双源螺旋CT大螺距Flash模式在肺动脉CTA成像技术中的应用 被引量:5

Application of dual-source spiral CT with high-pitch flash mode in CT pulmonary angiography
下载PDF
导出
摘要 目的:探讨双源螺旋CT大螺距Flash模式、时间-密度曲线测试技术在肺动脉CT成像(computer tomography pulmonary angiography,CTPA)中运用的可行性。方法:将拟行CTPA检查的60例患者随机分为2组。对照组采用单源常规模式扫描,对比剂示踪技术。实验组采用双源大螺距Flash模式扫描,时间-密度曲线测试技术。对比剂均采用400 mg I/ml碘迈伦,其余扫描参数均相同。扫描结束后,将原始数据传至后处理工作站进行处理,分别采用图像质量评分及CT值、对比噪声比评价图像质量。结果:实验组及对照组共60例病人均成功完成CTPA检查,2组一般资料如性别、年龄、身高、体质量均无明显统计学差异(P值分别为0.752,0.605,0.319,0.523);实验组对比噪声比高于对照组(P值分别为0.028,0.030,0.001,0.012,0.020),2组肺动脉干、左右上肺动脉、左右下肺动脉CT值均无统计学差异(P值分别为0.362,0.986,0.373,0.584,0.931),实验组升主动脉、左右上肺静脉、左右下肺静脉CT值均低于对照组,差异有统计学意义(P值分别为0.000,0.001,0.000,0.003,0.000),且实验组的图像质量评分优于对照组(χ2=12.25,P=0.016;χ2=7.53,P=0.023);实验组对比剂用量(24 ml vs.50 ml)和辐射剂量[(5.27±1.94)m Sv vs.7.30±2.73)m Sv]低于对照组(P=0.011),实验组扫描时间明显短于对照组[(0.94±0.07)s vs.(4.71±0.37)s,P=0.000]。结论:在CTPA检查中,双源CT大螺距Flash模式、时间-密度曲线测试技术不仅能采集到理想的肺动脉影像,而且降低了患者的辐射剂量及对比剂用量,还有效缩短了患者的屏气时间。 Objective:To explore the feasibility of dual-source spiral CT with high-pitch flash spiral mode as well as time-density curve in diagnosing the clinically suspected pulmonary embolism (PE). Methods:Sixty patients planed to receive CT pulmonary an- giography(CTPA) were randomly divided into two groups. Routine scan and bolus tracking monitoring were used in control group,while dual-source spiral CT with high-pitch flash spiral mode as well as time-density curve was used in experimental group. Contrast agents Iodine diftler with the concentration of 400 mgI/ml was applied in both groups. The parameter of CT scanning was the same in both groups. All CT original dates were transferred to Siemens workstation and were post-processed by VR, MIP, MPR and CPR tech- nology. The CTPA image quality was evaluated based on score of image quality, CT value, and contrast to noise ratio (CNR). Results: All patients in two groups completed CTPA successfully. There was no difference in baseline characters including sex, age, height and weight between the two groups (P=0.752,0.605,0.319,0.523). CNR of experimental group was higher than that of control group (P= 0.028,0.030,0.001,0.012,0.020). There was no difference in CT value of pulmonary artery trunk, RSPA, RIPA, LSPA and LIPA between the two groups (P=0.362,0.986,0.373,0.584,0.931 ), but CT value of the ascending aorta, RSPV, RIPV, LSPV and LIPV were lower in experimental group than in control group (P=0.000,0.001,0.000,0.003,0.000). Scores of image quality were better in experimental group than in control group0(x^2=12.25 ,P=0.016;X^2=7.53 ,P=0.023). Both the contrast agents(24 ml vs. 50 ml) and radi- ation dose[(5.27 ± 1.94) mSv vs. (7.30 ± 2.73) mSv] were lower in experimental group than in control group(P=0.011). The scanning time was significantly shorter in experimental group than in control group[(0.94±0.07) s vs. (4.71 ± 0.37) s,P=0.000)]. Conclusion:CTPA with flash spiral mode and time-density curve could col- lect the satisfied image, reduce patients' contrast agents and radiation dose, and decrease the breathless time.
出处 《重庆医科大学学报》 CAS CSCD 北大核心 2014年第10期1494-1499,共6页 Journal of Chongqing Medical University
基金 国家临床重点专科建设资助项目
关键词 肺动脉 X线计算机 对比剂 血管造影术 pulmonary artery X-ray computer contrast agent angiography
  • 相关文献

参考文献9

二级参考文献52

共引文献117

同被引文献22

引证文献5

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部