Detector729在放射治疗剂量学中的应用研究 被引量:1

Application of Detector729 in radiotherapy dosimetry
下载PDF
导出
摘要 目的:探讨二维电离室矩阵探测器Detector729剂量测量的准确性及其在放射治疗质量控制中的应用。方法:采用0.6 cm3的参考电离室在X射线下对Detector729进行剂量刻度,然后两者进行剂量和输出因子测量比对,最后应用Detector729对调强放射治疗(intensity modulated radiation therapy,IMRT)治疗计划进行实际机架角剂量验证。结果:参考电离室对Detector729电离室的刻度因子是0.991,两者测量结果比对相对测量偏差≤0.57%,在4 cm×4 cm^20 cm×20 cm不同方形野下输出因子测量结果基本一致,应用Detector729对IMRT剂量验证的γ通过率均>90%。结论:应用Detector729进行放疗剂量测量和质量控制工作,结果具有较高的准确度和可信度,可为规范我国放疗剂量测量工具的使用提供理论依据。 Objective :To study accuracy of 2-D matrix Detector729 in dosimetric measurement and its application in radiotherapy quality control. Methods :Tests and output factor measurement were performed to evaluate and compare the accuracy of Detector729 taking 0.6 cm3 ion chamber under the X-ray as reference. Detector729 was used as a tool to verify dosimetric distribution for inten- sity modulated radiation therapy(IMRT). Results :The scale factor of Detector729 from the reference ionization chamber was 0.991, and the comparison results showed the maximum deviation of measuring point dose was 0.57%. Meanwhile, under the square fields of 4 cm × 4 cm-20 cm ~ 20 cm,the estimated output factor matched very well with the reference ion chamber measurements. Conclusion : With high accuracy and reliability, Detector729 can be used in dosimetric measurement and quality control of radiotherapy and pro- vide theoretical basis for the standard use of radiation dosimetric measurement instruments.
出处 《重庆医科大学学报》 CAS CSCD 北大核心 2014年第10期1474-1477,共4页 Journal of Chongqing Medical University
基金 国家自然科学基金面上资助项目(编号:81272511) 协和青年教师基金资助项目(编号:2012Y03中央高校基本科研业务费) 中国医学科学院放射医学研究所基金资助项目(编号:所开1427)
关键词 Detector729 剂量验证 剂量比对 输出因子 Detector729 dosimetric verification dosimetric comparison output factor
  • 相关文献

参考文献13

  • 1Low DA,Moran JM,Dempsey JF,et al.AAPM TG-120:Dosimetry tools and techniques for IMRT[J].Med Phvs, 2011,38(3) : 1313-1338.
  • 2翟贺争,程金生.二维矩阵探测器在IMRT剂量验证中的研究进展[J].中华放射医学与防护杂志,2013,33(2):215-217. 被引量:11
  • 3Spezi E,Angelini AL, Romani F,et al.Characterization of a 2D ion chamber array for the verification of radiotherapy treatments[J].Phys Med Biol, 2005,50(14) : 3361-3373.
  • 4Hussein M,Adams E J, Jordan TJ, et al.A critical evaluation of the PTW 2D-ARRAY seven29 and OCTAVIUS lI phantom for IMRT and VMAT verifieation[J].J Appl Clin Med Phys, 2013,14(6) :274-292.
  • 5Saminathan S,Maniekan R,Chandraraj V,et al.Dosimetrie study of 2D ion chamber array matrix for the modem radiotherapy treatment ver- ifieation[J].J Appl Clin Med Phys, 2010,11 (2) : 116-127.
  • 6国家质量监督检验检疫总局.JJG912-2010治疗水平电离室剂量计检定规程[S].北京,2010.
  • 7Low DA,James FD.Evaluation of the gamma dose distribution com- parison method[J].Med Phys, 2003,30(9) :2455-2464.
  • 8Fraass B,Doppke K,Hunt M,et al.AAPM TG-53:Quality assurance for clinical radiotherapy treatment planning[J].Med Phys, 1998 (25) : 1773-1829.
  • 9Agazarysa N,Solberg TD,DeMarco JJ.Patient specific quality assur- ance for the delivery of intensity modulated radiotherapy[J].J Applied Clinical Med Phys, 2003,4( 1 ) : 40-50.
  • 10Ieybovich LB, Sethi A, Dogan N.Comparison of ionization cham- bers of various volumes for IMRT absolute dose verification[J].Med Phys, 2003,30(2) : 119-123.

二级参考文献17

  • 1吴昊,蒋璠,韩树奎,包尚联.常见探测器的空间分辨率比较[J].中国医学物理学杂志,2011,28(2):2501-2504. 被引量:10
  • 2张惠玲,李勤,全红,韩军.MatriXX在IMRT剂量验证中的应用[J].中国医学物理学杂志,2011,28(2):2455-2459. 被引量:16
  • 3周振山,于广辉,乔艳萍.调强放射治疗计划的剂量验证[J].中华放射医学与防护杂志,2005,25(1):69-70. 被引量:5
  • 4Agazaryan N, Solberg TD, DeMarco JJ. Patient specific quality assurance for the delivery of intensity modulated radiotherapy [ J ]. J App Clin Med Phys ,2003 ,4(1 ) :40-50.
  • 5Dong L, Antolak J, SalehpourM, et al. Patient - specific point dose measurement for IMRT monitor unit verification [ J ]. Int J Radiat Oncol Biol Phys, 2003, 56 (3) :867 - 877.
  • 6Sulman EP, Schwartz DL, Le TT, et al. IMRT reirradiation of head and neck cancer- disease control and morbidity outcomes [ J]. Int J Radiat Oncol Biol Phys ,2009,73 ( 2 ) :399 - 499.
  • 7Vincent G, Wilfried DN, Avraham E, et al. Intensity - modulated radiation therapy for head and neck carcinoma [ J]. Oncologist, 2007,12(5) :555 -564.
  • 8Leonid B Leybovieh, Anil Sethi, Nesrin Dogan. Comparison of ionization chambers of various volumes for IMRT absolute dose verification[J] . Med Phys,2003,30(2) :119 -123.
  • 9Zhu X.R, Jursinic PA, Grimm DF, et al. Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a static muhileaf collimator[ J]. Med Phys, 2002,29 (8) :1687 - 1692.
  • 10谢耩,王运来,马林,徐寿平,戴相昆.电离室灵敏体积对调强放射治疗剂量学验证准确性的影响[J].中华放射医学与防护杂志,2007,27(6):582-584. 被引量:7

共引文献12

同被引文献6

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部