期刊文献+

核酸适配体生物传感器对心肌肌钙蛋白Ⅰ的检测研究(英文) 被引量:1

The Detection of cTnⅠ by The Aptamer Biosensor
下载PDF
导出
摘要 核酸适配体生物传感器是利用固定在电极表面的适配子与被测溶液中心肌肌钙蛋白Ⅰ(cTnⅠ)发生特异性结合,从而达到检测的目的.我们对玻碳电极进行阳极氧化、氨基化修饰,通过碳二亚胺盐酸盐(carbodiimide hydrochloride,EDC)、N-羟基琥珀酰亚胺(N-hydroxysuccinimide,NHS)活化作用将适配子结合在电极表面.cTnⅠ最佳检测范围是0.05-5 nmol/L,最低检测限为0.05 nmol/L,检测时间为5 min. A new aptamer biosensor for detection of cTn I was established. It can detect cTn I in samples. The surface of GC was modified with anode-oxidation, amidation and activation modification. We used EDC and NHS as activators. Due to aptamer's specificity for cTn I, we can detect the concentration of cTn I in sample solution. The linear range of the analytical signal is observed from 0.05 -5 nmol/L. The lowest detected concentration is 0.05 nmol/L. The detection time is 5 min.
出处 《生物化学与生物物理进展》 SCIE CAS CSCD 北大核心 2014年第9期916-920,共5页 Progress In Biochemistry and Biophysics
基金 supported by grants from the Ministry of Science and Technology of China(2013DFG52490) Tianjin Municipal Science and Technology Commission(12ZXCXSY07500) The National Natural Science Fundation of China(21076157)~~
关键词 核酸适配体 生物传感器 cTnⅠ aptamer, biosensor, cTn I
  • 相关文献

参考文献17

  • 1Zimmet H. The twilight zone of troponins. Heart, Lung and Circulation, 2003,12:S90~S94.
  • 2Keller T, Zeller T, Peetz D, et al. Sensitive troponin I assay in early diagnosis of acute myocardial infarction. New England Journal of Medicine, 2009,361:868~877.
  • 3Reichlin T, Hochholzer W, Bassetti S, et al. Early diagnosis of myocardial infarction with sensitive cardiac troponin assays. New England Journal of Medicine, 2009,361:858~867.
  • 4Barry JAW, Barth JH, Howell SJ. Cardiac troponins: their use and relevance in anaesthesia and critical care medicine. Continuing Education in Anaesthesia, Critical Care Pain, 2008,8:62~66.
  • 5Tsang WM, Owen AM, Collinson PO, et al. Surveys on the use of cardiac markers in the UK. Annals of clinical biochemistry, 2003,40:138~142.
  • 6Eggers KM, Oldgren J, Nordenskj?ld A, et al. Combining different biochemical markers of myocardial ischemia does not improve risk stratification in chest pain patients compared to troponin I alone. Coronary artery disease, 2005,16:315~319.
  • 7Cummins B, Auckland ML, Cummins P. Cardiac-specific troponin-l radioimmunoassay in the diagnosis of acute myocardial infarction. American heart journal, 1987,113:1333~1344.
  • 8Bodor GS, Porter S, Landt Y, et al. Development of monoclonal antibodies for an assay of cardiac troponin-I and preliminary results in suspected cases of myocardial infarction. Clinical chemistry, 1992,38:2203~2214.
  • 9Oh S-K, Foster K, Datta P, et al. Use of a dual monoclonal solid phase and a polyclonal detector to create an immunoassay for the detection of human cardiac troponin I. Clinical Biochemistry, 2000,33:255~262.
  • 10Cheng AK, Ge B, Yu H-Z. Aptamer-based biosensors for label-free voltammetric detection of lysozyme. Analytical chemistry, 2007,79:5158~5164.