Flash双源CT低剂量扫描在冠脉成像中的应用 被引量:2

Low-Dose Scan Protocols in Dual-Source FLASH CT Coronary Angiography
下载PDF
导出
摘要 目的:比较双源FLASH CT不同扫描方案在冠脉成像中图像质量和辐射剂量的差别。材料与方法:对需做冠脉CT心率小于等于65次/分,心律整齐患者118例,随机分为A、B两组,分别采用Flash模式与前瞻性心电门控触发模式进行冠脉扫描,盲法评价分析A、B两组患者冠脉图像质量及X线辐射剂量。结果:Flash扫描模式下扫描与前门控触发扫描所得具有诊断价值的血管节段数和平均图像质量评分无显著统计学差异。Flash模式下扫描辐射剂量显著低于前瞻性心电门控触发模式。结论:Flash扫描模式能显著降低患者冠脉CTA的辐射剂量。 Purpose: To compare image quality and radiation dose using different low-dose scan protocols in dualsource FLASH CT coronary angiography. Materials and Methods: One hundred eighteen patients (67 men, 51women) with heart rate less than 65 beats per minute were randomized to divide into group A and group B. In group A, CT was performed using FLASH mode; In group B, CT was performed using the prospectively electrocardiography (ECG)-gated mode. The image quality and radiation in all coronary segments were evaluated by two blinded and independent observers. Resnlts: Diagnostic image quality was obtained 875 segments in group A and 878 segments in group B. Mean image quality grades showed no significant difference between two groups. Mean estimated radiation dose was signi? cantly lower in group A than in group B. Conclusion: A high-pitch spiral CT coronary angiographic protocol will reduce the radiation dose in patients with heart rates less than 65 beats per minute.
出处 《现代医用影像学》 2014年第5期497-499,共3页 Modern Medical Imageology
关键词 FLASH 辐射剂量 冠脉CTA Flash Radiation dose CT coronary angiography
  • 相关文献

参考文献4

二级参考文献24

  • 1张森,杜湘珂,李剑颖.心电图调制电流对于容积CT心脏扫描图像质量及曝光剂量影响的评估[J].中华放射学杂志,2006,40(9):974-976. 被引量:45
  • 2Trabold T, Buchgeister M, Kuttner A, et al. Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Rofo, 2003, 175 : 1051-1055.
  • 3Mallet NR, Cademartiri F, van Mieghem CA, et al. High- resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation, 2005,112 : 2318-2323.
  • 4Flohr TG, Schoepf U J, Kuettner A, et al. Advances in cardiac imaging with 16-section CT systems. Acad Radiol, 2003, 10: 386-401.
  • 5Hsieh J, Londt J, Vass M, et al. Step-and-shoot data acquisition and reconstruction for cardiac X-ray computed tomography. Med Phys, 2006,33:4236-4248.
  • 6Leschka S,Stolzmann P,Desbiolles L,et al.Diagnostic accuracyof high-pitch dual-source CT for the assessment of coronary ste-noses:first experience[J].Eur Radiol,2009,19:2896-2903.
  • 7Austen WG,Edwards JE,Frye RL,et al.A reporting system on pa-tients evaluated for coronary artery disease.Report of the Ad HocCommittee for Grading of Coronary Artery Disease,Council on Car-diovascular Surgery,American Heart Association[J].Circulation,1975,51:5-40.
  • 8Achenbach S,Anders K,Kalender WA.Dual-source cardiac com-puted tomography:image quality and dose considerations[J].EurRadiol,2008,18:1188-1198.
  • 9Achenbach S,Marwan M,Ropers D,et al.Coronary computedtomography angiography with a consistent dose below 1mSv u-sing prospectively electrocardiogram-triggered high-pitch spiralacquisition[J].Eur Herat J,2010,31:340-346.
  • 10Blankstein R,Shah A,Pale R,et al.Radiation dose and imagequality of prospective triggering with dual-source cardiac compu-ted tomography[J].Am J Cardiol,2009,103:1168-1173.

共引文献202

同被引文献16

引证文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部