摘要
目的:建立同时测定杆菌肽原料药中主成分的含量及有关物质的方法。方法:采用高效液相色谱法。色谱柱为XTerra MS?C18,流动相为水-0.2 mol/L磷酸氢二钾溶液(用磷酸调节p H至6.0)-甲醇-乙腈(28∶14∶52∶6,V/V/V/V),主成分杆菌肽A检测波长为254 nm,杆菌肽F(杂质F)检测波长为300 nm,柱温为30℃,流速为1.0 ml/min,进样量为100μl。结果:杆菌肽A质量浓度在0.4~4 mg/ml范围内与峰面积呈良好的线性关系(r=0.999 9);精密度、稳定性、重复性试验的RSD≤0.5%;平均加样回收率为100.1%,RSD为0.5%(n=9)。主成分杆菌肽A与其他成分分离度较好;杂质F含量在6.0%以下。结论:本方法操作简便、快捷,结果准确、可靠,可用于杆菌肽原料药的质量控制。
OBJECTIVE: To establish a method for simultaneous determination of main components and related substance in hacitracin. METHODS: HPLC method was adopted. The determination was performed on XTerra MS C18 column with mobile phase consisted of water-0.2 mol/L dibasic potassium phosphate solution (pH adjusted to 6.0)-methanol-acetonitrile (28 : 14 : 52 : 6, V/V/V/V). The detection wavelength for bacitracin A was set at 254 nm for main components and 300 nm for bacitracin F. The col- umn temperature was set at 30 ℃, the flow rate of 1.0 ml/min and the sample size was 100 μl. RESULTS:The linear range of baci- tracin A was 0.4-4 mg/ml(r=0.999 9) with an recovery of 100.1% (RSD=0.5% ,n=9). RSDs of precision, stability and reproduc- ibility tests were lower than 0.5%. Main component bacitracin A was separated from other components well. The content of bacitra- cin F was less than 6.0%. CONCLUSIONS: The method is simple, rapid, accurate and reliable, and it is applicable for the quality control of bacitraein.
出处
《中国药房》
CAS
北大核心
2015年第6期834-837,共4页
China Pharmacy