摘要 在机械加工中测量切削温度有多种方法,其中红外辐射测温法是通过测量出物体的辐射能来计算得出物体的温度方法,这种测量方法需要知道被测量物体的辐射率。在机械加工中,很难获得工件材料和刀具材料的准确辐射率,所以在测量切削温度的过程中如何避免辐射率对测温结果的影响很重要。文中利用光纤型红外测温仪,通过比值处理In As和In Sb的辐射能的方法巧妙滤掉了辐射率的影响,并基于新型标定装置最终处理得出了红外光纤测温仪的电势-温度关系曲线。 There are a variety of ways to measure the cutting temperature in machining. Infrared radiation temperature measurement can measure the object radiation for calculating the temprature of the object, this method needs to know the emissivity of the object to be measured. In machining,it's difficult to obtain the accurate emissivity of the workpiece material and tool material, so it is very important to avoid the effects of radiation on temperature results in process of cutting. By using the fiber infrared thermodetector, ratio processing of InAs and InSb radiation is used to filter out the effects of radiation rate clever, and the new calibration device is used to obtain the thermometer-temperature relationship curve of the fiber infrared thermodetector.
作者 郭卡 曹春风
出处 《机械工程师》 2015年第1期54-56,共3页 Mechanical Engineer
关键词 辐射率 红外光纤 测温仪 比值处理 电势-温度曲线 emissivity infrared fiber thermodetector ratio processing potential-temperature curve
  • 相关文献

参考文献8

  • 1周庆福,杨永军,吕国义.红外辐射测温仪及校准方法探讨[J].计测技术,2008,28(B10):33-35. 被引量:11
  • 2黄长生.快速杯定装置的改进及多种材料的切削温度实验研究[D].哈尔滨:哈尔滨工业大学,1994.
  • 3黄宏伟.红外测温仪检定方法探讨[J].铁道技术监督,2004,32(9):35-35. 被引量:4
  • 4许开设,朱平.红外测温仪的校准方法[J].工业计量,2001,11(S1):251-252. 被引量:7
  • 5周智鵬.小径ボールエンドミルによる高速加工[D].日本金泽:金泽大学.2004.
  • 6細川晃,周智鵬小径ボールエンドミルによる高速ミロリングに関する研究[J].精密工学会誌,2004,70(12):1527-1532.
  • 7細川晃,小田佳作,山口启司等.断続切削における逃け面温度-高速エンドミルに関する研究(第一報)[J].精密工学会誌, 2000,66( 11) : 1786.
  • 8上田隆司,细川晃..光ファイパを用いた赤外線輔射温度計の開[J].材料,1987(36):404-409.