双能CT区分两种造影剂的研究 被引量:2

Differentiation of Two Contrast Media Using Dual-energy CT
下载PDF
导出
摘要 目的探讨双能CT扫描区分铋造影剂和钆造影剂的可行性。资料与方法将两组不同比例的铋造影剂和钆造影剂混合物置于试管内,第一组为碘海醇注射液和钆特酸葡胺注射液,体积比分别为0∶1、1∶6、1∶1、6∶1、1∶0;第二组为碘海醇注射液和果胶胶体铋溶液,质量浓度分别为0∶1、1∶6、1∶1、6∶1、1∶0。使用双能CT扫描后,以80 keV进行单能量图像重建,应用Liver VNC软件将双能数据进行虚拟平扫及碘含量分析。结果在80 keV虚拟单能量成像的条件下,第一组混合液的CT值为379~383 HU,第二组混合液的CT值为170~173 HU,且两组实际碘浓度与测量碘浓度之间差异均无统计学意义(P〉0.05)。物质分离图像中,碘造影剂和钆造影剂区分不明显,碘和铋造影剂区分明显。结论双能CT可以区分铋和钆这两种衰减率不同的造影剂,原子序数相差越大,区分效果越明显。 Purpose To evaluate the feasibility of using dual-energy CT scanner to differentiate two contrast media bismuth and gadolinium. Materials and Methods Two phantoms containing contrast media of different ratio were scanned on dual-energy CT. Group 1 was mixture of iodine and gadolinium solution with volume ratio of 0 : 1, 1 : 6, 1 : 1, 6 : 1 and 1 : 0. Group 2 was mixture of iodine and bismuth solution with mass ratio of 0: 1,1 : 6, 1 : 1, 6 : 1, 1 : 0. Monoenergetic image reconstruction was performed at 80 keV. Liver VNC software was used for virtual scanning and iodine concentration analysis. Results Under 80 keV reconstruction, the measured CT attenuation of Group 1 was 379- 383 HU, and 170-173 HU in group 2. The iodine concentration of two groups was not significantly different between the calculated and actual iodine concentration (P〉0.05). The dual-energy CT can distinguish two contrast media. Iodine and gadolinium contrast media were not statistically different, while iodine and bismuth contrast media agent could be easily differentiated. Conclusion Dual-energy CT can distinguish two contrast media with different attenuation.
出处 《中国医学影像学杂志》 CSCD 北大核心 2014年第9期701-703,共3页 Chinese Journal of Medical Imaging
关键词 体层摄影术 X线计算机 造影剂 Tomography, X-ray computed Contrast media Bismuth Gadolinium
  • 相关文献

参考文献10

二级参考文献52

  • 1兰永树,黄文龙,朱德强,唐光才.影响CT值测量的因素及对策的探讨[J].实用放射学杂志,2006,22(3):352-355. 被引量:27
  • 2张志浩,李严斌,梅劲,彭田红,Christopher R.Geddes,唐茂林.应用放射造影术进行血管3D可视化研究初探[J].中国临床解剖学杂志,2006,24(3):255-258. 被引量:52
  • 3楼新法,梅劲,Christopher R.Geddes,唐茂林.明胶-氧化铅血管造影术的优化[J].中国临床解剖学杂志,2006,24(3):259-262. 被引量:133
  • 4刘成伟,苏唏,宋丹,彭剑,熊辉,朱国英.8937例冠状动脉造影分析及临床意义的探讨[J].中华内科杂志,2007,46(7):574-575. 被引量:3
  • 5王同权,魏晓东,李宏杰,路伟.X射线的衰减和能量沉积计算[J].原子能科学技术,2007,41(4):399-403. 被引量:20
  • 6Matsumoto K, Jinzaki M, Tanami Y, et al. Virtual monochromatic spectral imaging with fast kilovohage switching: improved image quality as compared with that obtained with conventional 120 kVp CT. Radiology, 2011 ,In press.
  • 7Lv P, Lin XZ, Li J, et al. Differentiation of small hepatic hemangioma from small hepatocellular carcinoma: recently introduced soectral CT method. Radioloav. 2011. In nress.
  • 8Lin XZ, Miao F, Li JY, et al. High-definition CT gemstone spectral imaging of the brain: initial results of selecting optimal monochromatic image for beam-hardening artifacts and image noise reduction. J Comput Assist Tomogr, 2011, 35:294-297.
  • 9Zhao LQ, He W, Li JY, et al. Improving image quality in portal venography with spectral CT imaging. Eur J Radiol, 2011, In press.
  • 10Feuerlein S, Roessl E, Proksa R, et al. Muhienergy photoncounting K-edge imaging: potential for improved luminal depiction in vascular imaging. Radiology, 2008, 249 : 1010-1016.

共引文献429

同被引文献17

引证文献2

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部