, 56:89 - 105.

共引文献30

同被引文献36

  • 1崔明明,闫镔,曹鸿涛,陈健,曾磊.基于全局二值特征描述子的三维CBCT图像快速匹配算法[J].计算机辅助设计与图形学学报,2015,27(4):666-673. 被引量:3
  • 2戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 3Dzierma Y,Beyhs M,Palm J,et al.Set-up errors and planning margins in planar and CBCT image-guided radiotherapy using three different imaging systems:A clinical study for prostate and headand-neck cancer[J].Phys Med,2015,31(8):1055-1059.
  • 4Zhang Y,Wu H,Chen Z,et al.Concomitant imaging dose and cancer risk in image guided thoracic radiation therapy[J].Int J Radiat Oncol Biol Phys,2015,93(3):523-531.
  • 5Ciardo D,Alterio D,Jereczek-Fossa BA,et al.Set-up errors in head and neck cancer patients treated with intensity modulated radiation therapy:Quantitative comparison between three-dimensional cone-beam CT and two-dimensional kilovoltage images[J].Phys Med,2015,31(8):1015-1021.
  • 6Xin S.Setup errors in cone-beam computed tomography and their effects on acute radiation toxicity in cervical cancerradiotherapy[J].Genet Mol Res,2015,14(3):10937-10943.
  • 7Heijkoop ST,Langerak TR,Quint S,et al.Quantification of intrafraction changes during radiotherapy of cervical cancer assessed with pre-and post-fraction Cone Beam CT scans[J].Radiother Oncol,2015,117(3):536-541.
  • 8Huang Y,Gardner SJ,Wen N,et al.Radiobiologically optimized couch shift:A new localization paradigm using cone-beam CT for prostate radiotherapy[J].Med Phys,2015,42(10):6028-6032.
  • 9Wood TJ,Moore CS,Horsfield CJ,et al.Accounting for patient size in the optimization of dose and image quality of pelvis cone beam CT protocols on the Varian OBI system[J].Br J Radiol,2015,88(1055):20150364.
  • 10Dolezel M,Slezak P,Odrazka K,et al.Interfraction variation in prostate cancer-analysis of 11726cone-beam CT[J].J BUON,2015,20(4):1081-1087.

引证文献2

二级引证文献26

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部