摘要
目的:建立盐酸雷洛昔芬及其片剂中有关物质测定的HPLC法,对其主要杂质A、C和D进行定量研究。方法:采用Venusil XBP C8色谱柱(250 mm×4.6 mm,5μm),以0.066 mol·L^-1磷酸二氢钾缓冲液(磷酸调节p H至3.0)为流动相A,乙腈为流动相B,梯度洗脱,流速1.0 m L·min-1,检测波长280 nm,对有关物质进行定性、定量分析。结果:在选定的色谱条件下,盐酸雷洛昔芬与各杂质分离良好,杂质A、C和D的检测限分别为21.3、23.3和23.0 ng,质量浓度分别在0.532-213.0、0.582-233.0和0.538-215.0μg·m L^-1范围内与峰面积呈良好的线性关系。结论:本方法简便、灵敏,专属性好,线性范围宽,可用于盐酸雷洛昔芬及其片剂的有关物质检测和质量控制。
Objective:To establish an HPLC method for the determination of impurities in raloxifene hydrochloride and its tablets. Methods: The separation was performed on a Venusil XBP C8 column (250 mm × 4. 6 mm, 5 μm) with a mobile phase consisting of a mixture of acetonitrile (mobile phase B) and 0. 066 mol · L^-1 potassium dihydro- gen phosphate buffer ( mobile phase A, pH 3.0) by gradient elution at a flow rate of 1.0 mL ·min ^- 1 The detective wavelength was 280 nm. Results: Raloxifen hydrochloride and its impurities were well separated under the selected liquid chromatographic system. The detection limits of impurity A, C and D were 21.3 ng,23.3 ng,23.0 ng and the calibration curves revealed good linearities over the ranges of 0. 532 -213.0 μg · mL^-1, 0. 582 -233.0 μg ·mL^- 1 and 0. 538 -215.0 μg ·mL^-1. respectively. Conclusion: The method is easy, sensitive, and specific which can be used for the test of the related substances of raloxifene hydrochloride and its tablets and quality control.
出处
《药物分析杂志》
CAS
CSCD
北大核心
2014年第12期2192-2197,共6页
Chinese Journal of Pharmaceutical Analysis