摘要
HPGeγ谱分析中,一般通过测量222Rn的子体来确定样品中226Ra的含量,这种方法需要对样品密封数周后测量,无法实现样品中226Ra含量的快速测量。文章分析了样品γ谱中186 ke V全能峰的计数来源,对使用186 ke V全能峰计算出的226Ra的含量结果进行修正,得到较好的实验结果,实现了用186 ke V全能峰快速测量样品中Ra的含量。
In general , the content of 226 Ra in a sample is deduced from results of 222 Rn progeny in HPGeγspectral analysis .Because the sample should be sealed for several weeks before the measurement , we can not get the results quickly through this method .The paper analyzes the source of the counts of the full -en-ergy peak in 186 keV in the full-energy peak , and modifies the result of 226 Ra content calculated by using the full-energy peak in 186 keV.The study gets good experiment results and realizes the rapid measurement of Ra in samples using the full -energy peak in 186 keV.
出处
《核电子学与探测技术》
CAS
CSCD
北大核心
2014年第9期1132-1134,共3页
Nuclear Electronics & Detection Technology