期刊文献+

基于微流控芯片的圆褐固氮菌浓度快速检测方法 被引量:3

Rapid Detection of Azotobacter chroococcum Concentration Based on Microfluidic Chip
下载PDF
导出
摘要 针对当前微生物肥料中圆褐固氮菌浓度检测方法耗时长、自动化程度低等缺点,提出了一种基于微流控芯片的圆褐固氮菌浓度检测方法。设计了具有自动进样、混合稀释以及电化学循环伏安检测等功能的微流控芯片,构建了微流控检测系统实验平台,优化了系统中的进样速度、扫描速率等工艺参数,并对检测系统的性能指标进行了实验验证。结果表明,与国标平板计数法相比,所提出方法不仅能够满足微生物肥料检测的精度要求,而且具备较快的检测速度及较高的检测自动化程度。 Aiming at the problem that the concentration detection of Azotobacter chroococcum in microbial fertilizer suffers from time-consuming and low automation level, a detecting method of Azotobacter chroococcum concentration based on mierofluidic chip was proposed. A dedicated microfluidic chip with automatic sample injecting, dilution and cyclic vohammetry detection was designed, and the experiment platform based on microfluidic detection system was bulit. The optimum scan rate and feed rate of concentration detection in the microfluidic detection system were determined, and the performance index of the system was experimentally verified. Experiment results showed that, compared with plate counting method, the proposed method met the requirement of microbial fertilizer detection and realized rapid detection with increased automation.
出处 《农业机械学报》 EI CAS CSCD 北大核心 2015年第1期155-159,共5页 Transactions of the Chinese Society for Agricultural Machinery
基金 江苏省自然科学基金资助项目(BK20131250) 江苏省高校优势学科建设工程资助项目(PAPD)(苏政办发(2011)6号)
关键词 微生物肥料 圆褐固氮菌 微流控芯片 快速检测 Microbial fertilizer Azotobacter chroococcum Microfluidic chip Rapid detection
  • 相关文献

参考文献13

  • 1刘鹏,刘训理.中国微生物肥料的研究现状及前景展望[J].农学学报,2013,3(3):26-31. 被引量:75
  • 2Mohammadi K, Heidari G, Khalesro S, et al. Soil management, microorganisms and organic matter interactions: a review[ J]. African Journal of Biotechnology, 2012, 86 (10) : 19840 -19849.
  • 3李俊,姜昕,李力,沈德龙.微生物肥料的发展与土壤生物肥力的维持[J].中国土壤与肥料,2006(4):1-5. 被引量:135
  • 4GB20287-2006农用微生物菌剂[S].2006.
  • 5张荣标,黄义振,孙晓军,杨宁,洪杨英.基于图像处理的圆褐固氮菌浓度快速检测方法[J].农业机械学报,2012,43(10):174-178. 被引量:3
  • 6李武,王凌华,赵勇,张晓君,潘迎捷,赵立平.分子生态学方法在微生物肥料质量监测中的应用[J].微生物学通报,2006,33(1):53-58. 被引量:4
  • 7Setterington E B, Alocilja E C. Rapid electrochemical detection of polyaniline-labeled Escherichia coli O157: H7[ J]. Biosensors and Bioelectronics, 2011, 26 (5) : 2208 - 2214.
  • 8Foudeh A M, Didar T F, Veres T, et al. Microfluidic designs and techniques using lab-on-a-chip devices for pathogen detection for point-of-care diagnostics[ J]. Lab on a Chip, 2012, 12(18) : 3249 - 3266.
  • 9Velusamy V, Arshak K, Korostynska O, et al. An overview of foodborne pathogen detection: in the perspective of biosensors[ J]. Biotechnology Advances, 2010, 28 (2) : 232 - 254.
  • 10Choi J, Sung J H, Shuler M L, et al. Investigation of portable in situ fluorescence optical detection for microfluidic 3D cell culture assays[J]. Optics Letters, 2010, 35(9) : 1374 -1376.