期刊文献+

气相-质谱联用法测定主流烟气总粒相物中苯并[a]芘测量不确定度评定 被引量:1

Uncertainty Assessment in Determination of Benzoapyrene in Particulate Matter of Mainstream Smoke by GC-MS
下载PDF
导出
摘要 通过对气相-质谱联用法测定主流烟气总粒相物中苯并[a]芘测量不确定度的分析研究,确定了测量不确定度的主要来源为标准溶液的配制,标准曲线拟合,测量重复性及方法回收率.测量结果的标准不确定度为0.338 ng/cig.在95%的置信度下,取包含因子为k=2,其扩展不确定度为0.676 ng/cig. Uncertainty in determination of Benzoapyrene in particulate matter of mainstream smoke by gas chromatography/mass spectrometry (GC - MS) has been assessed. It showed that the preparation of standard solution, the standard curve fitting, the repeatability of detection and the recovery of the method are main factors which influence the uncertainty. The standard uncertainty was 0. 338 ng/cig, the extend uncertainty (a = 0.05, k = 2) was 0. 676 ng/cig.
出处 《哈尔滨师范大学自然科学学报》 CAS 2014年第5期82-86,共5页 Natural Science Journal of Harbin Normal University
关键词 气相-质谱联用法 苯并[A]芘 不确定度 Gas chromatography/mass spectrometry Benzoapyrene Uncertainty
  • 相关文献

参考文献12

二级参考文献67

共引文献82

同被引文献31

  • 1尚德军,王军.测量不确定度的研究和应用进展[J].理化检验(化学分册),2004,40(10):623-627. 被引量:36
  • 2杨琴,彭忠,郑利锋.卷烟主流烟气中CO测量不确定度的评定[J].现代测量与实验室管理,2004,12(6):37-38. 被引量:5
  • 3冯秀文,高俊琴.用回收率检验分析方法的准确度[J].光谱学与光谱分析,1995,15(3):87-90. 被引量:10
  • 4CNAS-GL06.化学分析中不确定度的评估指南[S].,2006.133.
  • 5谢剑平.烟草与烟气化学成分[M].化学工业出版社,2010:28-57.
  • 6BIPM,ISO.Guide to the Expression of Uncertainty in Measurement.Geneve:ISO,1995.
  • 7Ellison S L R,Rosslein M,Williams A.Quantifying uncertainty in analytical measurement.2nd ed.UK:EURACHEM/CITAC,2000.
  • 8Max R.APLAC interpretation and guidance on the estimation of uncertainty of measurement in testing[J].J AOAC international,2003,86(5):1070-1076.
  • 9Silva GMS.Uncertainty of analytical determination[J].J AOAC international,2003,86(5):1077-1184.
  • 10JJF 1059.1测量不确定度评定与表示[S].2012.