湿度对UV-FAIMS基于谱峰位置的谱图识别影响 被引量:3

Effects of Humidity on the UV-FAIMS Spectrum Identification Based on the Positions of the Spectrum Peaks
下载PDF
导出
摘要 湿度是高场不对称离子迁移谱(FAIMS)检测的一个重要环境参数。常温常压下,以6种常见的挥发性有机物(丙酮、乙醇、异丙醇、异丁醇、邻二甲苯和对二甲苯)为样品,研究了载气湿度对紫外灯高场不对称离子迁移谱(UV-FAIMS)基于谱峰位置的谱图识别影响。实验结果表明,随着载气湿度的增大,各样品最大谱峰数增加。干燥条件下所选样品的主峰位置都很靠近,大都在5 V附近。随着载气湿度增大,各样品主峰位置发生偏移,且偏移程度不同。因此,选择合适的湿度使各样品谱峰数较少,且谱峰位置差异较大,能提高UV-FAIMS基于谱峰位置谱图识别的准确性。 Humidity is a key environment parameter to the high-field asymmetric waveform ion mobility spectrometry(FAIMS).At the ambient temperature and pressure,with the six common volatile organic compounds(VOCs)(acetone,ethanol,2-propanol,2-butanol,O-xylene and P-xylene)as the samples,the effects of the carrier gas humidity on the ultraviolet lamp high-field asymmetric waveform ion mobility spectrometry(UV-FAIMS)spectrum identification based on the position of the spectrum peak were studied.The results show that the number of the maximum peak of the samples increases with the increase of the carrier gas humidity.Under the dry condition,the main peak positions of the selected samples are very close,mostly around 5 V.With the increase of the carrier gas humidity,the main peak positions of the samples offset and the deviation degrees are different.Therefore,through choosing a suitable humidity,the spectral peak number of the samples is less while the difference of the spectral peak position is greater,thus the accuracy of the UV-FAIMS spectrum identification based on the spectrum peak location can be improved.
出处 《微纳电子技术》 CAS 北大核心 2014年第10期655-658,683,共5页 Micronanoelectronic Technology
基金 国家自然科学基金资助项目(61374016) 中国科学院对外合作重点项目(GJHZ1218) 中国科学院青年创新促进会资助项目
关键词 高场不对称离子迁移谱(FAIMS) 紫外灯(UV) 湿度 谱图识别 挥发性有机物(VOC) high-field asymmetric waveform ion mobility spectrometry(FAIMS) ultraviolet lamp(UV) humidity spectrum identification volatile organic compound(VOC)
  • 相关文献

参考文献11

  • 1SHVARTSBURG A A. Differential ion mobility spectrometry [M]. Boca Ration, USA: CRC Press Taylor &. Francis Group, 2009: 144- 146.
  • 2张东风,孔德义,梅涛,陶永春.离子迁移谱仪微型化的现状与进展[J].仪器仪表学报,2006,27(2):199-204. 被引量:22
  • 3KOLAKOWSKI B M, MESTER Z. Review of applications of high-field asymmetric waveform ion mobility spectrometry (FAIMS) and differential mobilityspectrometry (DMS) [J]. Analyst, 2007, 132 (9):842-864.
  • 4SABA J, BONNEIL E, POMIES C, et al. Enhanced sensitivi- ty in proteomics experiments using FAIMS coupled with a hy- brid linear ion trap/orbitrap mass spectrometer [J].Journal of Proteome Research, 2009, 8 (7): 3355- 3366.
  • 5VARESIO E, le BLANC J C Y, HOPFGARTNER G. Real time 2D separation by LC & differential ion mobility hyphena ted to mass spectrometry [J]. Anal Bioanal Chem, 21112, 402 (8): 2555-2564.
  • 6SCHNEIDER B B, COVEY T R, COY S L, et al. Chemical effects in the separation process of a differential mobility/mass spectrometer system [J]. Analytical Chemistry, 2010, 82 (5): 1867- 18811.
  • 7林丙涛,陈池来,孔德义,李庄,王焕钦,程玉鹏,王电令,梅涛.载气流速对高场不对称波形离子迁移谱的影响[J].分析化学,2010,38(7):1027-1030. 被引量:24
  • 8KRYLOV E V, COY S L, NAZAROV E G. Temperature effects in differential mobility spectrometry [J]. International Journal of Mass Spectrometry, 2009, 279 (2/3): 119-125.
  • 9KRYI.OVA N, KRYILV E, EICEMAN G, et al. Effect of moisture on the field dependence of mobility for gas-phase ions of organophosphorus compounds at atmospheric pressure with field asymmetric ion mobility spectrometry [J].The Journal of Physi- cal Chemistry: A, 2003, 107 (19) : 3648- 3654.
  • 10陈池来,赵聪,王电令,程玉鹏,孔德义,王焕钦,高钧,高理升,林新华,王英先,殷世平,张瑞.基于MEMS的新型高场不对称波形离子迁移谱仪[J].微纳电子技术,2011,48(2):112-117. 被引量:6

二级参考文献49

  • 1张东风,孔德义,梅涛,陶永春.离子迁移谱仪微型化的现状与进展[J].仪器仪表学报,2006,27(2):199-204. 被引量:22
  • 2时迎国,劭士勇,李安林,姚琏,王宾,李芳,王俊德,李海洋.迁移管的电场强度对真空紫外电离-离子迁移谱仪性能的影响[J].分析化学,2006,34(9):1353-1356. 被引量:13
  • 3Buryakov I A,Krylov E V,Nazarov E G,Rasulev U K.Int.J.Mass Spectrom.Ion Processes,1993,128(3):143-148.
  • 4Kolakowski B M,Mester Z.Analyst,2007,132(9):842-864.
  • 5Hatsis P,Kapron J T.Rapid Commun.Mass Spectrom.,2008,22(5):735-738.
  • 6Ells B,Barnett D A,Purves R W,Guevremont R.J.Environ.Monit.,2000,2(5):393-397.
  • 7Miller R A,Nazarov E G,Eiceman G A,Thomas K A.Sensors and Actuators.A,2001,91(3):301-312.
  • 8Purves R W,Guevremont R,Day S,Pipich C W,Matyjaszczyk M S.Rev.Sci.Instrum.,1998,69(12):4094-4105.
  • 9Krylov E V.Int.J.Mass Spectrom.,2003,225(1):39-51.
  • 10YANG Jin-Ji(杨津基).Gas Discharge(气体放电).Beijing(北京):Science Press(科学出版社),1983:68-70.

共引文献46

同被引文献17

引证文献3

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部