基于肝血管树的三维超声自动图像融合初步研究 被引量:1

A preliminary study of three-dimensional ultrasound images automatic registration based on hepatic vessel
原文传递
导出
摘要 目的 探讨基于肝血管树的三维超声(3DUS)图像自动融合新技术在临床应用的可行性及准确性.方法 应用基于肝血管树3DUS图像融合技术对10例健康志愿者肝进行三维超声容积图像自动融合,数据采集和融合部位包括:左肝外叶、左肝内叶、右肝前叶及右肝后叶各10对三维超声容积数据,共40对,评估该技术在肝上述各部位的成功率及准确性.结果 36对三维超声容积图像自动融合成功,成功率为90%(36/40),配准误差约1.48 mm(0.98~2.76 mm);左肝外叶、左肝内叶、右肝前叶及右肝后叶融合成功率差异无统计学意义(P=0.891),而配准误差右肝前叶小于左肝外叶(P=0.014).结论 基于肝血管树自动配准的3DUS图像融合技术是可行的,在将来临床应用中具有广阔的前景. Objective To evaluate the feasibility and accuracy of three-dimensional ultrasound (3DUS) automatic image fusion based on hepatic vessel in clinical application.Methods Forty pairs of 3D ultrasound volumetric images from 10 healthy volunteers were acquired and enrolled in the study,including 10 pairs of each following lobe of the liver:left lateral lobe,left medial lobe,right anterior lobe and right posterior lobe.3DUS automatic registrations were performed based on hepatic vessel.The technical successful rate and accuracy of the image fusion using the 3DUS data from four hepatic lobes were observed and compared.Results Thirty-six pairs of volumetric data were successfully fused together by automatic registration.The technical successful rate was 90% (36/40) and the registration error distance was 1.48 mm (0.98-2.76 mm).There were no statistical differences in the successful rate for the four different hepatic lobes (P =0.891),but the registration error of right anterior lobe was less than left lateral lobe of liver (P =0.014).Conclusions The new technology of 3DUS automatic registration based on hepatic vessel was feasible and reliable,and it has a wide prospect in future clinical application.
出处 《中华超声影像学杂志》 CSCD 北大核心 2014年第10期865-868,共4页 Chinese Journal of Ultrasonography
基金 国家自然科学基金(81271669,81301931)
关键词 超声检查 三维 图像融合 Ultrasonography,three-dimensional Liver Image fusion
  • 相关文献

参考文献3

二级参考文献59

  • 1赵永明,肖昌炎,陈亚珠.多模态图像配准在HIFU定位系统中的应用[J].北京生物医学工程,2004,23(4):246-250. 被引量:2
  • 2严昆,陈敏华,杨薇,戴莹,尹珊珊,王艳滨,霍苓.超声造影评价肝恶性肿瘤射频治疗疗效──与常规超声及增强CT比较[J].中华超声影像学杂志,2005,14(9):655-658. 被引量:43
  • 3刘广健,吕明德,谢晓燕,周旭辉,徐辉雄,匡铭,徐作峰,郑艳玲,梁瑾瑜,黄蓓.实时虚拟导航系统引导消融治疗肝癌[J].中华超声影像学杂志,2006,15(10):758-760. 被引量:15
  • 4Maintz JB,Viergever MA. A survey of medical image registration. Med Image Anal, 1998,2 : 1-36.
  • 5Barra V, Boire JV. A general framework for the fusion of anatomical and functional medical images. Neuroimage, 2001,13 : 410-424.
  • 6Comeau RM, Sadikot AF, Fenster A, et al. Intraoperative ultrasound for guidance and tissue shift correction in image guided neurosurgery. Med Phys, 2000,27 : 787-800.
  • 7Pelizzari CA, Chen GT, Spelbring DR, et al. Accurate three-dimensional registration of CT, PET and/or MR images of the brain. J Comput Assist Tomogr, 1989,13:20-26.
  • 8Yu JN, Fahey FH, Gage HD, et al. Intermodality, retrospective image registration in the thorax. J Nucl Med, 1995,36:2333-2338.
  • 9Hata N, Dohi T, Iseki H, et al. Development of a frameless and armless stereotactic neuronavigation system with ultrasonographic registration. Neurosurgery, 1997,41:608-614.
  • 10Lindseth F, Kaspersen JH, Ommedal S, et al. Multimodal image fusion in ultrasound-based neuronavigation: improving overview and interpretation by integrating preoperative MRI with intraoperative 3D ultrasound. Comput Aided Surg, 2003,8 : 49-69.

共引文献31

同被引文献8

引证文献1

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部