样本收集和保存对血液透析患者iPTH检测的影响 被引量:1

Influence of sample collection and storage on detection of iPTH in patients with hemodialysis
下载PDF
导出
摘要 目的评估样本收集和保存对全段甲状旁腺素(iPTH)检测结果的影响。方法 10份血液透析患者血液样本同时注入4种采血管:促凝管、空白管、EDTA管及枸橼酸钠管,采用罗氏cobase411免疫分析仪分别检测在采血后4h内,以及4℃保存12、24、48h的血清iPTH水平。结果采血4h内促凝管血清iPTH快速发生降解,而空白管中iPTH水平最高,且明显高于12、24、48h。而在EDTA管和枸橼酸钠管,iPTH水平在48h才会开始降解。结论首次揭示了促凝剂对血液透析患者血清iPTH有降解或抑制作用,进一步证实EDTA管是iPTH最佳的采血管。 Objective To evaluate the influence of sample collection and storage on the detection results of intact parathyroid hormone(iPTH). Methods 10 blood samples from the hemodialysis patients were simultaneously injected into four types of blood collection tube:pro-coagulation tube, blank tube, EDTA tube and sodium citrate tube. The iPTH levels in the fresh samples within 4 h after blood collection and serum/plasma preserving for 12,24,48 h under 4OC were detected by the Roche cobase411 immunoassay instrument. Results Serum iPTH in the pro-coagulation tube within 4 h after blood collection was rapidly degraded,while the iPTH level in the blank tube was highest, which was significantly higher than that the serum/plasma preserved for 12,24,48 h. The iPTH level in the EDTA tube and the sodium citrate tube began to degrade only at 48 h. Conclusion It is the first time to reveal that the coagulant has the degrading or inhibiting effect on serum iPTH, which verifies that the EDTA tube is the optimal blood collection tube.
作者 荀春华
出处 《国际检验医学杂志》 CAS 2015年第1期98-99,共2页 International Journal of Laboratory Medicine
关键词 全段甲状旁腺素 样本收集 电化学发光 intact parathyroid hormone sample collection electrochemiluminescence
  • 相关文献

参考文献15

  • 1Holland NT,Smith MT,Eskenazi B, et al. Biological sample col- lection and processing for molecular epidemiological studies[J].Murat Res,2003,543(2):217-234.
  • 2Smith JJ, Lewis S, Holbrook J, et al. Effect of heparin and citrate on measured concentrations of various analytes in plasma[J]. Clin Chem,1987,33(1) :814-816.
  • 3Tammen H, Schulte I, Hess R, et al. Peptidomic analysis of human blood specimens:comparison between plasma specimens and ser- um by differential peptide display[J]. Proteomics, 2005 (5) :3414- 3422.
  • 4Haab BB, Geierstanger BH, Michailidis G, et al. Immunoassay and antibody microarray analysis of the HUPO Plasma Proteome Pro- ject reference specimens: systematic variation between sample types and calibration of mass spectrometry data[J]. Proteomies, 2005,5 (13) : 3278-3291.
  • 5Lopez JB, Peng CL. Can fluoride-oxalate and sodium citrate sta- bilise homocysteine levels after blood collection [J]. Clin Chem Lab Med. 2003,41(1) : 1369-1372.
  • 6Sonia L, La' ulu, William L. Roberts. Performance Characteristics of Six Intact Parathyroid Hormone Assays[J]. Am J Clin Pathol, 2010,134(1) : 930-938.
  • 7Tepel M, Armbruster FP, Gron HJ, et al. Nonoxidized, biological- ly active parathyroid hormone determines mortality in hemodialy- sis patients [J]. J Clin Endoerinol Metab, 2013, 98 (12) : 4744- 4751.
  • 8Nishida H, Masakane I, Kudo K, et al. Chronic hemodialysis pa- tients without marked elevation of intact parathyroid hormone are also good candidates for early intervention with cinacalcet[J]. T- her Apher Dial, 2013,17 (3) : 325-331.
  • 9Kinnaert P,Tielemans C, Dhaene M, et al. Evaluation of surgical treatment of renal hyperparathyroidism by measuring intact parathormone blood levels on first postoperative day[J]. World J Surg, 1998,22(7) : 695-699.
  • 10Bieglmayer C,Prager G, Niederle B. Kinetic analyses of parathy- roid hormone clearance as measured by three rapid immunoassays during parathyroidectomy [J]. Clin Chem, 2002,48 (10) : 1731- 1738.

同被引文献1

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部