导出
摘要 以Bi(Ⅲ)做内标,结合丝网印刷电极与微型化的电化学检测系统,建立了一种警报型即时检测方法,可以对自来水中铅含量是否超标进行快速准确的判断。采用方波溶出伏安法测定了含10μg/L铅标准溶液的IPb/IBi并计算得出阈值。讨论了3种电极预处理方法以及不同浓度Cu2+对IPb/IBi测定的干扰与消除。实验结果表明:电极经1.8 V阳极氧化处理2 min后,电子转移电阻大大减小,由4 348Ω减小至715Ω,峰电流增加1.6倍,峰电位之差△Ep由212 m V减小至113 m V。当加入2.0 mmol/L Fe(CN)64-,Cu2+干扰可有效消除。将这种方法用于自来水样的IPb/IBi测定,与阈值对比给出Yes/No判断结果,通过ICP-MS验证方法的有效性,结果令人满意。作为一种简捷、快速和便携的检测平台,该方法无需测出Pb2+的准确浓度,具有巨大的应用潜力。 Combined with screen printed electrode and custom USB electrochemical system(μECS), bismuth served as internal standard is used for the construction of alert-type point-of-care testing method to decide whether the lead level has excessed a regulated threshold. The threshold is obtained by determining standard solution containing 10 μg/L Pb^2 +with square wave anodic stripping voltammetry(SWV). Three electrode pretreatment methods as well as the interference and removal of Cu^2+were discussed. Results indicated that a remarkable decreased charge transfer resistance from 4 348 Ω to 715Ω was observed after electrode was pretreated at 1.8 V for 2 min and the peak current increased 1.6-fold with peak potential separation decreasing from 212 m V to 113 m V. The detrimental effect of Cu^2+can be circumvented effectively by adding 2.0mmol/L Fe(CN)6^4-. The proposed method is applied to the decision of Pb2+in tap water samples and to give a Yes/No signal.ICP-MS analysis is performed to validate the reliability, and the results are satisfying. Simple, rapid and portable, the platform holds great potential for the application in daily life.
出处 《环境科学与技术》 CAS CSCD 北大核心 2015年第1期49-54,共6页 Environmental Science & Technology
基金 国家自然科学基金面上基金(41276093)
关键词 警报型 即时检测 铅离子 方法研究 alert-type point-of-care testing lead method study
  • 相关文献

参考文献26

  • 1Arduini F,Calvo J Q,Amine A,et al.Bismuth-modified electrodes for lead detection[J].Tr AC-Trends in Analytical Chemistry,2010,29(11):1295-1304.
  • 2吴志珊,宋伟,韩倩,杨丽莉,许丹科,钟文英.铋膜丝网印刷电极现场快速测定工业废水中Pb^(2+)、Cd^(2+)[J].环境化学,2013,32(4):692-697. 被引量:7
  • 3Tercier Waeber M L,Taillefert M.Remote in situ voltammetric techniques to characterize the biogeochemical cycling of trace metals in aquatic systems[J].Journal of Environmental Monitoring,2008,10(1):30-54.
  • 4Chen C,Niu X H,Chai Y,et al.Determination of lead(Ⅱ)using screen-printed bismuth-antimony film electrode[J].Electroanalysis,2013,25(6):1446-1452.
  • 5Song W,Zhang L,Shi L,et al.Simultaneous determination of cadmium(Ⅱ),lead(Ⅱ)and copper(Ⅱ)by using a screenprinted electrode modified with mercury nano-droplets[J].Microchimica Acta,2010,169(3/4):321-326.
  • 6Fang H L,Zheng H X,Ou M Y,et al.One-step sensing lead in surface waters with screen printed electrode[J].Sensors and Actuators B-Chemical,2011,153(2):369-372.
  • 7Pan D W,Zhang L,Zhuang J M,et al.On-line determination of lead in tap waters at two-step prepared bismuth electrode[J].International Journal of Electrochemical Science,2011,6(7):2710-2717.
  • 8Quintana J C,Arduini F,Amine A,et al.Part two:analytical optimization of a procedure for lead detection in milk by means of bismuth-modified screen-printed electrodes[J].Analytica Chimica Acta,2012,736:92-99.
  • 9何洪亮,王永凤.铅测定方法的研究进展[J].职业与健康,2012,28(11):1395-1397. 被引量:8
  • 10Choi H S,Kim H D.Development of a portable heavy metal ion analyzer using disposable screen-printed electrodes[J].Bulletin of the Korean Chemical Society,2009,30(8):1881-1883.

二级参考文献46