摘要
目的建立测定大鼠血浆中黄体酮的液相色谱-串联质谱(LC-MS/MS)法。方法血浆样品采用液-液萃取法,经正己烷-二氯甲烷-异丙醇(100∶50∶5,V/V/V)提取后,以甲醇-水(75∶25,V/V)为流动相,以Hypersil ODS柱(50 mm×2.1 mm,5μm)为分析柱,采用ESI源以多反应监测(MRM)方式进行正离子检测。用于定量分析的离子反应为m/z 315.4→m/z(97.0+109.2)(黄体酮)和m/z 385.4→m/z 267.4(醋酸甲地孕酮,内标)。结果黄体酮血浆浓度测定方法的线性范围为0.5~200 ng/ml,日内、内间精密度(RSD)均〈6.7%,准确度(RE)在±6.2%之间。结论该方法分析时间短、操作简便、灵敏度高、专属性强,适用于黄体酮的药动学研究。
Objective To develop and validate an LC-M S/MS method for determination of progesterone in rat plasma.Methods Plasma samples were prepared by liquid-liquid extraction with a mixture of n-hexane/dichloromethane/isopropanol(100∶50∶5,V/V/V). Isocratic chromatographic separation was performed on a Hypersil ODS column(50 mm × 2.1 mm,5 μm,Thermo).The mobile phase was methanol-water(75∶25,V/V). Detection of progesterone and the internal standard(IS) megestrol acetate was achieved by ESI MS/MS in the positive ion mode using m/z 315.4 → m/z(97.0 + 109.2) and m/z 385.4 → m/z 267.4 transitions,respectively. Results The method was linear from 0.5 to 200 ng/ml when 100 μl plasma was analyzed. The lower limit of quantification was 0.5 ng/ml. Both the inter- and intra-day precision values were below 6.7%,and the accuracy(relative error)was within ±6.2% in all quality control samples. Conclusion The method provides a specific,sensitive and rapid means for the determination of progesterone in rat plasma,and completely meets the requirements for pharmacokinetic study.
出处
《国际药学研究杂志》
CAS
CSCD
北大核心
2015年第1期107-111,126,共6页
Journal of International Pharmaceutical Research
基金
国家自然科学基金青年基金资助项目(81001468)
国家"重大新药创制"科技重大专项资助项目(2012ZX09301003001)