BC-5000五分类血细胞分析仪性能评价

Performance evaluation of BC-5000 five categories hematology analyzer
下载PDF
导出
摘要 目的评价BC-5000五分类血细胞分析仪(BC-5000)的主要性能。方法对BC-5000的本底及空白值、携带污染率、精密度、准确度、正确度、线性范围及其检测结果与参比仪器(XE-2100)、人工分类的可比性进行评价。结果 BC-5000各参数的本底及空白值均合格,携带污染率均小于1%。所有参数均具有较好的批内和批间精密度,准确度及正确度验证合格。WBC、RBC、HGB、PLT具有良好的线性(r>0.990)。与XE-2100及人工分类结果进行对比,只有嗜碱性粒细胞百分比(Bas%)的相关性略差(r=0.744,P>0.05)。结论 BC-5000具有良好的性能,是一款理想的小型五分类血细胞分析仪。 Objective To evaluate the performance of BC-5000 five categories hematology analyzer(BC-5000).Methods The background and blank values,carryover rate,precision,accuracy,trueness,and linear range were evaluated,as well as the comparability of its results with reference instrument(XE-2100)detection and artificial classification of leukocytes.Results The background and blank values were both qualified.The carryover rates of all the detection items were less than 1%.The intra-and interprecision of all the detection items were good,and the accuracy and trueness were verified eligibility.The linearity of WBC,RBC,HGB,and PLT were good(r〉0.990).Comparing with XE-2100 and artificial classification,Bas% was the only one item with poor correlation coefficient(r=0.744,P〉0.05).Conclusion BC-5000 has good performance in hematology detection,and it is an ideal small five categories hematology analyzer.
出处 《国际检验医学杂志》 CAS 2014年第22期3116-3118,共3页 International Journal of Laboratory Medicine
关键词 血细胞分析仪 性能评价 血细胞计数 hematology analyzer performance blood cell count
  • 相关文献

参考文献8

二级参考文献29

  • 1梁洪璋.不同浓度抗凝剂对血细胞参数的影响[J].国际医药卫生导报,2003,9(22):39-40. 被引量:4
  • 2彭明婷,谷小林,陆红,陈文祥,申子瑜.我国血细胞分析参考系统的建立[J].中华检验医学杂志,2006,29(3):196-198. 被引量:33
  • 3卫生部临床检验中心血细胞分析溯源专家组.建立血细胞分析溯源体系的有关问题[J].中华检验医学杂志,2006,29(4):381-381. 被引量:13
  • 4何敏,敖小梅,李向饶.标本保存时间和温度对血细胞分析仪测定结果的影响[J].中华临床医学杂志,2006,7(9):17-18. 被引量:8
  • 5NCCLS (EP 9-P) [ S ]. User comparison of quantitative clinical laboratory using patient samples, Proposed Guideling (1986).
  • 6Cornbleet J. Spurious results from automated hematology cell counters[J]. Lab Med, 1983,4( 1 ) :509- 514.
  • 7Bain, Barbara J. Detecting erroneous blood counts, chapter 4 (132-146) of blood ceils -a practical guide-2nd Edition [ S ]. Blackwell Science. United Kingdom, 1995, 12.
  • 8卫生部临床检验中心,中国医院协会临床检验管理专业委员会.《医疗机构临床实验室管理办法》及配套文件[S].2006,9.
  • 9Barnes PW, McFadden SL, Machin SJ, et al. The International Group for Hematology Review: suggested criteria for action following automated CBC and WBC differential analysis [ J ]. Lab Hematol, 2005, 11(1): 83-90.
  • 10罗春丽.临床检验基础[M].北京:人民卫生出版社,2004:98-99.