血糖仪与全自动生化仪检测末梢血和静脉血血糖的比较 被引量:10

Comparison of finger capillary glucose versus vein blood glucose levels between three glucose meters and automatic biochemistry analyzer
原文传递
导出
摘要 目的 比较血糖仪与全自动生化仪检测末梢血和静脉血的一致性. 方法 选取2012年11~12月于我科门诊就诊的糖尿病和非糖尿病患者197例,使用3款血糖仪和日立7180全自动生化分析仪分别检测末梢血和静脉血血糖,并行线性回归分析和配对t检验. 结果 当红细胞压积(Hct)在35.1%~51.6%时,3款血糖仪与日立7180相关性顺序由高到低为罗氏Accuchek(末梢血:R2=0.968,静脉血:R2=0.966)、爱科莱GT-1920、强生One Touch.爱科莱GT-1920检测末梢血与日立7180的偏倚较小,检测静脉血的偏倚较大.罗氏Accuchek检测静脉血的均值低于日立7180.强生One Touch的偏倚约为10%. 结论3 款血糖仪与全自动生化分析仪相关性均较好,Hct可对两者的相关性造成影响. Objective To compare the agreement for glucose assessing in capillary versus vein blood between glucose meters and automatic analyzer.Methods 197 patients were collected from outpatients from Nov.2012 to Dec.2012.Fingertip and vein blood glucose were tested by three glucose meters and automatic biochemistry analyzer (Hitachi 7180)separately.Linear regression and paired t test were performed.Results When hematocrit (Hct) had the range of 35.1%~51.6%,the correlation between glucose meters and automatic analyzer (Hitachi 7180)was in the following agreement order:Accuchek (R2 =0.968 for finger blood and R2 =0.966 for vein),GT-1920 and One Touch versus Hitachi 7180,respectively.GT-1920 showed the lowest bias to Hitachi 7180 for capillary blood versus the vein blood.One Touch had the bias of 10%.Conclusion The three glucose meters show good correlation with Hitachi 7180 and Hct influences the correlation.
出处 《中国糖尿病杂志》 CAS CSCD 北大核心 2014年第10期931-933,共3页 Chinese Journal of Diabetes
关键词 血糖仪 全自动生化仪 血糖 比较 Glucose meters Automatic biochemistry analyzer Blood glucose Comparison
  • 相关文献

参考文献4

  • 1周博锋,何本进.POCT血糖仪与全自动生化仪对血糖测定结果的比对分析[J].广西中医学院学报,2009,12(3):23-24. 被引量:22
  • 2Kulkarni A, Saxena M,Price G, et al. Analysis of blood glucose measurements using capillary and arterial blood samples in in- tensive care patients. Intensive care medicine, 2005, 31: 142-145.
  • 3Finkielman JD, Oyen LJ, Afessa B. Agreement between bed- side blood and plasma glucose measurement in the ICU setting. CHEST Journal, 2005,127 : 1749-1751.
  • 4Whitley RJ, Santrach PJ, Phillips DL Establishing a quality man- agement system for unit-use testing based on NCCLS proposed guideline (EP18-P). Clinica ehimica acta, 2001,307 : 145-149.

二级参考文献1

共引文献21

同被引文献96

  • 1韩秀玲.血糖检测新进展[J].医学信息(医学与计算机应用),2014,0(32):390-390. 被引量:1
  • 2刘凤凤,周芳.快速血糖仪与全自动生化分析仪的对比研究[J].中国医药导报,2006,3(35). 被引量:7
  • 3丁红香,徐晓杰,张美芬,倪莉.POCT血糖仪与生化分析仪血糖检测结果的比对试验及分析[J].中华检验医学杂志,2007,30(12):1374-1375. 被引量:49
  • 4Rebel A, Rice MA, Fahy BG. The accuracy of Point-of-Care glu-cose measurements[J]. J Diabetes Sci Technol, 2012,6(2):396-411.
  • 5O' kane MJ. The accuracy of point-of-care glucose measurement[J]. Ann Clin Biochem, 2012 ,49(2) : 108-109.
  • 6Tonyushkina K,Nichols JH. Glucose meters:a review of technicalchallenges to obtaining accurate results[J]. J Diabetes Sci Techn-ol, 2009 ,3(4) : 971-980.
  • 7中华人吳共和国卫生部.医疗机构便携式血糖检测仪管理和临床操作规范(试行)[R].北京:中华人民共和国卫生部,2010.
  • 8Cook A, Laughlin D,Moore M,et al. Differnces in glucose valuesobtained from point-of-care glucose meters and laboratory analysisin critically ill patients[J]. Am J Crit Care,2009 ,18(1 ) : 65*72.
  • 9Shearer A, Boehmer M, Closs M, et al. Comparison of glucosepoint-of-care values with laboratory values in critically ill patients[J]. Am J Crit Care,2009,18(3)=224-230.
  • 10Bland JM, Altman DG. Statistical methods for assessing agree-ment between two methods of clinical measurement[J j. Lancet,1986,1(8476):307-310.

引证文献10

二级引证文献66

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部