自动管电流调节下不同管电压及噪声指数对下肢动脉CTA患者辐射剂量的影响 被引量:4

下载PDF
导出
摘要 目的探讨采用自动管电流调节技术下不同管电压及噪声指数对下肢CTA扫描患者辐射剂量的影响。方法将72例行下肢CTA扫描患者随机分为A、B、C 3组,管电压和噪声指数分别为100 kV、15,80 kV、15及100 kV、20。扫描时在配合自动管电流调节技术(ATCM)条件下,根据原始图像和最大密度投影(MIP)重组图像对3组下肢动脉图像质量进行评分,测量血管CT值、SNR、CNR及患者所接受辐射剂量。统计学分析采用单因素方差分析,P<0.05为差异有统计学意义。结果 1血管CT值、CNR及SNR,B组高于C组和A组,差异有统计学意义(P<0.05),A、C组间差异无统计学意义;2辐射剂量C组最低,容积CT剂量指数(CTDvol)、剂量长度乘积(DLP)及有效剂量(ED)较A组分别减低50.45%、52.09%、51.89%,较B组减低34.78%、34.90%、35.59%,C组与A组、B组间差异有统计学意义(P<0.05),A组与B组之间差异无统计学意义(P>0.05)。结论下肢CTA使用ATCM在适当范围内噪声指数增加辐射剂量降低,而管电压降低图像对比增加。
出处 《中外医疗》 2014年第23期178-179,共2页 China & Foreign Medical Treatment
  • 相关文献

参考文献9

  • 1Albrecht T, Foert E, Hohkamp R, et al. 16-MDCT angiography of aor- toiliac and lower extremity arteries: comparison with digital subtraction angiagraphy[J]. A JR Am J Roentgenol,2007,189(3): 702-711.
  • 2McCollough CH, Primak AN, Braun N, et al. Strategies for reducing ra- diation dose in CT [J]. Radiol Clin North Am, 2009. 47(1): 27-40.
  • 3Huda W, Sealzetti EM, Levin G. Technique factors and image quality as functions of patient weight at abdominal CT[J].Radiology,2000,217(2): 430-435.
  • 4Zhang WL, Li M, Zhang B, et al. CT Angiography of the Head-and- Neck Vessels Acquired with Low Tube Voltage, Low Iodine, and hera- tire Image Reconstruction: Clinical Evaluation of Radiation Dose and Image Quality[J]. PLoS One,2013,8(12): e81486.
  • 5Beitzke D, Wolf F, Edelhauser G, et al. Computed tomography angiogra- phy of the carotid arteries at low kV (p) settings_ a prospective ran- domised trial assessing radiation dose and diagnostic confidence[J]. Eur Radiol, 2011,21(11 Germany) : 2434-2344.
  • 6Sano T, Matsutani H, Kondo T, et al. [Prospective electrocardiogram gat- ed coronary 320-row area detector computed tomography angiography using low tube current scanning with full reconstruction] [J]. Nihon Hnshasen Gijutsu Gakkai zasshi,2013, 69(3 Japan):244-250.
  • 7Lee EJ, Lee SK, Agid R, et al.Comparison of image quality and radiation dose between fixed tube current and combined automatic tube current modulation in craniocervical CT angiography[J]. AJNR Am J Neuroradi- ol,2009,30(9): 1754-759.
  • 8Sookpeng S, Martin C J, Gentle D J, et al. Relationships between patient size, dose and image noise under automatic tube current[J], modulation systems. J Radiol Prot,34(1):2013,103-123.
  • 9王益钢,林俊,丁建平,张祖峰,陈丹娜.噪声指数指导自动管电流调制技术在肺结核CT复查中降低辐射剂量的应用价值[J].浙江临床医学,2013,15(12):1814-1816. 被引量:3

二级参考文献9

共引文献2

同被引文献30

引证文献4

二级引证文献19

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部