期刊文献+

螺旋断层放射治疗系统的影响因素分析及对策 被引量:2

Research on factors affecting operation rate of Tomo Therapy system and their countermeasures
下载PDF
导出
摘要 目的:针对螺旋断层放射治疗(Tomo Therapy)系统故障率较传统加速器高的问题,分析影响Tomo Therapy开机率的相关因素,并提出相应对策,以提高设备的使用效率。方法:回顾性分析日常故障记录,总结影响Tomo Therapy开机率的主要因素。结果:影响Tomo Therapy开机率的主要因素分别为:温度冷却系统、空气压缩系统、易损性部件和数据安全系统。结论:对影响螺旋断层放射治疗系统开机率的4个因素提出了相应的保养和维护对策,以降低设备故障率,保证系统安全、高效的运行。 Objective: To analyze the factors affecting operation rate of TomoTherapy system, which had higher failure rate compared to the traditional accelerator. Relevant countermeasures had been proposed to increase the operation rate. Methods:To discuss the main factors affecting operation rate of TomoTherapy, the failure record had been analyzed. Results: Temperature cooling system, compressed air system, deteriorating parts and database system had been summarized to be the main factors caused the failure of the TomoTherapy. Conclusion:Some corresponding maintenance measures specific to the four factors had been proposed in order to reduce TomoTherapy failure rate and increase TomoTherapy service efficiency.
出处 《中国医学装备》 2015年第1期27-29,共3页 China Medical Equipment
基金 广东省社会发展领域科技计划(2013B021800051)"非小细胞肺癌螺旋断层放疗食管损伤相关因素分析与预测"
关键词 螺旋断层放射治疗系统 开机率 温度冷却系统 空气压缩系统 易损性部件 数据系统 TomoTherapy system Operation rate Temperature cooling system Compressed air system Deteriorating parts Database system
  • 相关文献

参考文献15

二级参考文献111

  • 1沈亚丽,许峰.影像引导放射治疗系统[J].国际肿瘤学杂志,2006,33(9):707-710. 被引量:9
  • 2姜瑞瑶.数字化放疗科的设备配置[J].医疗设备信息,2006,21(12):42-43. 被引量:5
  • 3周颖娟,黄劭敏,邓小武.用放射性铬胶片进行调强放疗剂量验证的研究[J].中华放射肿瘤学杂志,2007,16(4):307-312. 被引量:12
  • 4Maekie TR, Holmes T, Swerdloff S, et al. Tomotherapy: A new concept for the delivery of conformal radiotherapy [J]. Med Phys, 1993, 20(6): 1709-1719.
  • 5Mackie TR, Balog J, Ruchala K, et al. Tomotherapy [J]. Semin Radiat Oncol, 1999, 9(1): 108-117.
  • 6Balog J, Holmes T, Vaden R. Helical tomotherapy dynamic quality assurance [J]. Med Phys, 2006, 33:3939-3950.
  • 7Fenwick JD, Tome WA, Jaradat HA, et al. Quality assurance of a hellcal tomotherapy machine [J]. Phys Med Biol, 2004, 49:2933-2953.
  • 8Kutcher GJ, Coia L, Gillin M, et al. Comprehensive QA for radiation oncology: report of AAPM Radiation Therapy Committee Task Group 40 [J]. Med Phys, 1994, 21(4):581-618.
  • 9Almond PR, Biggs PJ, Coursey BM, et al. AAPM's TG51 protocol for clinical reference dosimetry of high-energy photon and electron bearps[J]. Med Phys, 1999, 26(9):1847-1870.
  • 10Balog J, Mackie TR, Pearson D, et al. Benchmarking beam alignment for a clinical helical tomotherapy device [J]. Med Phys, 2003, 30(6): 1118-1127.