在患者胸部正电子发射成像(PET)过程中,呼吸运动会造成成像质量下降。本文使用三层网格B-样条弹性配准的方法修正重组和降噪后的正弦图来进行呼吸运动校正。使用GATE仿真NCAT呼吸运动体模数据进行实验,结果显示图像质量得到了很大地提高,运动伪影和结构信息得到了修复,表明本文的方法是有效的。
In the process of positron emission tomography (PET) data acquiring, respiratory motion reduces the qual- ity of PET imaging. In this paper, we present a correction method using three level grids B-spline elastic method to correct denoised and reorganized sinograms for respiratory motion correction. Using GATE simulates NCAT respira- tory motion model to generate raw data which are used in experiment, the experiment results showed a significantly improved respiratory image with higher quality of PET, and the motion blur and structural information were fixed. The results proved the method of this paper would be effective for the elastic registration.
出处
《生物医学工程学杂志》
EI
CAS
CSCD
北大核心
2014年第5期1005-1010,共6页
Journal of Biomedical Engineering
基金
国家自然科学基金资助项目(11265007)
教育部回国人员科研启动基金资助项目(2010-1561)
关键词
运动校正
弹性配准
正电子发射成像
正弦图
motion correction
elastic registration
positron emission tomography
sinogram