导出
摘要 目的:探索研究6D-skull(六维颅骨追踪)的曝光参数选择。方法 :以模型LUCY(体模名称)为实验模体,改变曝光参数的电压、电流和时间值,采集影响追踪的误差与参考数据,对数据进行定量分析。结果:6D-skull的曝光参数选择为:电压97~107 k V为最佳(电流100 m A、曝光时间75 ms时),电流100~200 m A为最佳(电压90 k V、曝光时间75 ms时),曝光时间100~125 ms为最佳(电压90 k V、电流100 m A时)。结论 :所得结果可指导曝光参数选择,保证治疗的稳定性,一定程度上缩短了治疗时间。 Objective To explore the selection of exposure parameters for 6D skull tracing of Cyberknife G3 system.Methods Model LUCY was used as the experiment phantom, and the exposure parameters such as voltage, current and time were changed to acquire the error and reference data influencing the tracing. The data were analyzed quantitatively.Results The optimal voltage was between 97 and 107 k V with the current of 100 m A and exposure time of 75 ms. The optimal current was between 100 and 200 m A with the voltage of 90 k V and the exposure time of 75 ms. The optimal exposure time was between 100 and 125 ms with the voltage of 90 k V and the current of 100 m A. Conclusion The obtained results can be used to guide the selection of the exposure parameters, with the treatment stability ensured and the treatment time decreased.
机构地区 解放军
出处 《医疗卫生装备》 CAS 2015年第2期46-48,124,共4页 Chinese Medical Equipment Journal
关键词 射波刀 影像引导放疗 曝光参数 6D-skull Cyberknife image-guided radiotherapy exposure parameter 6D-skull
  • 相关文献

参考文献11

二级参考文献57

  • 1Olivier Morin,Amy Gillis,Josephine Chen,Michèle Aubin,M.Kara Bucci,Jean Pouliot,索京涛,陆嘉德.兆伏级锥形束CT:系统说明及IGRT临床应用介绍[J].中国癌症杂志,2006,16(6):513-524. 被引量:9
  • 2John R.Adler.An Overview of CyberKnife Radiosurgery[J].Chinese Journal of Clinical Oncology,2006,3(4):229-243. 被引量:7
  • 3陈光耀.临床肿瘤医师如何善用放射外科的新利器[J].中国肿瘤临床,2007,34(1):1-5. 被引量:16
  • 4陈光耀.放射外科的最新进展[J].肿瘤,2007,27(7):588-590. 被引量:7
  • 5Adler JR Jr,Murphy MJ,Chang SD,et al.Image-guided robotic radiosergery.Neurosurgery,1999,44:1299-1306.
  • 6Harada T,Shirato H,Ogura S,et al.Real-time tumor-tracking radiation therapy for lung carcinoma by the aid of insertion of a gold mark using bronchofiberscopy.Cancer,2002,95:1720-1727.
  • 7Welsh JS,Berta C,Borzillary S,et al.Fiducial markers implanted during prostate brachytherapy for guiding conformal external beam radiation therapy.Technol Cancer Res Treat,2004,3:359-364.
  • 8Olender D.Fiducials for target localization//Heilbrun MP,eds.Cyberknife radiosurgery:a practical guide.Sunnyvale.CA:Cyberknife Society,2003:80-94.
  • 9Reichner CA,Collins BT,Gagnon GJ,et al.The placement of gold fiducials for cyberknife stereotactic radiosurgery using a modified transbronchial needle aspiration technique.J Bronchology,2005,12:193 -195.
  • 10Fu D,Kuduvalli G,Maurer CR,et al.3D target localization using 2D local displacements of skeletal structures in orthogonal X-ray images for image-guided spinal radiosurgery.Int J Comp Assis Radiol Surg,2006,Suppl 1:198-200.

共引文献58

同被引文献12

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部