导管室放射防护评价 被引量:2

Evaluation of Radiological Protection in Catheter Room
下载PDF
导出
摘要 目的:评价放射防护在导管室介入治疗中的应用价值。方法在导管室60例介入手术中应用各类防护措施,利用BH3103B便携式辐射仪检测X线辐射平均剂量率,分析防护措施的效果。结果铅玻璃防护屏有效防护率为95.2%,铅防护服防护效率为94.9%,床下铅橡胶帘防护效率为88.9%,上述防护器材前后辐射剂量率差异均具统计学意义(P<0.01)。距X线球管1~2 m处的射线衰减量为60.2%,距球管1~3 m处的射线衰减量为90.9%,1~2m和1~3m之间的X线剂量率差异具有统计学意义(P<0.01)。结论采用防护措施后可明显降低X射线对医护人员的危害。 Objective To evaluate the application value of radiological protection in interventional treatments conducted in catheter room. Methods The effects of various kinds of radiological protection measures used in 60 cases of interventional operations were analyzed with BH3103B portable radiation detector which was used to measure the average dose rate of X-ray. Results There were significant differences between the radiation dose rate which was measured in front of lead-glass protecting screen and the radiation dose rate which was measured behind of lead-glass protecting screen (P<0.01), whose effective protection rate was 95.2%. There were signiifcant differences between the radiation dose rate which was measured in front of lead protective clothing and the radiation dose rate which was measured behind of lead protective clothing (P〈0.01), whose effective protection rate was 94.9%. There were signiifcant differences between the radiation dose rate which was measured in front of under-bed lead-rubber shield and the radiation dose rate which was measured behind of under-bed lead-rubber shield (P〈0.01), whose effective protection rate was 88.9%. The attenuation rate of X-ray which was 1~2 m away from the tube was 60.2%while that of X-ray which was 1~3 m away from the tube was 90.9%. There were signiifcant differences between X-ray dose rate which was 1~2 m away from the tube and X-ray dose rate which was 1~3 m away from the tube (P〈0.01). Conclusion The adoption of radiological protection measures can signiifcantly reduce the damage of X-ray to medical staff.
作者 陶功福 张枫
出处 《中国医疗设备》 2014年第10期115-116,159,共3页 China Medical Devices
关键词 辐射防护 介入治疗 便携式辐射仪 防护器材 radiological protection interventional treatment portable radiation detector protective devices
  • 相关文献

参考文献12

  • 1Valentin J.Avoidance of radiation injuries from medical intervenfional procedures[J].ICRP Pubficafion 85,2000,30(2):57-67.
  • 2Valentin J.Avoidance of radiation injuries:from medical interventional procedures[J].Annals of the ICRP,2000,30(2):7-67.
  • 3Papageorgiou E,Tsapaki V,Tsalafoutas IA,et al.Comparison of patient doses in interventional radiology procedures performed in two large hospitals in Greece[J].Radiation Protection Dosimetry,2007,124(2) :97 - 102.
  • 4Sandborg M,Fransson SG,PetterssonH.Evaluation of patient-absorbed doses during coronary angiography and intervention by femoral and radial artery access[J].European P-adiology,2004,14(4) : 653-658.
  • 5Efstathopoulos EP,Katritsis DG,Kottou S,et al.Patient and staff radiation dosimetry during cardiac electrophysiology studies and catheter ablation procedures:a comprehensive analysis [J]. Europace,2006,8(6):443 -448.
  • 6陈胜利,黄齐好,邹蓉珠,陈国东,黄子诚,慕容爱.介入放射治疗中医师的辐射防护评价[J].中华放射医学与防护杂志,2002,22(6):451-452. 被引量:5
  • 7张冠石.医护人员的放射防护[J].中国医疗设备,2013,28(5):15-16. 被引量:8
  • 8杨新芳,赵进沛,刘士敏,李秀芹.介入放射学的防护问题与管理对策[J].中国辐射卫生,2010,19(4):416-417. 被引量:7
  • 9郁鹏,尉可道,李田昌,程玉玺,燕树林,刘澜涛.介入诊疗区域内辐射场的测定与评价[J].中华放射医学与防护杂志,2005,25(3):270-271. 被引量:11
  • 10师勤莹,张锦.心血管介入诊疗中术者站立区域辐射剂量的测量与分析[J].中国医疗设备,2014,29(1):39-41. 被引量:8

二级参考文献36

  • 1胡利丰,董文骏,王群利.介入放射学中的辐射防护问题[J].中国辐射卫生,2004,13(4):304-306. 被引量:4
  • 2郁鹏,尉可道,李田昌,程玉玺,燕树林,刘澜涛.介入诊疗区域内辐射场的测定与评价[J].中华放射医学与防护杂志,2005,25(3):270-271. 被引量:11
  • 3[1]GB 8279-87,医用诊断X线卫生防护标准[S].
  • 4Valentin J. Avoidance of radiation injuries from medical interven- tinal procedure[J]. Ann ICRP, 2000, 30:7 - 67.
  • 5Miller DL, Balter S, Cole PE, et al. Radiation doses in interventional radiology procedures: the RAD-IR study pare I: overall measures of dose [J]. J Vasc Inter Radiol, 2003, 14: 711 - 727.
  • 6UNSCEAR. Sources and Effects of loniziong Radiation [ M ]. New York: UN, 2000.
  • 7白玫,刘彬,郑钧正.心脏介入患者受照剂量研究[J].介入放射学杂志,2007,16(4):222-225. 被引量:14
  • 8Padovani R, Rodella CA. Staff dosimetry in interventional Cardiology.Radiat Prot Dosim, 2001,94(1-2) : 99-103.
  • 9Vano E, Gonzalez L, Guibelade E, et al. Radiation exposure to medical staff in interventional and cardiac radiology. Br J Radiol, 1998, 71:954-960.
  • 10Mcketty MH. Study of radiation doses to Personnel in a Cardiac catheterization laboratory. Health Phys,1996, 70: 563-567.

共引文献45

同被引文献34

  • 1汪洋.应用CCD成像的脉冲X射线透视技术研究[J].贵州大学学报(自然科学版),2005,22(4):419-422. 被引量:1
  • 2林国辉.浅议医用X光机焦点的检测[J].上海计量测试,2007,34(4):25-26. 被引量:4
  • 3乔文,严惠民.CR系统的现状及进展[J].光学仪器,2007,29(4):79-83. 被引量:5
  • 4乔文,严惠民.X光机数字化的研究进展[J].光学仪器,2007,29(5):80-84. 被引量:8
  • 5Strender L.Incidence of heart disease and functional significance of change in the electrocardiogram 10 year after radiotherapy for breast cancer].Cancer,1986,57(5):929.
  • 6Mitrael A.Cardiovascular sepuelae of therapeutic thoracic radiation[J] .Prog Cardiovasc Dis, 1991,33(5) :299.
  • 7Tamirisa PK,Holland MR,Miller JG,et al.Ultrsonic tissue characterization:Review of an approach to assess hypertrophic myocardium[j].Echocardiography,2001,18:593-597.
  • 8Schecter SO,Teichholz LF,Klig V,et al.Ultrsonic tissue characterization:Review of noninvasive technique for assessing myocardial viability[J].Echocardiography,1996,13(4):415-430.
  • 9Zuber M,Gerber K,Erne P,et al.Myocardial tissue characterization in heart failure by real-time integrated backscatter[J].Eur J Ultra, 1999,9(2):135-143.
  • 10Wagdi P,Rouvinez G,Fluri M,et al.Cardioprotection in chemo and radiotherapy for malignant diseases:An echocardiographic pilot study[J].Schweiz Rundsch Med prax,1995,84(43):1220- 1223.

引证文献2

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部