摘要 目的利用VITEK-2 Compact全自动细菌鉴定仪对BACTEC 9120血培养阳性标本进行直接菌种鉴定,并将结果与传统检测方法相比较,以探讨快速获得阳性血培养结果的可行性。方法共收集60例单细菌血培养阳性标本,其中36例为革兰阴性杆菌,24例为革兰阳性球菌。应用VITEK-2 Compact全自动细菌鉴定仪对血培养阳性标本进行直接鉴定,同时与传统操作方法结果相比较。结果血培养阳性标本直接经VITEK-2 Compact全自动细菌鉴定仪鉴定后,97.2%的革兰阴性杆菌和75.0%的革兰阳性球菌与传统操作方法所得鉴定结果相一致。结论在直接鉴定革兰阴性杆菌血培养阳性标本时,应用VITEK-2 Compact全自动细菌鉴定仪进行鉴定是可行的,但在鉴定革兰阳性球菌时可靠性稍差。该方法在一定程度上可以有效缩短获得血培养结果的周期,对指导临床治疗起到积极的作用。 Objective We performed a study to compare identification profile of positive BACTEC 9120 blood cultures by VITEK- 2Compact automatic microbial analysis system with traditional method,in order to provide the clinician with early information about blood culture results. Methods Sixty positive blood cultures with monomicrobial bacteremia( 36 gram- negative bacilli and 24 gram- positive cocci) were identified by both VITEK- 2 Compact automatic microbial analysis system and traditional method. The results were compared. Results Comparison between the results obtained with VITEK- 2 Compact instrument and traditional method revealed that 97. 2% of the gram- negative bacilli and75. 0% of the gram- positive cocci identified by VITEK- 2 Compact instrument were coincident with the results obtained with traditional method.Conclusion These data suggest that direct identification of the gram- negative bacilli from blood culture bottles leads to the rapid and reliable results via VITEK- 2 Compact automatic microbial analysis system,but the reliability of this performance on gram- positive cocci is relatively poor.The direct identification may reduce turnaround time and allow a better prognosis.
出处 《临床和实验医学杂志》 2014年第22期1895-1897,共3页 Journal of Clinical and Experimental Medicine
关键词 血培养 VITEK-2 Compact全自动细菌鉴定仪 革兰阳性球菌 革兰阴性杆菌 Blood culture VITEK-2 Compact instrument Gram-positive cocci Gram-negative bacilli
  • 相关文献

参考文献8

  • 1黄磊,孙立颖,严岩.血培养阳性报警时间预测金黄色葡萄球菌血流感染患者的临床特征与预后[J].中华医院感染学杂志,2013,23(20):5107-5109. 被引量:6
  • 2Beekmann SE, Diekema DJ, Chapin KC, et al. Effects of rapid detec- tion of bloodstream infections on length of hospitalization and hospital charges[ J]. J Clin Microbiol, 2003,41 (7) :3119 - 3125.
  • 3郭建,吴文娟.血流感染分子诊断的研究进展[J].检验医学,2014,29(6):584-589. 被引量:22
  • 4Quesada MD, Gim:nez M, Molinos S, et al. Performance of VITEK - 2 Compact and overnight MicroScan panels for direct identification and susceptibility testing of Gram - negative bacilli from positive FAN BacT/ ALERT blood culture bottles[J]. Clin Microbiol Infect, 2010,16(2) : 137 - 140.
  • 5Ling TK, Liu ZK, Cheng AF. Evaluation of the VITEK 2 system for rapid direct identification and susceptibility testing of gram - negative bacilli from positive blood cultures [J]. J Clin Microbiol, 2003,41 (10) :4705 -4707.
  • 6Lupetti A, Barnini S, Morici P, et al. Saponin promotes rapid identifi- cation and antimicrobial susceptibility profiling of Gram - positive and Gram- negative bacteria in blood cultures with the Vitek 2 system [ J ]. Eur J Clin Microbiol Infect Dis, 2013,32(4) :493 -502.
  • 7Gherardi G, Angeletti S, Panitti M, et al. Comparative evaluation of the Vitek - 2 Compact and Phoenix systems for rapid identification and anti- biotic susceptibility testing directly from blood cultures of Gram - nega- tive and Gram -positive isolates[J]. Diagn Microbiol Infect Dis, 2012, 72(1) :20 -31.
  • 8De Cueto M, Ceballos E, Martinez -Martinez L, et al. Use of positive blood cultures for direct identification and susceptibility testing with the vitek 2 system[ J]. J Clin Microbiol, 2004,42(8 ) :3734 -3738.

二级参考文献35

  • 1Naber CK. Staphylococcusaureus bacteremia: epidemiology pathophysiology, and management strategies[J]. Clin Infect Dis,2009,48(Suppl 4) :2231-237.
  • 2Laupland KB, Ross T,Gregson DB. Staphylococcus aureus hloodstream infections: risk factors, outcomes, and the in- fluence of methicillin resistance in Calgary, Canada, 2000- 2006[J]. J Infect Dis,2008,198(3),336-343.
  • 3Marra AR, Edmond MB, Forbes BA,et a l. Time to blood cul- ture positivity as a predictor of clinical outcome of Staphylo- coccus aureus bloodstream infection[J]. J Clin Microbiol, 2006,44(4) : 13427-1346.
  • 4Kim J, Gregson DB, Ross T, et al. Time to blood culture posik tivity in Staphylococcus aureus bacteremia: association with 30-day mortality[J]. J Infect,2010,61(3) :197-204.
  • 5Alvarez R, Vinas-Castillo L, Lepe-Jimenez JA, et al. Time to positivity of blood culture association with clinical presenta- tion, prognosis and ESBLs-production in Escherichla coli bacteremia[J]. Eur J Clin Microbiol Infect Dis, 2012,31 (9) : 2191-2195.
  • 6Skvarc M, Stubljar D, Rogina P, et al. Non-culture- based methods to diagnose bloodstream infection: does it work [ J ]. Enr J nicrobiol Immunol ( Bp), 2013, 3(2) :97-104.
  • 7Afshari A, Schrenzel J, Ieven M, et al. Bench-to- bedside review: rapid molecular diagnostics for bloodstream infection-a new frontierr[ J ]. Crit Care,2012,16(3) :222.
  • 8Loonen AJ, de Jager CP, Tosserams J, et al. Biomarkers and molecular analysis to improve bloodstream infection diagnostics in an emergency care unit[J]. PLoS One, 2014,9(1):e87315.
  • 9Deck MK, Anderson ES, Buckner RJ, et al. Multieenter evaluation of the Staphylococcus QuickFISH method for simultaneous identification of Staphylococcus aureus and coagulase-negative staphylococci directly from blood culture bottles in less than 30 minutes[J]. J Clin Microbiol, 2012,50(6) : 1994-1998.
  • 10Forrest GN, Roghmann MC, Toombs LS, et al. Peptide nucleic acid fluorescent in situ hybridization for hospital-acquired enterococcal bacteremia: delivering earlier effective antimicrobial therapy [ J ]. Antimicrob Agents Chemother, 2005,52 (10) : 3558- 3563.

共引文献26

同被引文献19

引证文献3

二级引证文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部