摘要
目的:探讨二维电离室矩阵探测器Detector729剂量测量的准确性及其在放射治疗质量控制中的应用。方法:采用0.6 cm3的参考电离室在X射线下对Detector729进行剂量刻度,然后两者进行剂量和输出因子测量比对,最后应用Detector729对调强放射治疗(intensity modulated radiation therapy,IMRT)治疗计划进行实际机架角剂量验证。结果:参考电离室对Detector729电离室的刻度因子是0.991,两者测量结果比对相对测量偏差≤0.57%,在4 cm×4 cm^20 cm×20 cm不同方形野下输出因子测量结果基本一致,应用Detector729对IMRT剂量验证的γ通过率均>90%。结论:应用Detector729进行放疗剂量测量和质量控制工作,结果具有较高的准确度和可信度,可为规范我国放疗剂量测量工具的使用提供理论依据。
Objective :To study accuracy of 2-D matrix Detector729 in dosimetric measurement and its application in radiotherapy quality control. Methods :Tests and output factor measurement were performed to evaluate and compare the accuracy of Detector729 taking 0.6 cm3 ion chamber under the X-ray as reference. Detector729 was used as a tool to verify dosimetric distribution for inten- sity modulated radiation therapy(IMRT). Results :The scale factor of Detector729 from the reference ionization chamber was 0.991, and the comparison results showed the maximum deviation of measuring point dose was 0.57%. Meanwhile, under the square fields of 4 cm × 4 cm-20 cm ~ 20 cm,the estimated output factor matched very well with the reference ion chamber measurements. Conclusion : With high accuracy and reliability, Detector729 can be used in dosimetric measurement and quality control of radiotherapy and pro- vide theoretical basis for the standard use of radiation dosimetric measurement instruments.