期刊文献+

基于模型的迭代重建算法优化低管电压儿童胸部CTA图像质量 被引量:13

Full model-based iterative reconstruction in optimizing image quality in low tube voltage chest CTA images of children
下载PDF
导出
摘要 目的评价基于模型的迭代重建算法(MBIR)优化低管电压儿童胸部CTA图像质量的价值。方法收集接受胸部低管电压(80kVp)CTA的儿童患者41例,对原始数据进行MBIR、30%自适应迭代重建(ASiR)与70%滤波反射投影(FBP)混合及FBP重建,评价图像整体质量及细小血管显示能力,测量降主动脉、同层面背部肌肉和脂肪噪声值,计算降主动脉SNR和CNR。结果 MBIR图像整体质量评分和细小血管显示能力评分为均明显优于30%ASiR与70%FBP混合图像及FBP图像(P均<0.05);MBIR图像中各组织噪声值明显低于30%ASiR与70%FBP混合图像和FBP图像(P均<0.05),降主动脉SNR和CNR(28.30±7.85、24.61±7.52)明显高于30%ASiR与70%FBP混合图像(15.80±5.73、14.19±4.22)和FBP图像(12.06±3.66、10.59±3.53,P均<0.05)。结论 MBIR用于处理儿童低管电压胸部CTA图像,图像质量明显高于ASiR及FBP图像,值得临床推广应用。 Objective To evaluate the feasibility of full model-based iterative reconstruction(MBIR)in optimizing image quality in low tube voltage chest CTA images of children.Methods Totally 41 children who underwent low tube voltage(80kVp)chest CTA were included.The images were retrospectively reconstructed with MBIR,30% of adaptive statistical iterative reconstruction(ASiR)+70% filtered back-projection(FBP)and FBP.The image quality and ability to display small arteries were evaluated,the image noise value of descending aorta,muscle and fat were measured and SNR,CNR of descending aorta were calculated.Results The score of image quality and ability to display small vessels were both higher in MBIR image compared with 30% ASiR+70% FBP image and FBP image(all P〈0.05).The noise value of each tissue in MBIR image was significantly lower than that in 30% ASiR+70% FBP image and FBP image(all P〈0.05).SNR and CNR of descending aorta in MBIR image(28.30±7.85,24.61±7.52)were both significantly higher compared with 30%ASiR+70% FBP image(15.80±5.73,14.19±4.22)and FBP image(12.06±3.66,10.59±3.53,all P〈0.05).Conclusion Compared with FBP and ASiR image,the image quality is significantly improved by the MBIR in children low tube voltage chest CTA,which is worthy of popularizing.
出处 《中国医学影像技术》 CSCD 北大核心 2015年第2期302-305,共4页 Chinese Journal of Medical Imaging Technology
基金 首都临床特色应用研究(Z141107002514005)
关键词 体层摄影术 X线计算机 图像处理 计算机辅助 噪声 儿童 辐射剂量 Tomography X-ray computed Image processing computer-assisted Noise Child Radiation dosage
  • 相关文献

参考文献16

  • 1Coakley FV, Gould R, Yeh BM, et al. CT radiation dose: What can you do right now in your practice? AJR Am J Roentgenol, 2011,196(3) :619-625.
  • 2张祺丰,彭芸,李剑颖,刘玥,曾津津.64排螺旋CT自动管电流调节技术降低儿童胸部扫描剂量的应用研究[J].中国医学影像技术,2009,25(9):1591-1594. 被引量:42
  • 3Lee S, Yoon SW, Yoo SM, et al. Comparison of image quality and radiation dose between combined automatic tube current mod- ulation and fixed tube current technique in CT of abdomen and pelvis. Acta Radiol, 2011,52(10) : 1101-1106.
  • 4Hara AK, Paden RG, Silva AC, et al. Iterative reconstruction technique for reducing body radiation dose at CT: Feasibility study. AJR Am J Roentgenol, 2009,193(3) :764-771.
  • 5Kundel HL, Polansky M. Measurement of observer agreement. Radiology, 2003,228(2) : 303-308.
  • 6Naidich DP, Marshall CH, Gribbin C, et al. Low-dose CT of the lungs: preliminary observations. Radiology, 1990, 175 (3) : 729- 731.
  • 7Leipsic J, Nguyen G, Brown J, et al. A prospective evaluation of dose reduction and image quality in chest CT using adaptive statis- tical iterative reconstruction:&JR Am J Roentgenol, 2010, 195 (5) : 1095-1099.
  • 8Vorona GA, Ceschin RC, Clayton BL, et al. Reducing abdominal CT radiation dose with the adaptive statistical iterative recon- struction technique in children: A {easibility study. Pediatr Radi- ol, 2011,41(9):1174-1182.
  • 9徐姝,侯阳,郭启勇.应用迭代重建技术的低剂量增强CT评估肝脏肿瘤的可行性[J].中国医学影像技术,2011,27(8):1633-1636. 被引量:47
  • 10Prakash P, Kalra MK, Kambadakone AK, et al. Reducing ab- dominal CT radiation dose with adaptive statistical iterative re- construction technique. Invest Radiol, 2010,45(4) :202 210.

二级参考文献23

  • 1Andrea Laghi,Franco Iafrate,Marco Rengo,Cesare Hassan.Colorectal cancer screening:The role of CT colonography[J].World Journal of Gastroenterology,2010,16(32):3987-3994. 被引量:6
  • 2李剑颖.GE VCT——高图像质量和低剂量的和谐[J].CT理论与应用研究(中英文),2006,15(3):57-60. 被引量:5
  • 3郭洪涛,刘勇,袁淑华.CT剂量指数(CTDI)测量研究[J].中国测试技术,2007,33(4):33-36. 被引量:32
  • 4Itoh S,Ikeda M,Schweden F,et al.Lung cancer screening:minimum tube current required for helical CT.Radiology,2000,215(1):175-183.
  • 5Brener DJ,Elliston CD,Hall EJ,et al.Estimated risks of radiation-induced fatal cancer from pediatric CT.AJR Am J Roentgenol,2001,176(2):289-296.
  • 6Hojreh A,Kainberger F,Puig S.Low dose multislice CT in the pediatric patient.Radiology,2003,43(12):1051-1055.
  • 7Mettler FA Jr,Wiest PW,Locken JA,et al.CT scanning:patterns of use and dose.J Radiol Prot,2000,20(4):353-359.
  • 8Greess H,Nmayr A,Wolf H,et al.Dose reduction in CT examination of children by an attenuation-based on-line modulation of tube current (CARE Dose).Eur Radiol,2002,12(6):1571-1576.
  • 9Prasad SR,Wittram C,Shepard JA,et al.Standard-dose and 50%-reduced-dose chest CT:comparing the effect on image quality.AJR Am J Roentgenol,2002,179(2):461-465.
  • 10中島正俊.撮影条件自動適正化機能を用いたル-チン化への検討.日本放射線技術学会雑誌,2002,58(4):502-510.

共引文献90

同被引文献58

引证文献13

二级引证文献68

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部