期刊文献+

聚季铵盐为结合相的薄膜梯度扩散技术富集测量自来水中痕量Cr(VI) 被引量:2

Determination of Cr(VI) in tap water with poly quaternary ammonium salt as a binding phase in diffusive gradients in thin-films
下载PDF
导出
摘要 建立了以0.05mol/L聚季铵盐(PQAS)溶液为结合相的薄膜梯度扩散(DGT)装置(PQAS DGT)富集测量自来水中痕量Cr(VI)的分析方法。DGT法测得配制水中Cr(VI)的回收率为93.5%-102.4%,相对标准偏差(RSD)为2.25%-4.33%;测得自来水中Cr(VI)的浓度为2.68-3.45g/L,加标回收率为94.7%-102.6%;DGT方法对水中Cr(VI)的检出限为0.065g/L(采样48h),可应用于自来水中痕量Cr(VI)的定量检测与早期预警。 Cr (Vl) in tap water was measured by the diffusive gradients in thin-films (DGT) technique using0.05mol/L poly quaternary ammonium salt(PQAS) solution as binding phase(PQAS DGT). The recoveries of Cr(Vl) in prepared solutions measured by PASP DGT were 93.5%-102.4% with relative standard deviation (RSD)2.25%N4.33%. The concentration of Cr(Vl) in tap water measured by PQAS DGT was 2.68-3.45g/L withrecoveries of 94.7%-102.6%. The method detect limit for PQAS DGT was 0.065g/L(deployment time 48h).Therefore, PQAS DGT could be used to measure trace Cr(Vl) and gave an early alarm in tap water.
出处 《食品工业科技》 CAS CSCD 北大核心 2014年第20期69-71,85,共4页 Science and Technology of Food Industry
基金 “十二五”国家科技支撑计划(2012BAD29B06) 辽宁省科学事业公益研究基金项目(2012003001) 辽宁省食品安全重点实验室开放课题(LNSAKF2011038)
关键词 自来水 铬(VI) 薄膜扩散梯度 聚季铵盐 tap water Cr (Vl) diffusive gradients in thin-films poly quaternary ammonium salt