对解剖倾斜位X线平片判断桡骨远端螺钉穿破关节方法的评估 被引量:2

Evaluation of Anatomic Tilt X-Ray Films in Judging Intra-Articular Screw Penetration of Distal End of Radius
原文传递
导出
摘要 目的评估解剖倾斜位X线平片判断桡骨远端螺钉穿破关节方法的可靠性、可重复性和诊断价值。方法选择在标准正侧位X线平片显示可疑桡骨远端螺钉穿破关节的31例患者行腕关节CT扫描和解剖倾斜位X线平片检查。CT扫描螺钉位置作为标准,选择两位观察者依据解剖倾斜位X线平片,采用双盲法独立判断螺钉位置,计算判断的敏感性、特异性和预估值;采用Kappa系数计算一致性。结果 CT扫描未穿破17例,穿破14例。两位观察者首次判断螺钉未穿破的敏感性分别为94.12%和88.24%,特异性分别为87.50%和93.33%,预估值均为100%;判断螺钉穿破的敏感性分别为93.33%和87.50%,特异性分别为106.25%和93.75%,预估值分别为93.33%和87.50%。两位观察者之间的Kappa系数为0.81-0.94,具有高度可靠性;二次重复判断之间的Kappa系数均为0.87,具有高度可重复性。结论解剖倾斜位X线平片判断桡骨远端螺钉穿破关节具有高度的可靠性、可重复性和较高的诊断价值。 Objective To evaluate the reliability, repeatability and diagnostic value of anatomic tilt X-ray films in judging intra-articular screw penetration of distal end of radius. Methods A total of 31 patients, in whom the distal radius screws were suspected to penetrate into the joint on standard AP and lateral films, were enrolled in this study. CT scanning and anatomic tilt X-ray films of the wrist joint were performed in all the patients. Taking the screw position displayed on CT scans as the reference standard and using double-blind method, the anatomic tilt X-ray films of the wrist joint were independently analyzed by two radiologists to judge the screw position. The sensitivity, specificity and predictive value were calculated, and the consistency between two radiologists was evaluated using Kappa coefficient analysis. Results CT scans indicated that no intra-articular screw penetration was detected in 17 cases and intra-articular screw penetration was observed in 14 cases. The sensitivity, specificity and predictive value for the absence of intra-articular screw penetration judged by two radiologists were 94.12% and 87.50% ; 87.50% and 93.33% ; 100% and 100% respectively, while the sensitivity, specificity and predictive value for the presence of intra-articular screw penetration judged by two radiologists were 93.33% and 87.50% ; 106.25% and 93.75% ; 93.33% and 87.50% respectively. The intraobserver Kappa coefficient of initial judgment was 0.81 -0.94, indicating the presence of a higher reliability. The intra-observer Kappa coefficient of second judgment was 0.87, indicating the presence of higher repeatability. Conclusion Anatomic tilt X-ray films has higher reliability and repeatability in judging the presence or absence of intra-articular screw penetration, therefore it has great diagnostic value.
出处 《临床放射学杂志》 CSCD 北大核心 2014年第12期1931-1934,共4页 Journal of Clinical Radiology
关键词 X线 桡骨远端骨折 螺钉位置 解剖倾斜位 X-ray Distal radius fractures Screw position Anatomic tilt position
  • 相关文献

参考文献8

  • 1Boyer MI, Korcek K J, Gelberman RH, et al. Anatomic tilt X-rays of the distal radius : an ex vivo analysis of surgical fixation. J Hand Surg Am ,2004,29 : 116.
  • 2Soong M, Got C, Katarincic J, et al. Fluoroscopic evaluation of intra- articular screw placement during locked volar plating of the distal ra- dius:a cadaveric study. J Hand Surg Am,2008,33:1720.
  • 3Kumar D, Breakwell L, Deshmukh SC, et al. Tangential views of the articular surface of the distal radius-aid to open reduction and inter- nal fixation of fractures. Injury,2001,32:783.
  • 4Pace A, Cresswell T. Use of articular wrist views to assess intra-artic- ular screw penetration in surgical fixation of distal radius fractures. J Hand Surg Am,2010,35:1015.
  • 5唐涛,马霄虹,王安庆,赵利,冯建璞,刘克敏,闫家智,刘四海,王飞,崔志刚.不同腕关节影像学位置成像效果比较[J].中国康复理论与实践,2006,12(6):529-530. 被引量:4
  • 6Matullo KS, Dennison DG. Lateral tilt wrist radiograph using the con- tralateral hand to position the wrist after volar plating of distal radius fractures. J Hand Surg Am,2010,35:900.
  • 7Waris E, Paavola M. Current treatment of distal radius fractures. Duodecim ,2012,128:386.
  • 8Matthew L,Tweet MD, Ryan P, et al. Rotational Fluoroscopy Assists in Detection of Intra-Articular Screw Penetration During Volar Plat- ing of the Distal Radius. J Hand Surg Am,2010,35:619.

二级参考文献10

  • 1Hastings H Ⅱ,Leibovic SJ.Indications and techniques of open reduction.Internal fixation of distal radius fractures[J].Orthop Clin North Am,1993,24:309-326.
  • 2Fernandez DL.Should anatomic reduction be pursued in distal radial fractures?[J].J Hand Surg,2000,25B:1-6.
  • 3Kambouroglou GK,Axelrod TS.Complications of the AO/ASIF titamium distal radius plate system (pi plate) in internal fixation of the distal radius:a brief report[J].J Hand Surg Am,1998,23:737-741.
  • 4Hutchinson ST,Strenz GO,Cautilli RA.Pin and plaster vs.external fixation in the treatment of unstable distal radial fractures.A randomized prospective study[J].J Hand Surg,1999,20(3):365-372.
  • 5Werber SC,Szabo RM.Severely comminuted distal radial fractures as an unsolved problem:complications associated with external fixation and pins and plaster techniques[J].J Hand Surg,1986,11A:157-165.
  • 6Cooney WP,Dobyns JH,Linscheid RL.Complications of Colle's fracture[J].J Bone Joint Surg,1980,62A:613-619.
  • 7Kaempffe FA,Wheeler DR,Peimer CA,et al.Severe fractures of the distal radius:Effect of amount and duration of external fixator distraction on outcome[J].J Hand Surg,1993,18A:33-41.
  • 8Sander RA,Keppel FL,Waldrop JL.External fixation of distal radius fractures:Results and complications[J].J Hand Surg,1991,16A:1010-1016.
  • 9Cooney WP,Linscheid RL,Dobyns JH.Fractures and dislocations of the wrist[M]//Rockwood CA,Green DP,Bucholz RW,et al.ed.Fractures in Adults.Philadelphia.PA:Lippincott Raven,1996:745-868.
  • 10王启华,孙博.四肢分册[M]//临床解剖学丛书.人民卫生出版社,1991:9-14.

共引文献3

同被引文献8

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部