头体一体式体位固定板在全中枢螺旋断层放射治疗中的应用 被引量:12

Positioning research on helical tomotherapy for craniospinal irradiation
下载PDF
导出
摘要 目的:观察分析头体一体式体位固定板在全中枢神经系统放射治疗中的优势。方法:回顾性分析20例使用螺旋断层放疗设备实施全中枢放疗的患者资料,其中10例使用了头体一体式体位固定板,10例使用了头体分体式固定板,将两组患者在治疗过程中各个方向的摆位误差数据进行分析。结果:头体一体式体位固定板组患者X轴和Y轴方向摆位误差显著小于头体分体式固定板组,其差异有统计学意义(t=3.489,t=4.750;P<0.05);Z轴方向摆位误差两组比较无显著差异。结论:头体一体式体位固定板在全中枢放疗时可以减小头脚方向的摆位误差,与螺旋断层调强放疗相结合可提高治疗精度,缩短治疗时间,值得在临床工作中广泛应用。 Objective: To observe the advantages of single head-body board in decreasing positioning error for cranniospinal irradiation. Methods: Twenty patients accepting craniospinal irradiation treated with tomotherapy. Ten of the patients used the head-board plus body-board separately to fix up the body in treatment, and another 10 used the single head-body board. The data of positioning error and side effects during the treatment were analyzed. Results: The positioning error in patients using head-body board was less than that using the head-board plus body-board in X and Y direction(t=3.489, t=4.750; P〈0.05), and the compare data of the positioning error in Z directions had no statistical significance. Conclusion: The single head-body board can decrease the positioning error in superior-inferior direction for craniospinal irradiation, and it can shorten the treatment time, so it is worth wide usage in clinical treatment. The advice is to design different types and plastic models to fit patients with different ages and different body types.
出处 《中国医学装备》 2014年第9期52-55,共4页 China Medical Equipment
基金 中国人民解放军总医院苗圃基金(11KMM45)"全骨髓和全淋巴系统螺旋断层放疗新技术的研究"
关键词 全中枢神经系统 放射治疗 螺旋断层放疗 头体一体式体位固定板 Central nervous system Radiotherapy Helical tomotherapy Head-body board
  • 相关文献

参考文献14

  • 1Bauman G,Yartsev S,Coad T,et al. Helical tomotherapy for craniospinal radiation[J].Br J Radiol, 2005,78(930). 548-552.
  • 2曲宝林,徐寿平,戴相昆,张慧娟,崔书祥,马林.全中枢神经系统螺旋断层放射治疗初探[J].军医进修学院学报,2008,29(2):137-139. 被引量:14
  • 3李雪南,修霞,李高峰.大孔径螺旋CT在放疗模拟定位中的应用研究[J].中国医学装备,2013,10(10):16-18. 被引量:6
  • 4姚伟荣,徐寿平,杜镭,解传滨,马林.螺旋断层治疗机实施剂量重建和剂量引导放疗的研究[J].中国医疗器械杂志,2012,36(5):375-377. 被引量:7
  • 5Sterzing F,Schubert K,Sroka-Perez G,et al.HeEcal tomotherapy--Experience of the first 150 patients in Heidelberg[J].Strahlenther Onkol, 2008, 184(1):8-14.
  • 6Langen KM,Zhang Y,Andrews RD,et al.Initial experience with megavoltage(MV)CT guid-ance daily prostate alignments[J].int J Radiat Oncoi Biol Phys, 2006,66(3): 876-882.
  • 7Amish P,Katja M,Kenneth J,et al. Patient close from megavoltage computed tomography imaging[J].Int J Radiat Oncol Biol Phys, 2008,70(5) : 1579-1587.
  • 8Li B,Lu H,Luo LM,et al.Novel technique for craniospinal radiotherapy with patient supine[J].J Southeast University(English Edition),2009,25(3): 326-329.
  • 9Yoon M,Shin DH,Kim J,et al. Craniospinal irradiation techniques:a dosimetric comparison of proton beams with standard and advanced Dhoton radiotherapy[J].Int J Radiat Oncol Biol P hys, 2011,81(3): 637-646.
  • 10Sharma DS,Gupta T,Jalali R,et al. High- precision radiotherapy for craniospinal irradiation:evaluation of three-dimensional conformal radiotherapy, intensity-modulated radiation therapy and helical tomotherapy[J].Br J Radiol, 2009,82(984): 1000-1009.

二级参考文献32

  • 1亓恒涛,秦维昌,刘传亚,王巍.CT机质量控制检测的探讨[J].医学影像学杂志,2006,16(9):978-981. 被引量:24
  • 2王蒙.MATLAB7辅助信号处理技术与应用[M].北京.电子工业出版社:2005
  • 3郑穗生,高斌,刘斌,等.CT诊断与临床[M].2版.合肥:安徽科学技术出版社,2011:116-117.
  • 4郭晶,孙伟娟.小波分析理论与MATLAB7实现[M].北京:电子工业出版社,2005.
  • 5Fenwick JD, Tome WA, Kissick MW, et al. Modelling simple helically delivered dose distributions [ J ]. Phys Med Biol, 2005, 50 : 1505-1517.
  • 6Robert Jeraj, Thomas R. Radiation characteristics of helical tomotherapy[J]. Med Phys, 2004, 31(12) :396-404.
  • 7Kissick MW, Fenwick J, James JA, et al. The helical tomotherapy thread effect[J]. Med Phys , 2005, 32:1414-1423.
  • 8Bauman G, Yartsev S, Coad T, et al. Helical tomotherapy for craniospinal radiation [ J ]. The British Journal of Radiology, 2005, 78:548-552.
  • 9Shepard DM, Olivera GH, Reekwerdt PJ, et al. Iterative approaches to dose optimization in tomotherapy [ J ]. Phys Med Biol, 2000.45:69-90.
  • 10Ruchala K J, Olivera GH, Kapatoes JM, et al. Megavoltage CT image reconstruc.tion during tomotherapy treatments [ J ]. Phys Med Biol, 2000, 45:3545-3562.

共引文献23

同被引文献78

引证文献12

二级引证文献71

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部