摘要 PET/CT成像中的人体呼吸运动会造成图像运动模糊,会严重影响图像质量,对医生的诊断造成影响。目前常用的呼吸门控技术能够在一定程度上改善图像质量,但是均存在其局限性。提出了一种基于CT图像提取呼吸运动特征的B样条方法,对呼吸运动图像进行校正。在PET/CT上获取与PET图像周期匹配的CT图像序列,通过B样条配准方式对在呼吸周期内的CT图序列像提取运动信息;对CT图像所匹配的PET图像做基于运动特征信息的校正变换进行运动校正。几何位移形变和像素人体模实验结果表明,提出的方法对PET/CT呼吸运动图像质量改进明显,具有研究价值。 The degradation of image quality of PET/CT caused by respiratory motion will affect the physician' s diagnosis. The common developed technology of respiratory motion correction is gating, but it still has some limitations. This paper proposes a new method that using CT images extract the features of respiratory motion based on B-Spline to correct respi- ration. Firstly, it obtains the sequence of CT images corresponding with PET images within same respiratory motion cycle, and extracts the features of the motion of CT sequences within respiratory cycle by B-Spline. Next it transforms the feature parameters of CT sequences into the corresponding PET image sequences for the motion correction. The geometric defor- mation phantom and voxelised phantom tests show that the proposed method can obviously improve the quality of image of PET/CT for respiratory motion, and has a value of study further.
出处 《计算机工程与应用》 CSCD 北大核心 2015年第1期156-161,共6页 Computer Engineering and Applications
基金 国家自然科学基金(No.11265007) 教育部留学回国人员科研启动基金(No.2010-1561)
关键词 PET/CT成像 呼吸运动校正 B样条 PET/CT imaging respiratory motion correction B-spline
  • 相关文献

参考文献22

  • 1许全盛,袁克虹,于丽娟,王文志,叶大田.PET/CT图像呼吸运动伪影校正研究进展[J].中国生物医学工程学报,2009,28(4):573-580. 被引量:22
  • 2Schaerer1 J,Fassi A,Riboldi M,et al.Multi-dimensional respiratory motion tracking from markerless optical surface imaging based on deformable mesh registration[J].Physics in Medicine and Biology,2012,57:357-373.
  • 3Chang G,Chang T,Pan T,et al.Implementation of an automated respiratory amplitude gating technique for PET/CT:clinical evaluation[J].Journal of Nuclear Medicine,2010,51(1):16-24.
  • 4Chamberland M,Wassenaar R,Spencer B,et al.Performance evaluation of real-time motion tracking using positron emission fiducial markers[J].Medical Physics,2011,38(2):810-819.
  • 5Kesnera A,Kuntner C.A new fast and fully automated software based algorithm for extracting respiratory signal from raw PET data and its comparison to other methods[J].Medical Physics,2010,37(10):5550-5559.
  • 6Kesner A,Bundschuh R.Respiratory gated PET derived in a fully automated manner from raw PET data[J].IEEE Transactions on Nuclear Science,2009,56(3):677-686.
  • 7Buther F,Dawood M,Stegger L,et al.List mode-driven cardiac and respiratory gating in PET[J].The Journal of Nuclear Medicine,2009,50(5):674-681.
  • 8He J,G.Keefe O,Gong S,et al.A novel method for respiratory motion gated with geometrical sensitivity of the scanner in3D PET[J].IEEE Transactions on Nuclear Science,2008,55(5):2557-2565.
  • 9Schleyer P,Doherty M,Marsden P.Extension of a datadriven gating technique to 3D,whole body PET studies[J].Physics in Medicine and Biology,2011,56:3953-3965.
  • 10Nehmeh S,Erdi Y.Respiratory motion in positron emission tomography/computed tomography:a review[J].Seminars in Nuclear Medicine,2008,38(3):167-176.

二级参考文献54

  • 1Pan Tinsu, Mawlawi O, Nehmeh S, et al. Attenuation correction of PET images with respiration-averaged CT image in PET/CT [ J ]. Nuclear Medicine, 2005, 36(6) : 1481 - 1487.
  • 2王学明,沈克涵.医学成像系统[M].北京:清华大学出版社.2006.
  • 3Daou D. Respiratory motion handling is mandatory to accomplish the high-resolution PET destiny [J]. European Journal of Nuclear Med Mol Imaging, 2008, 35:1961- 1970.
  • 4Visvikis D, Lamare F, Bruyant P, et al. Respiratory motion in positron emission tomography for oncology applications: Problems and solutions [ J]. Nuclear Instruments and Methods in Physics Research, 2006, A569: 453- 457.
  • 5Chuang Keh-Shih, Chen Tzong-Jer, Chang Chia-Chi, et al. Reduction of motion artifacts for PET imaging by respiratory correlated dynamic scanning [J]. Nuclear Instruments and Methods in Physics Research, 2006, A569: 458- 461.
  • 6Zhu Zhiyu, Tsui BMW, Segars WP. A Simulation Study of the Effect of Gating Scheme on respiratory Motion Blurting in FDG Lung PET [J]. IEEE Transactions on Nuclear Science, 2003, 13(3): 1554 - 1558.
  • 7Segars WP. Development of a new dynamic NURBS-based cardiactorso (NCAT) phantom [ D]. North Carolina: University of North Carolina, 2001.
  • 8Thomdyke B, Koong A, Xing L. Reducing respiratory motion artifacts in radionuclide imaging through retrospective stacking: A simulation study [J]. Linear Algebra and ils Applications, 2008, A428:1325 - 1344.
  • 9Dawood M, Lang N, Jiang Xiaoyi, et al. Lung Motion Correction on Respiratory Gated 3-D PET/CT Images [ J ]. IEEE Transactions on Medical Imaging, 2006, 25(4): 476- 485.
  • 10Asma E, Manjeshwar R, Thielemans K. Theoretical comparison of motion correction techniques for PET image reconstruction [ A ]. In: 2006 IEEE Nuclear Science Symposium Conference Record [ C ]. San Diego: INSPEC, 2006. 1762- 1767.

共引文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部