ArcCheck系统在鼻咽癌容积旋转调强剂量验证中的应用 被引量:13

Dosimetric Verification of Volumetric Modulated Arc Therapy in Nasopharyngeal Carcinoma using ArcCheck System
下载PDF
导出
摘要 目的:研究ArcCheck验证系统在鼻咽癌容积旋转调强剂量验证中的应用。在病人接受放射治疗之前,进行容积调强计划的验证,使其满足计划设计要求。保证患者实际照射中接受到准确的剂量,确保治疗的安全。方法:选择34例鼻咽癌病例用热塑型体部网罩固定,CT扫描定位,将计划CT传到Pinnacle计划系统(9.2版本)中进行容积旋转调强治疗计划设计,然后将设计好的计划传输至MOSAIQ网络系统上。同时把治疗计划导入ArcCheck模体中重新新型剂量计算。在验证之前,对Arccheck系统进行本底校准,矩阵校准,绝对剂量校准。分析方法采用的是Gamma方法在ArcCHECK自带软件中对治疗计划系统中计算的结果和实际测量的结果二者的差异,阈值标准选择3%,3mm,10%。采用微软Excel2003数据统计软件。结果:Gamma分析方法相对量通过率97.9%,绝对量通过率为95.9%;实测剂量分布图与理论计算剂量分布图在高低剂量点分布上重合很好;另外X方向和Y方向的平坦度,实测结果与计算结果很接近。结论:ArcCheck验证系统在鼻咽癌容积旋转剂量验证中是可行的。它是一种方便快捷准确的旋转调强剂量验证方法。 Objective:To investigate the dosimetric performance of ArcCHECK system,to validate the designed plan before patients radiotherapy,to make sure patient gain accuracy dosimetry.Methods:34 VMAT treatment plans of nasopharyngeal carcinoma patients were performed with Pinnacle treatment planning system(TPS),and then these plans were exported to MOSAIQ system,treatment plans were transplanted into ArcCHECK phantom and recomputed.Calibrated Arccheck system,matrix and absolute dose Before validation.Usingbuild-in the ArcCHECK software to analyze the result.Comparison of Pinnacle calculated dose versus the real measured result by Gamma methods,which threshold were 3mm,3%,10%,statistics software is Microsoft Excel 2003.Results:The relative dosage pass rate was 97.9%,and the absolute dosage pass rate was 95.9%.The measured and calculated dose distribution and the profile of X and Y direction is also coincided with the actual measured and calculated dose distribution.Conclusions:The Arc CHCK system was an ideal tool for the VMATverification.
出处 《中国医学物理学杂志》 CSCD 2014年第5期5136-5138,5168,共4页 Chinese Journal of Medical Physics
关键词 ArcCheck系统 容积旋转调强 鼻咽癌 Gamma方法 ArcCheck verification volumetric modulated arc therapy nasopharyngeal cancer γ method
  • 相关文献

参考文献7

二级参考文献29

  • 1耿辉,戴建荣,李晔雄,张可,王绿化.一种简单调强放疗技术应用的初步研究[J].中华放射肿瘤学杂志,2006,15(5):411-415. 被引量:58
  • 2Otto K. Volumetric modulated arc therapy:IMRT in a single gantry arc[J].Medical Physics,2008,(01):310-317.doi:10.1118/1.2818738.
  • 3Bzdusek K,Friberger H,Eriksson K. Development and evaluation of an efficient approach to volumetric arc therapy planning[J].Mad Phys,2009,(06):2328-2339.
  • 4Bhagwat MS,Han ZH,Ng SK. An oscillating sweeping gap test for VMAT quality assurance[J].Phya Med Biol,2010,(17):5029-5044.
  • 5Alvarez-Moret J,Pohl F,Koelbl O. Evaluation of volumetric medulated arc therapy (VMAT) with Oncentra MasterPlan(⑧) for the treatment of head and neck cancer[J].Radiation Oncology,2010,(05):110-120.
  • 6Ran M,Yang WS,Chen F. Comparison of Elekta VMAT with helical tomotherapy and fixed field IMRT:plan quality,delivery efficiency and accuracy[J].Medical Physics,2010,(03):1350-1359.
  • 7Bortfeld T,Webb S. Single-arc IMRT[J].Physics in Medicine and Biology,2009,(01):N9-N20.
  • 8Yon G,Lu B,Kozelka J. Calibration of a novel four-dimensional diode array[J].Medical Physics,2010,(01):108-115.doi:10.1118/1.3266769.
  • 9Létourneau D,Publieover J,Kozelka J. Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy[J].Medical Physics,2009,(05):1813-1821.
  • 10Yoda K,Nakagawa K,Shiraishi K. Dose verification of intensity modulated arc therapy using an ERGO + + treatment planning system and Elekta internal multileaf collimators for prostate cancer treatment[J].British Journal of Radiology,2009,(976):328-331.doi:10.1259/bjr/13535955.

共引文献50

同被引文献79

  • 1鄢佳文,高靖琰,刘旭红,朱思瑾,柏晗.ArcCHECK和MatriXX验证系统对非小细胞肺癌患者立体定向放射治疗计划剂量验证通过率的影响[J].医疗装备,2019,0(21):21-22. 被引量:3
  • 2张俊红,邓涤,谢丛华,周云峰.非霍奇金淋巴瘤中枢神经系统侵犯的治疗[J].武汉大学学报(医学版),2005,26(5):639-641. 被引量:2
  • 3Darby S, McGale P, Correa C. Early Breast Cancer Trialists' Collabora- tive Group (EBCTCG). Effect of radiotherapy after breast-conserving surgery on 10-year recurrence and 15-year breast cancer death: meta-analysis of individual patient data for 10, 801 women in 17 rand-omised trials[J]. Lancet, 2011,378(9804): 1707-1716.
  • 4Mackie T, Holmes T, Swerdloff S, et al. Tomothempy: a new concept for the delivery of dynamic conformal radiotherapy[J]. Med Phys, 1993, 20(2): 1709-1720.
  • 5Herzen J, Todorovic M, Cremers F. Dosimetric evaluation of a 2D pix- el ionization chamber for implementation in clinical routine [J]. Phys Med Biol, 2007, (04): 1197-1208.
  • 6Low DA, Harms W-B, Mutic S evaluation of dose distributions[J] A technique for the quantitative Med Phys, 1998, (5): 656-661.
  • 7Buonamici FB, Compagnucci A, Marrazzo L. An intercomparison between film dosimetry and diode matrix for IMRT quality assurance [J]. Med Phys, 2007, (4): 1372-1379.
  • 8Nelms BE, Simon JA. A survey on planar IMRT QA analysis[J]. J Appl Clin Med Phys, 2007, 8(3): 2448.
  • 9Neilson C, Klein M, Barnett R, et al. Delivery quality assurance with AreCHECK [J]. Med Dosim, 2013, 38(1): 77 -80.
  • 10STUDENSKI M T, SHEN X, YU Y, et al. Intensity-modulated radiation therapy and volumetric-modulated arc therapy for adult craniospinal irradiation-a comparison with traditional techniques [J 1. Med Dosim, 2013, 38(1 ): 48-54.

引证文献13

二级引证文献58

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部