摘要 目的通过对磁共振常规T2WI施加驱动平衡技术,探究该技术在保证图像对比度和图像质量的前提下能否显著缩短扫描时间、加快扫描速度。方法随机选取10名志愿者对其踝关节进行磁共振检查。每名志愿者进行3种T2WI扫描,分别为常规T2WI扫描(TR=4250 ms),短TR时T2WI扫描(TR=2500 ms),短TR时且施加驱动平衡的T2WI扫描(TR=2500 ms)。分析各种扫描图像的软组织对比度和图像质量。结果的部分未复制上。结论在T2WI中施加驱动平衡技术,可在保证图像质量的前提下缩短扫描时间,值得推广。 Objective To test whether applying the driven equilibrium technology to MRI T2 weighted image(T2WI) can obviously reduce scanning time and improve scanning speed on the premise of ensuring preferable image contrast and image quality. Methods Ten volunteers were selected randomly to undergo MRI examinations on their knees. All the volunteers received three kinds of T2 WI scan: conventional T2 WI scan(TR=4250 ms), T2 WI scan with short TR(TR=2500 ms), T2 WI scan with driven equilibrium technology and short TR(TR=2500 ms). Then the soft-tissue contrasts and quality of images were contrastively analyzed. Results With the application of driven equilibrium technology, the contrast and quality of images could be improved and the scanning time could be significantly reduced. Conclusion Applying driven equilibrium technology to T2 WI can reduce scanning time on the premise of ensuring image quality, which is worth widely promotion.
出处 《中国医疗设备》 2015年第1期60-62,共3页 China Medical Devices
关键词 踝关节 驱动平衡技术 T2WI 图像对比度 ankle joint driven equilibrium technique T2 weighted image image contrast
  • 相关文献

参考文献10

二级参考文献15

  • 1林克江,尤启冬,林东海,吴梧桐,叶波平.药物设计中的生物核磁共振技术[J].中国药科大学学报,2004,35(5):387-392. 被引量:2
  • 2陈彬,孔继烈.天然产物结构分析中质谱与核磁共振技术应用新进展[J].化学进展,2004,16(6):863-870. 被引量:7
  • 3王进贤,席永盛.^(13)C化学位移与定位效应[J].西北师范大学学报(自然科学版),1997,33(1):101-103. 被引量:8
  • 4[6]JohnCavangh[美],等.蛋白质核磁共振波谱学原理与应用[M].北京:科学出版社,2008.
  • 5Rothpearl A,Frager D,Subramanian A,et al.MR urography:technique and application[J].Radiology,1995,194(1):125-30.
  • 6Hennig J,Friedburg H,and Strobel B.Rapid nontomographic approach to MR myelography without contrast agents[J].J Comput Assist Tomogr,1986,10(3):375-378.
  • 7O'Malley M E,Soto J A,Yucel E K,et al.MR urography:evaluation of a three-dimensional fast spin-echo technique in patient with hydronephrosis[J].Am J Roentgenol,1997,168(2):387-392.
  • 8Kawashima A,Glockner J F,King B F Jr.CT urography and MR urography[J].Radiol Clin North Am,2003,41(5):945-961.
  • 9Nolte-Ernsting CC,Staatz G,Tacke J,et al.MR urography today[J].Abdom Imaging,2003,28(2):191-209.
  • 10Ernst R, Bodenhausen G, Wokaun A. 一维和二维核磁共振原理[M].北京:科学出版社,1997:6—12.

共引文献47

同被引文献13

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部