通过剂量验证检测MLC运动的准确性 被引量:5

Accuracy detection of multileaf collimator movement by dose verification
下载PDF
导出
摘要 目的:通过剂量学验证检测加速器多叶准直器(multi-leaf collimator,MLC)运动的准确性。方法:使用MatriXX测量Open野与MLC动态射野的剂量分布,验证5例鼻咽癌动态调强放疗患者在MLC校准前后的剂量分布。结果:(1)Open野的中心点剂量与TPS一致,3%/3 mmγ通过率为98.91%;(2)MLC动态射野的中心点剂量测量值和TPS的差异校准前和校准后分别为3.23%和-0.36%,校准前和校准后3%/3 mmγ通过率分别为95.89%和98.94%;(3)5例鼻咽癌患者计划校准前后3%/3 mmγ通过率分别为(90.234±5.485)%和(99.528±0.321)%。结论:动态调强放疗中MLC的运动精度直接影响靶区内的剂量准确性,通过剂量学验证可以检测MLC运动的准确性。 Objective To detect the accuracy of the muhileaf collimator (MLC) by dose verification. Methods MatriXX was used to measure the dose distribution of Open field and dynamic MLC field. The dose distribution of 5 patients with nasopharyngeal carcinoma (NPC) treated with dynamic IMRT was verified before and after MLC calibration. Results The absolute dose of centre points of Open field were consistent with TPS, and 3%/3 mm γ passing rate was 98.91%. The abso- lute dose of centre points of dynamic MLC field was different with TPS by 3.23% before calibration while -0.36% after calibration, and 3%/3 mm γ passing rate was 95.89% before calibration while 98.94% after calibration. The 3%/3mm γ passing rate of the five patients was (90.234±5.485)% and (99.528±0.321)% respectively before and after calibration. Conclusion The motion accuracy of the MLC directly influences the dose accuracy of the target volume, and the dose veri-fication can test the motion accuracy of MLC.
出处 《医疗卫生装备》 CAS 2014年第9期97-99,共3页 Chinese Medical Equipment Journal
关键词 MatriXX 多叶准直器 动态调强放疗 剂量验证 MatriXX multileaf collimator dynamic intensity modulated radiation therapy dose verification
  • 相关文献

参考文献7

二级参考文献80

共引文献81

同被引文献45

  • 1肖锋,孙朝阳,石梅,张学成,陆宙,陆军,刘晓莉.鼻咽癌动态调强放疗计划的剂量学验证[J].现代肿瘤医学,2007,15(2):178-180. 被引量:17
  • 2胡杰,陶建民,孙光荣.医用直线加速器等中心的质量控制和质量保证[J].中国医疗器械杂志,2007,31(3):213-215. 被引量:24
  • 3Poppe B, Blechschmidt A, Djouguela A ,et al. Two dimensional ion- ization chamber arrays for IMRT plan verification[J]. Med Phys, 2006, 33(4) : 1 005-1 015.
  • 4LI J G, YAN G, LIU C. Comparison of two commercial detector arrays for IMRT quality assurance [J]. J Appl Clin Med Phys, 2009,10 (2) : 2 942.
  • 5ZHANG Yu-hai, LI Yue-min, XIA Huo-sheng. Impact of dose rates on the position accuracy of multi-leaf collimator[J]. Radiat Phys C hem, 2012,81(12) : 1 813-1 816.
  • 6Angelo F M, Chiara B, Stefania G. Gantry angle dependence in IMRT pre-treatment patient-specific quality controls[J]. Phys Medica, 2013,29 ( 2 ) : 204-207.
  • 7Anup K B,Suresh C S, Bhupendra R,et ol. Study of 2D ion chamber array for angular response and QA of dynamic MLC and pretreatment IMRT plans[J]. Rep Prac Oncol Radiother, 2009,14( 3 ) : 89-94.
  • 8BA! S, LI G J, WANG M J,et al. Effect of MLC leaf position, collima- tor rotation angle, and gantry rotation angle errors on intensity-modu- lated radiotherapy plans for nasopharyngeal carcinoma[J]. Med Dosi, 2013,38(2) : 143-147.
  • 9Ghasroddashti E, Smith W L, Quirk S, et al. Clinical consequences of changing the sliding window IMRT dose rate[J]. J Appl Clin Med Phys, 2012, 13(4):4-12.
  • 10Krzysztof S,Aleksandra G,Wojciech O. Beam rate influence on dose distribution and fluence map in IMRT dynamic technique[J]. Rep Prac Oncol Radiother, 2012,17 : 97-103.

引证文献5

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部