Sysmex XN-9000全自动血液体液分析仪性能评价 被引量:5

Performance Assessment of Sysmex XN-9000 Fully Automatic Blood and Body Fluid Analyzer
下载PDF
导出
摘要 目的:对Sysmex XN-9000全自动血液体液分析仪的主要性能进行初步评价。方法根据美国临床和实验室标准化协会(CLSI)指南文件EP15-A2、EP6-A2的要求,对XN-9000仪进行精密度、携带污染率、线性范围和准确度验证,以Beckman LH750全自动血液分析仪的检测结果作为评估依据,将验证结果与厂商声明的性能指标或公认的质量标准进行比较。结果 XN-9000仪测定白细胞(WBC)、红细胞(RBC)、血红蛋白(HGB)、红细胞压积(HCT)和血小板(PLT)主要参数的批内变异系数(CV%)为0.11%~2.42%,批间CV%为0.12%~2.50%,线性相关性分析r为0.99991~0.99999,携带污染率为0.00%~0.38%。对比相关性分析,各项指标r均〉0.999。准确度验证各项指标的偏倚为-0.67%~0.96%。结论 XN-9000仪的精密度、携带污染率、线性范围和准确度等主要评估指标均符合要求。 Objective To evaluate the main performance of Sysmex XN-9000 fully-automatic blood and body fluid analyzer. Methods According to the guidelines EP15-A2 and EP6-A2 of clinical and laboratory standards institute (CLSI), the precision, contamination rate, linearity range and accuracy of XN-9000 analyzer were verified based on the detection results of Beckman LH750 full-automatic blood analyzer which were choosed as assessment basis. Then the verification results of XN-9000 analyzer were compared with performance indicators declared by its manufacturer or recognized quality standards. Results The intra-assay and inter-assay coefficient variations (CV%) of major parameters including WBC, RBC, HGB, HCT and PLT measured by XN-9000 analyzer were 0.11%~2.42% and 0.12%~2.50%, respectively. The correlation coefficient (r ) of linear correlation analysis between observational values and theoretical values of the above parameters was 0.99991~0.99999 while the contamination rate of the above parameters was 0.00%~0.38%. The correlation coefficient (r ) of comparative correlation analysis between the detection indicators of XN-9000 analyzer and those of Backman LH750 analyzer was greater than 0.999. The Bias of accuracy verification of the above parameters of XN-9000 analyzer was -0.67%~0.96%. Conclusion The main parameters including precision, contamination rate, linearity range and accuracy of XN-9000 analyzer can meet the requirements of relevant standards.
作者 施俊柱
出处 《中国医疗设备》 2014年第11期131-133,120,共4页 China Medical Devices
关键词 全自动血液体液分析仪 精密度 携带污染率 线性范围 准确度 fully-automatic blood and body lfuid analyzer precision contamination rate linearity range accuracy
  • 相关文献

参考文献8

二级参考文献41

  • 1吴健民.影响肿瘤标志物检测的因素[J].中华检验医学杂志,2005,28(4):352-354. 被引量:22
  • 2戈俊,徐玉华.LH750全自动血细胞分析仪常见故障修理[J].医疗卫生装备,2006,27(10):106-106. 被引量:15
  • 3张莉,吴炯,郭玮,潘柏申.医学检验检测系统应用前的性能评价[J].检验医学,2006,21(6):560-563. 被引量:71
  • 4U. S Department of Health and Human Services. Medicare, Med icaid, and CLIA programs, laboratory requirements relating to quality systems and certain personnel qualifications. Final rule [S]. Fed Regist, 2003, 68(16): 3 639-3 714.
  • 5Anonymous. Medical devices, current good manufacturing practice (CGMP) final rule, quality system regulation-FDA. Final rule [S]. Fed Regist, 1996, 61(195):52 602-52 662.
  • 6National Committee for Clinical Laboratory Standards. Evaluation of the linearity of quantitative measurement procedures: a statisti cal approach. Approved Guideline [S]. Wayne, PA: EP6-A, NCCLS. 2003.
  • 7National Committee for Clinical Laboratory Standards. Preliminary evaluation of quantitative clinical laboratory methods,approved guideline [S]. 2nd ed. Wayne, PA: EP10-A2. NCCLS, 2002.
  • 8U. S. Department of Health and Human Services, Medi Care. Clinical and laboratory standards institute, CLSI [S]. Fed Regis St,2006,57(40):7002.
  • 9杨由业,张秀明.临床检验方法学评价[M].北京:人民卫生出版社,2008:118.
  • 10[4]熊立凡,李树仁.临床检验基础.第3版.北京:人民卫生出版社,2001.7-55.

共引文献399

同被引文献25

引证文献5

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部