摘要 目的验证Sysmex XT-2000i全血细胞分析仪的性能。方法按照美国临床实验室标准化委员会(NCCLS)EP5-A2、EP15-A2、EP6-A2、EP9-A2文件要求,对该仪器的准确度、精密度和线性进行评价,并与长期参加卫生部室间质评的仪器——Sysmex XE-2100i的测定结果进行比较。结果该仪器的准确度、精密度、携带污染率和线性均在允许范围内。结论 Sysmex XT-2000i全血细胞分析仪各方面性能良好,测定结果准确可靠,主要性能指标符合临床实验室要求,是一种较理想的全血细胞分析仪。 Objective To perform the performance verification of the Sysmex XT-2000i hematology analyzer.Methods Accord-ing to requirements of NCCLS EP5-A2,EP15-A2,EP6-A2,EP9-A2 documents,the accuracy,precision and linearity of the Symex XT-2000i hematology were evaluated and the detection results were compared with those detected by the Sysmex XE-2100i which participated in the external quality assessment(EQA)of Ministry of Public Health for long time.Results The accuracy,precision, linearity and contamination rate all are within the permissible range.Conclusion The performance of Sysmex XT-2000i is good,the determination results are reliable.As a perfect hematology analyzer,it can satisfy the hematological analysis of hospital clinical la-boratory.
出处 《国际检验医学杂志》 CAS 2014年第24期3413-3414,3416,共3页 International Journal of Laboratory Medicine
关键词 血细胞计数仪 性能验证 准确度 hematology analyzer performance qualification accuracy
  • 相关文献

参考文献10

  • 1中围合格评定国家认可委员会.CNAS2CL02:2008医学实验室质最和能力认可准则[M].北京:中国合格评定国家认可委员会,2008.
  • 2Interference Testingin Clinical chemistry: Approved Guideline- Second Editiong,CLSI document EP02-A2(2005).
  • 3李燕,李山.临床实验室管理学[M].3版.北京:人民卫生出版社,2013.
  • 4国家食品药品监督管理局.体外诊断试剂分析性能评估系列指导原则,2009:20.
  • 5胡丽涛,王薇,王治国.血液分析仪的方法确认和性能验证[J].国际检验医学杂志,2011,32(13):1497-1499. 被引量:8
  • 6Clinical and Laboratory Standards Institute. EP9 A2 Method com- parision and bias estimation using patient samples approved guide- line[J]. Wayne,PA:CLS1,2002.
  • 7NCCLS. EPS-A2 Evaluation of Precision Performanee of Quanti tative Measurement Methods. Approved Guide2 line2Second Edi- tionJ]. Wayne, PA : NCCLS, 2004.
  • 8National Committee forC lin icalLaboratory S tand ard s. Evaluat ion of the linearity of quant itat ivem easu rein en t procedu res: a star itica app roach[J]. Approved gu ideline. 2nd ed. NCCLS do cure ent EP6 p2,Vol 23,2002.
  • 9National Comm ittee for Clinical Laboratory Standards. Prelim in ary valuat ion of quantitative clinical laboratory methods. A pproved guideline[J]. NCCLS document EP10-A2,Vol 22,2002.
  • 10王志国.临床检验方法确认与性能验证[M].北京:人民卫生出版社,2009:289.