期刊文献+

亚麻木酚素缓释片的质量标准研究 被引量:1

Study on the Quality Specification of Secoisolariciresinol Diglucoside Sustained-release Tablets
原文传递
导出
摘要 目的:为建立亚麻木酚素缓释片的质量标准提供参考。方法:采用高效液相色谱法测定亚麻木酚素缓释片的含量,并测定其体外释放度。色谱柱为Waters Symmetny Shield-C18,流动相为甲醇-水(35∶65,V/V),流速为1.0 ml/min,检测波长为282 nm,柱温为室温,进样量为10μl。结果:亚麻木酚素的质量浓度在0.018~0.225 mg/ml范围内与峰面积呈良好的线性关系(r=0.999 5);精密度、稳定性、重复性试验的RSD〈2%;平均加样回收率为99.59%,RSD=1.95%(n=6)。亚麻木酚素缓释片在2、6、12 h的累积释放度分别为10%~30%、50%~70%、80%~100%,均符合规定。结论:该方法准确、简便,专属性、重复性好,可用于亚麻木酚素缓释片的质量控制。 OBJECTIVE:To provide reference to establish the quality specification of Secoisolariciresinol diglucoside(SDG)sustained-release tablets. METHODS:The content of SDG sustained-release tablets was determined by HPLC and the release property was studied in vitro. HPLC conditions were as follows:Waters Symmetny Shield-C18 column,methanol-water(35∶65,V/V)as mobile phase at a flow rate of 1.0 ml/min. The detection wavelength was 282 nm,the column temperature was room temperature and injection volume was 10 μl. RESULTS:The concentration of SDG had a good linear relationship with its peak area value in the range of 0.018~0.225 mg/ml(r=0.999 5). The RSDs of precision,stability and reputability test were less than 2%;the average sample recovery were 99.59%,RSD=1.95%(n=6). The accumulative release amount of SDG sustained-release tablets in 2,6,12 h were 10%-30%,50%-70% and 80%-100%,respectively,which conform to the provisions. CONCLUSIONS:The method is accurate,simple,specificity and reproducible,and can be used for the quality control of SDG sustained-release tablets.
出处 《中国药房》 CAS 北大核心 2015年第9期1286-1288,共3页 China Pharmacy
关键词 亚麻木酚素 缓释片 高效液相色谱法 体外释放度 Secoisolariciresinol diglucoside Sustained-release tablets HPLC in vitro release
  • 相关文献

参考文献7

二级参考文献52

共引文献44

同被引文献35

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部